Dislocation remains among the most common complications
of, and reasons for, revision of both primary and revision total
hip replacements (THR). Hence, there is great interest in maximising
stability to prevent this complication. Head size has been recognised
to have a strong influence on the risk of dislocation post-operatively.
As femoral head size increases, stability is augmented, secondary
to an increase in impingement-free range of movement. Larger head
sizes also greatly increase the ‘jump distance’ required for the
head to dislocate in an appropriately positioned cup. Level-one
studies support the use of
Large femoral heads have been used with increasing
frequency over the last decade. The prime reason is likely the effect
of large heads on stability. The larger head neck ratio, combined
with the increased jump distance of larger heads result in a greater
arc of impingement free motion, and greater resistance to dislocation
in a provocative position. Multiple studies have demonstrated clear
clinical efficacy in diminishing dislocation rates with the use
of large femoral heads. With crosslinked polyethylene, wear has
been shown to be equivalent between larger and smaller heads. However,
the stability advantages of increasing diameter beyond 38 mm have
not been clearly demonstrated. More importantly, recent data implicates
large heads in the increasing prevalence of groin pain and psoas impingement.
There are clear benefits with
Implantation of a large-diameter femoral head prosthesis with a metal-on-metal bearing surface reduces the risk of dislocation, increases the range of movement, minimises the risk of impingement and, in theory, results in little wear. Between February 2004 and March 2007 we implanted 100 consecutive total hip replacements with a metal-on-metal bearing and a large femoral head into 92 patients. There were 51 men and 41 women with a mean age of 50 years (18 to 70) at the time of surgery. Outcome was assessed using the Western Ontario McMaster University osteoarthritis index and the Harris hip score as well as the Devane activity score. These all improved significantly (p <
0.0001). At the last follow-up there were no cases of dislocation, no impingement, a good range of movement and no osteolysis, but seven revisions, two for infection and five for aseptic loosening. The probability of groin pain increased if the other acetabular component inclination exceeded 50° (p = 0.0007). At 4.8 years of follow-up, the projected survival of the Durom acetabular component, with revision for any reason, was 92.4% ( The design of the component made it difficult both to orientate and seat, which when combined with a poor porous coating, produced unpredictable fixation and a low survival at five years.
Objectives. An ongoing prospective study to investigate failing metal-on-metal
hip prostheses was commenced at our centre in 2008. We report on
the results of the analysis of the first consecutive 126 failed
mated total hip prostheses from a single manufacturer. Methods. Analysis was carried out using highly accurate coordinate measuring
to calculate volumetric and linear rates of the articular bearing
surfaces and also the surfaces of the taper junctions. The relationship
between taper wear rates and a number of variables, including bearing
diameter and orientation of the acetabular component, was investigated. Results. The measured rates of wear and distribution of material loss
from the taper surfaces appeared to show that the primary factor
leading to taper failure is the increased lever arm acting on this
junction in contemporary large-diameter metal-on-metal hip replacements. Conclusions. Our analysis suggests that varus stems, laterally engaging taper
systems and
Aims. A revision for periprosthetic joint infection (PJI) in total hip arthroplasty (THA) has a major effect on the patient’s quality of life, including walking capacity. The objective of this case control study was to investigate the histological and ultrastructural changes to the gluteus medius tendon (GMED) in patients revised due to a PJI, and to compare it with revision THAs without infection performed using the same lateral approach. Methods. A group of eight patients revised due to a PJI with a previous lateral approach was compared with a group of 21 revised THAs without infection, performed using the same approach. The primary variables of the study were the fibril diameter, as seen in transmission electron microscopy (TEM), and the total degeneration score (TDS), as seen under the light microscope. An analysis of bacteriology, classification of infection, and antibiotic treatment was also performed. Results. Biopsy samples from the GMED from infected patients revealed a
Objectives. Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of
Aims. The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. Methods. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083 metal-on-metal (MOM), 90 ceramic-on-ceramic (COC), and 81 surface arthroplasties (SAs). With the exception of the COC, all other groups used cobalt-chromium (CoCr) femoral heads. The mean follow-up was 10 (0.008 to 35) years. Descriptive statistics with revisions per 100 component years (re/100 yr) and survival analysis with revision for any reason as the endpoint were used to compare bearing surfaces. Results. XLPE liners demonstrated a lower cumulative incidence of revision at 15 years compared to the CPE-GA and CPE-GP groups owing to the absence of wear-related revisions (4% for XLPE vs 18%, p = 0.02, and 15%, p = 0.003, respectively). Revisions for adverse local tissue reactions occurred exclusively among the MOM (0.8 re/100 year) and SA groups (0.1 re/100 year). The revision rate for instability was lower among hips with 36 mm and larger head sizes compared to smaller head sizes (0.2% vs 2%, p < 0.001). Conclusion. The introduction of XLPE has eliminated wear-related revisions through 15-year follow-up compared to CPE-GP and CPE-GA. Dislocation incidence has been reduced with the introduction of
Aims. We have previously demonstrated raised cobalt and chromium levels in patients with
Aims. We sought to determine whether cobalt-chromium alloy (CoCr) femoral
stem tapers (trunnions) wear more than titanium (Ti) alloy stem
tapers (trunnions) when used in a
Aims. Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head
taper junction in two groups of retrieved hips; those with mixed
manufacturers (MM) and those from the same manufacturer (SM). . Materials and Methods. We identified 151 retrieved hips with large-diameter cobalt-chromium
heads; 51 of two designs that had been paired with stems from different
manufacturers (MM) and 100 of seven designs paired with stems from
the same manufacturer (SM). We determined the severity of corrosion
with the Goldberg corrosion score and the volume of material loss
at the head/stem junction. We used multivariable statistical analysis
to determine if there was a significant difference between the two
groups. . Results. We found no significant difference in the corrosion scores of
the two groups. The median rate of material loss at the head/stem
junction for the MM and SM groups were 0.39 mm. 3. /year
(0.00 to 4.73) and 0.46 mm. 3. /year (0.00 to 6.71) respectively;
this difference was not significant after controlling for confounding
factors (p = 0.06). . Conclusion. The use of stems with heads of another manufacturer does not
appear to affect the amount of metal lost from the surfaces between
these two components at total hip arthroplasty. Other surgical,
implant and patient factors should be considered when determining
the mechanisms of failure of
Iliopsoas impingement occurs in 4% to 30% of patients after undergoing total hip arthroplasty (THA). Despite a relatively high incidence, there are few attempts at modelling impingement between the iliopsoas and acetabular component, and no attempts at modelling this in a representative cohort of subjects. The purpose of this study was to develop a novel computational model for quantifying the impingement between the iliopsoas and acetabular component and validate its utility in a case-controlled investigation. This was a retrospective cohort study of patients who underwent THA surgery that included 23 symptomatic patients diagnosed with iliopsoas tendonitis, and 23 patients not diagnosed with iliopsoas tendonitis. All patients received postoperative CT imaging, postoperative standing radiography, and had minimum six months’ follow-up. 3D models of each patient’s prosthetic and bony anatomy were generated, landmarked, and simulated in a novel iliopsoas impingement detection model in supine and standing pelvic positions. Logistic regression models were implemented to determine if the probability of pain could be significantly predicted. Receiver operating characteristic curves were generated to determine the model’s sensitivity, specificity, and area under the curve (AUC).Aims
Methods
Aims. This study reports the mid-term results of total hip arthroplasty (THA) performed using a monoblock acetabular component with a large-diameter head (LDH) ceramic-on-ceramic (CoC) bearing. Patients and Methods. Of the 276 hips (246 patients) included in this study, 264 (96%) were reviewed at a mean of 67 months (48 to 79) postoperatively. Procedures were performed with a mini posterior approach. Clinical and radiological outcomes were recorded at regular intervals. A noise assessment questionnaire was completed at last follow-up. Results. There were four re-operations (1%) including one early revision for insufficient primary fixation (0.4%). No hip dislocation was reported. The mean University of California, Los Angeles (UCLA) activity score, 12-Item Short-Form Health Survey (SF-12) Mental Component Summary (MCS) score, SF-12 Physical Component Summary (PCS) score, Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) score, and Forgotten Joint Score (FJS) were 6.6 (2 to 10), 52.8 (25.5 to 65.7), 53.0 (27.2 to 66.5), 7.7 (0 to 63), and 88.5 (23 to 100), respectively. No signs of loosening or osteolysis were observed on radiological review. The incidence of squeaking was 23% (n = 51/225). Squeaking was significantly associated with
This study aims to assess the relationship between history of pseudotumour formation secondary to metal-on-metal (MoM) implants and periprosthetic joint infection (PJI) rate, as well as establish ESR and CRP thresholds that are suggestive of infection in these patients. We hypothesized that patients with a pseudotumour were at increased risk of infection. A total of 1,171 total hip arthroplasty (THA) patients with MoM articulations from August 2000 to March 2014 were retrospectively identified. Of those, 328 patients underwent metal artefact reduction sequence MRI and had minimum two years’ clinical follow-up, and met our inclusion criteria. Data collected included demographic details, surgical indication, laterality, implants used, history of pseudotumour, and their corresponding preoperative ESR (mm/hr) and CRP (mg/dl) levels. Multivariate logistic regression modelling was used to evaluate PJI and history of pseudotumour, and receiver operating characteristic curves were created to assess the diagnostic capabilities of ESR and CRP to determine the presence of infection in patients undergoing revision surgery.Aims
Methods
The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit. Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis.Aims
Methods
Resurfacing arthroplasty of the hip is being performed more frequently in the United Kingdom. The majority of these patients are younger than 55 years of age, and in this group the key benefits include conservation of femoral bone stock and the potential reduction in the rate of dislocation afforded by the larger resurfacing head. Early aseptic loosening is well recognised in patients younger than 55 years of age, and proponents of resurfacing believe that the improved wear characteristics of the metal-on-metal bearing may improve the long-term survival of this implant. There has been some concern, however, that resurfacing may not be conservative of acetabular bone. We compared a series of 33 consecutive patients who had a hybrid total hip arthroplasty with an uncemented acetabular component and a cemented femoral implant, with 35 patients undergoing a Birmingham hip resurfacing arthroplasty. We compared the diameter of the implanted acetabulum in both groups and, because they were not directly comparable, we corrected for patient size by measuring the diameter of the contralateral femoral head. The data were analysed using unpaired t-tests and analysis of covariance. There was a significantly larger acetabulum in the Birmingham arthroplasty group (mean diameter 56.6 mm vs 52.0 mm; p <
0.001). However, this group had a significantly
Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up. We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.Aims
Methods
Radiostereometric analysis (RSA) is the most accurate radiological method to measure in vivo wear of highly cross-linked polyethylene (XLPE) acetabular components. We have previously reported very low wear rates for a sequentially irradiated and annealed X3 XLPE liner (Stryker Orthopaedics, USA) when used in conjunction with a 32 mm femoral heads at ten-year follow-up. Only two studies have reported the long-term wear rate of X3 liners used in conjunction with larger heads using plain radiographs which have poor sensitivity. The aim of this study was to measure the ten-year wear of thin X3 XLPE liners against larger 36 or 40 mm articulations with RSA. We prospectively reviewed 19 patients who underwent primary cementless THA with the XLPE acetabular liner (X3) and a 36 or 40 mm femoral head with a resultant liner thickness of at least 5.8 mm. RSA radiographs at one week, six months, and one, two, five, and ten years postoperatively and femoral head penetration within the acetabular component were measured with UmRSA software. Of the initial 19 patients, 12 were available at the ten-year time point.Aims
Methods
This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.Aims
Methods
The aim of this study was to evaluate the incidence of liner malseating in two commonly used dual-mobility (DM) designs. Secondary aims included determining the risk of dislocation, survival, and clinical outcomes. We retrospectively identified 256 primary total hip arthroplasties (THAs) that included a DM component (144 Stryker MDM and 112 Zimmer-Biomet G7) in 233 patients, performed between January 2012 and December 2019. Postoperative radiographs were reviewed independently for malseating of the liner by five reviewers. The mean age of the patients at the time of THA was 66 years (18 to 93), 166 (65%) were female, and the mean BMI was 30 kg/m2 (17 to 57). The mean follow-up was 3.5 years (2.0 to 9.2).Aims
Methods
Total hip arthroplasty (THA) is a very successful and cost-effective operation, yet debate continues about the optimum fixation philosophy in different age groups. The concept of the 'cementless paradox' and the UK 'Getting it Right First Time' initiative encourage increased use of cemented fixation due to purported lower revision rates, especially in elderly patients, and decreased cost. In a high-volume, tertiary referral centre, we identified 10,112 THAs from a prospectively collected database, including 1,699 cemented THAs, 5,782 hybrid THAs, and 2,631 cementless THAs. The endpoint was revision for any reason. Secondary analysis included examination of implant survivorship in patients aged over 70 years, over 75 years, and over 80 years at primary THA.Aims
Methods