There is little information about the management
of peri-prosthetic fracture of the humerus after total shoulder replacement
(TSR). This is a retrospective review of 22 patients who underwent
a revision of their original shoulder replacement for peri-prosthetic
fracture of the humerus with bone loss and/or loose components.
There were 20 women and two men with a mean age of 75 years (61
to 90) and a mean follow-up 42 months (12 to 91): 16 of these had
undergone a previous revision TSR. Of the 22 patients, 12 were treated
with a long-stemmed humeral component that bypassed the fracture.
All their fractures united after a mean of 27 weeks (13 to 94).
Eight patients underwent resection of the proximal humerus with
endoprosthetic replacement to the level of the fracture. Two patients
were managed with a clam-shell prosthesis that retained the original
components. The mean Oxford shoulder score (OSS) of the original
TSRs before peri-prosthetic fracture was 33 (14 to 48). The mean
OSS after revision for fracture was 25 (9 to 31). Kaplan-Meier survival
using re-intervention for any reason as the endpoint was 91% (95%
confidence interval (CI) 68 to 98) and 60% (95% CI 30 to 80) at
one and five years, respectively. There were two revisions for dislocation of the humeral head,
one open reduction for modular humeral component dissociation, one
internal fixation for nonunion, one trimming of a prominent screw
and one re-cementation for aseptic loosening complicated by infection,
ultimately requiring excision arthroplasty. Two patients sustained
nerve palsies. Revision TSR after a peri-prosthetic humeral fracture associated
with bone loss and/or loose components is a salvage procedure that
can provide a stable platform for elbow and hand function. Good
rates of union can be achieved using a stem that bypasses the fracture.
There is a high rate of complications and function is not as good as
with the original replacement.
This is a retrospective review of the results of the Acclaim total elbow replacement in 11 older patients aged ≥ 65 years with primary osteoarthritis of the elbow, with a mean follow-up of 57.6 months (30 to 86.4). Significant reductions in pain and improvement in range of movement and function were recorded. Radiological review revealed two patients with 1 mm lucencies in a single zone, and one patient with 1 mm lucencies in two zones. No components required revision. There were no deep infections, dislocations or mechanical failures. Complications included one intra-operative medial condylar fracture and one post-operative transient ulnar neuropathy, which resolved. This study demonstrates that the Acclaim prosthesis provides good symptomatic relief and improvement of function in patients with primary osteoarthritis, with low rates of loosening or other complications. This prosthesis can therefore be considered for patients aged ≥ 65 years with primary osteoarthritis of the elbow.
The Acclaim total elbow replacement is a modular system which allows implantation in both unlinked and linked modes. The results of the use of this implant in primary total elbow replacement in 36 patients, operated on between July 2000 and August 2002, are presented at a mean follow-up of 36 months (24 to 49). Only one patient did not have good relief of pain, but all had improved movement and function. No implant showed clinical or radiological loosening, although one had a lucent area in three of seven humeral zones. The short-term results of the Acclaim total elbow replacement are encouraging. However, 11 patients (30.5%) suffered an intra-operative fracture of the humeral condyle. This did not affect the outcome, or the requirement for further surgery, except in one case where the fracture failed to unite. This problem has hopefully been addressed by redesigning the humeral resection guide. Other complications included three cases of ulnar neuropathy (8.3%) and one of deep infection (2.8%).
Ulnar nerve function, during and after open reduction and internal fixation of fractures of the distal humerus with subperiosteal elevation of the nerve, was assessed by intra-operative neurophysiological monitoring. Intermittent recording of the compound muscle action potentials was taken from the hypothenar muscles in 18 neurologically asymptomatic patients. The mean amplitude of the compound muscle action potential after surgery was 98.1% (