The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1731 - 1735
1 Nov 2021
Iobst CA Frost MW Rölfing JD Rahbek O Bafor A Duncan M Kold S

Aims. Limb-lengthening nails have largely replaced external fixation in limb-lengthening and reconstructive surgery. However, the adverse events and high prevalence of radiological changes recently noted with the STRYDE lengthening nail have raised concerns about the use of internal lengthening nails. The aim of this study was to compare the prevalence of radiological bone abnormalities between STRYDE, PRECICE, and FITBONE nails prior to nail removal. Methods. This was a retrospective case series from three centres. Patients were included if they had either of the three limb-lengthening nails (STYDE, PRECICE, or FITBONE) removed. Standard orthogonal radiographs immediately prior to nail removal were examined for bone abnormalities at the junction of the telescoping nail parts. Results. In total, 306 patients (168 male, 138 female) had 366 limb-lengthening nails removed. The mean time from nail insertion to radiological evaluation was 434 days (36 to 3,015). Overall, 77% of STRYDE nails (20/26) had bone abnormalities at the interface compared with only 2% of FITBONE (4/242) and 1% of PRECICE nails (1/98; p < 0.001). Focal osteolysis in conjunction with periosteal reaction at the telescoping interface was only observed in STRYDE nails. Conclusion. Bone abnormalities at the interface of telescoping nail parts were seen in the majority of STRYDE nails, but only very rarely with FITBONE or PRECICE nails. We conclude that the low prevalence of radiological changes at the junctional interface of 242 FITBONE and 98 PRECICE nails at the time of nail removal does not warrant clinical concerns. Cite this article: Bone Joint J 2021;103-B(11):1731–1735


Bone & Joint Open
Vol. 2, Issue 8 | Pages 599 - 610
1 Aug 2021
Hothi H Bergiers S Henckel J Iliadis AD Goodier WD Wright J Skinner J Calder P Hart AJ

Aims. The aim of this study was to present the first retrieval analysis findings of PRECICE STRYDE intermedullary nails removed from patients, providing useful information in the post-market surveillance of these recently introduced devices. Methods. We collected ten nails removed from six patients, together with patient clinical data and plain radiograph imaging. We performed macro- and microscopic analysis of all surfaces and graded the presence of corrosion using validated semiquantitative scoring methods. We determined the elemental composition of surface debris using energy dispersive x-ray spectroscopy (EDS) and used metrology analysis to characterize the surface adjacent to the extendable junctions. Results. All nails were removed at the end of treatment, having achieved their intended lengthening (20 mm to 65 mm) and after regenerate consolidation. All nails had evidence of corrosion localized to the screw holes and the extendable junctions; corrosion was graded as moderate at the junction of one nail and severe at the junctions of five nails. EDS analysis showed surface deposits to be chromium rich. Plain radiographs showed cortical thickening and osteolysis around the junction of six nails, corresponding to the same nails with moderate – severe junction corrosion. Conclusion. We found, in fully united bones, evidence of cortical thickening and osteolysis that appeared to be associated with corrosion at the extendable junction; when corrosion was present, cortical thickening was adjacent to this junction. Further work, with greater numbers of retrievals, is required to fully understand this association between corrosion and bony changes, and the influencing surgeon, implant, and patient factors involved. Cite this article: Bone Jt Open 2021;2(8):599–610


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 88 - 96
1 Jan 2023
Vogt B Rupp C Gosheger G Eveslage M Laufer A Toporowski G Roedl R Frommer A

Aims. Distraction osteogenesis with intramedullary lengthening devices has undergone rapid development in the past decade with implant enhancement. In this first single-centre matched-pair analysis we focus on the comparison of treatment with the PRECICE and STRYDE intramedullary lengthening devices and aim to clarify any clinical and radiological differences. Methods. A single-centre 2:1 matched-pair retrospective analysis of 42 patients treated with the STRYDE and 82 patients treated with the PRECICE nail between May 2013 and November 2020 was conducted. Clinical and lengthening parameters were compared while focusing radiological assessment on osseous alterations related to the nail’s telescopic junction and locking bolts at four different stages. Results. Osteolysis next to the telescopic junction was observed in 31/48 segments (65%) lengthened with the STRYDE nail before implant removal compared to 1/91 segment (1%) in the PRECICE cohort. In the STRYDE cohort, osteolysis initially increased, but decreased or resolved in almost all lengthened segments (86%) after implant removal. Implant failure was observed in 9/48 STRYDE (19%) and in 8/92 PRECICE nails (9%). Breakage of the distal locking bolts was found in 5/48 STRYDE nails (10%) compared to none in the PRECICE cohort. Treatment-associated pain was generally recorded as mild and found in 30/48 patients (63%) and 39/92 (42%) in the STRYDE and PRECICE cohorts, respectively. Temporary range of motion (ROM) limitations under distraction were registered in 17/48 (35%) segments treated with the STRYDE and 35/92 segments (38%) treated with the PRECICE nail. Conclusion. Osteolysis and periosteal reaction, implant breakage, and pain during lengthening and consolidation is more likely in patients treated with the STRYDE nail compared to the PRECICE nail. Temporary ROM limitations during lengthening occurred independent of the applied device. Implant-related osseous alterations seem to remodel after implant removal. Cite this article: Bone Joint J 2023;105-B(1):88–96


The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1168 - 1172
1 Jun 2021
Iliadis AD Wright J Stoddart MT Goodier WD Calder P

Aims. The STRYDE nail is an evolution of the PRECICE Intramedullary Limb Lengthening System, with unique features regarding its composition. It is designed for load bearing throughout treatment in order to improve patient experience and outcomes and allow for simultaneous bilateral lower limb lengthening. The literature published to date is limited regarding outcomes and potential problems. We report on our early experience and raise awareness for the potential of adverse effects from this device. Methods. This is a retrospective review of prospective data collected on all patients treated in our institution using this implant. We report the demographics, nail accuracy, reliability, consolidation index, and cases where concerning clinical and radiological findings were encountered. There were 14 STRYDE nails implanted in nine patients (three male and six female) between June 2019 and September 2020. Mean age at surgery was 33 years (14 to 65). Five patients underwent bilateral lengthening (two femoral and three tibial) and four patients unilateral femoral lengthening for multiple aetiologies. Results. At the time of reporting, eight patients (13 implants) had completed lengthening. Osteolysis and periosteal reaction at the junction of the telescopic nail was evident in nine implants. Five patients experienced localized pain and swelling. Macroscopic appearances following retrieval were consistent with corrosion at the telescopic junction. Tissue histology was consistent with effects of focal metallic wear debris. Conclusion. From our early experience with this implant we have found the process of lengthening to be accurate and reliable with good regenerate formation and consolidation. Proposed advantages of early load bearing and the ability for bilateral lengthening are promising. We have, however, encountered concerning clinical and radiological findings in several patients. We have elected to discontinue its use to allow further investigation into the retrieved implants and patient outcomes from users internationally. Cite this article: Bone Joint J 2021;103-B(6):1168–1172


The Bone & Joint Journal
Vol. 106-B, Issue 3 | Pages 293 - 302
1 Mar 2024
Vogt B Lueckingsmeier M Gosheger G Laufer A Toporowski G Antfang C Roedl R Frommer A

Aims. As an alternative to external fixators, intramedullary lengthening nails (ILNs) can be employed for distraction osteogenesis. While previous studies have demonstrated that typical complications of external devices, such as soft-tissue tethering, and pin site infection can be avoided with ILNs, there is a lack of studies that exclusively investigated tibial distraction osteogenesis with motorized ILNs inserted via an antegrade approach. Methods. A total of 58 patients (median age 17 years (interquartile range (IQR) 15 to 21)) treated by unilateral tibial distraction osteogenesis for a median leg length discrepancy of 41 mm (IQR 34 to 53), and nine patients with disproportionate short stature treated by bilateral simultaneous tibial distraction osteogenesis, with magnetically controlled motorized ILNs inserted via an antegrade approach, were retrospectively analyzed. The median follow-up was 37 months (IQR 30 to 51). Outcome measurements were accuracy, precision, reliability, bone healing, complications, and patient-reported outcome assessed by the Limb Deformity-Scoliosis Research Society Score (LD-SRS-30). Results. A median tibial distraction of 44 mm (IQR 31 to 49) was achieved with a mean distraction index of 0.5 mm/day (standard deviation 0.13) and median consolidation index of 41.2 days/cm (IQR 34 to 51). Accuracy, precision, and reliability were 91%, 92%, and 97%, respectively. New temporary range of motion limitations occurred in 51% of segments (34/67). Distraction-related equinus deformity treated by Achilles tendon lengthening was the most common major complication recorded in 16% of segments (11/67). In 95% of patients (55/58) the distraction goal was achieved with 42% unplanned additional interventions per segment (28/67). The median postoperative LD-SRS-30 score was 4.0 (IQR 3.6 to 4.3). Conclusion. Tibial distraction osteogenesis using motorized ILNs inserted via an antegrade approach appears to be a reliable and precise procedure. Temporary joint stiffness of the knee or ankle should be expected in up to every second patient. A high rate and wide range of complications of variable severity should be anticipated. Cite this article: Bone Joint J 2024;106-B(3):293–302


The Bone & Joint Journal
Vol. 101-B, Issue 2 | Pages 178 - 188
1 Feb 2019
Chaudhary MM Lakhani PH

Aims. Double-level lengthening, bone transport, and bifocal compression-distraction are commonly undertaken using Ilizarov or other fixators. We performed double-level fixator-assisted nailing, mainly for the correction of deformity and lengthening in the same segment, using a straight intramedullary nail to reduce the time in a fixator. Patients and Methods. A total of 23 patients underwent this surgery, involving 27 segments (23 femora and four tibiae), over a period of ten years. The most common indication was polio in ten segments and rickets in eight; 20 nails were inserted retrograde and seven antegrade. A total of 15 lengthenings were performed in 11 femora and four tibiae, and 12 double-level corrections of deformity without lengthening were performed in the femur. The mean follow-up was 4.9 years (1.1 to 11.4). Four patients with polio had tibial lengthening with arthrodesis of the ankle. We compared the length of time in a fixator and the external fixation index (EFI) with a control group of 27 patients (27 segments) who had double-level procedures with external fixation. The groups were matched for the gain in length, age, and level of difficulty score. Results. The mean gain in length was statistically similar in the two groups: 3.9 cm (1.5 to 9.0) in the study group and 4.2 cm (3.4 to 5.0) in the control group (p = 0.350). The mean time in a fixator was significantly less in the study group compared with the control group: 8.6 weeks (2.0 to 22.8) versus 30.2 weeks (25.0 to 35.4; p < 0.001). The mean EFI was significantly lower in the study group compared with the control group: 17.7 days/cm (10.6 to 35.6) versus 73.4 days/cm (44.5 to 102.3; p < 0.001). The ASAMI (Association for the Study and Application of the Method of Ilizarov) bone score was excellent in 22, good in four, and fair in one. The ASAMI functional score was excellent in 20 and good in seven. There were no infections, superficial or deep. Conclusion. Double-level osteotomies or two procedures using a custom-made straight nail and external fixation can be used to correct deformities or to treat nonunion or malunion and may be combined with arthrodesis of the ankle with lengthening. It is a reasonably safe procedure that allows accurate and cost-effective treatment with a relatively short time in a fixator


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 146 - 152
1 Jan 2010
Bilen FE Kocaoglu M Eralp L Balci HI

We report the results of using a combination of fixator-assisted nailing with lengthening over an intramedullary nail in patients with tibial deformity and shortening. Between 1997 and 2007, 13 tibiae in nine patients with a mean age of 25.4 years (17 to 34) were treated with a unilateral external fixator for acute correction of deformity, followed by lengthening over an intramedullary nail with a circular external fixator applied at the same operating session. At the end of the distraction period locking screws were inserted through the intramedullary nail and the external fixator was removed. The mean amount of lengthening was 5.9 cm (2 to 8). The mean time of external fixation was 90 days (38 to 265). The mean external fixation index was 15.8 days/cm (8.9 to 33.1) and the mean bone healing index was 38 days/cm (30 to 60). One patient developed an equinus deformity which responded to stretching and bracing. Another developed a drop foot due to a compartment syndrome, which was treated by fasciotomy. It recovered in three months. Two patients required bone grafting for poor callus formation. We conclude that the combination of fixator-assisted nailing with lengthening over an intramedullary nail can reduce the overall external fixation time and prevent fractures and deformity of the regenerated bone


Bone & Joint Research
Vol. 10, Issue 7 | Pages 425 - 436
16 Jul 2021
Frommer A Roedl R Gosheger G Hasselmann J Fuest C Toporowski G Laufer A Tretow H Schulze M Vogt B

Aims. This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported. Methods. This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors’ institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed. Results. Adjacent to the nail’s telescoping junction, osteolytic changes were observed in bi-planar radiographs of 20/57 segments (35%) after a mean of 9.5 months (95% confidence interval 7.2 to 11.9) after surgery. A total of 8/20 patients with osseous alterations (40%) reported rest and ambulation pain of the lengthened segment during consolidation. So far, 24 Stryde nails were retrieved and in 20 (83%) macroscopic corrosion was observed at the nail’s telescoping junction. Before implant removal 11/20 radiographs (55%) of lengthened segments with these 20 nails revealed osteolysis. Implant retrieval analysis by means of SEM showed pitting and crevice corrosion. EDX detected chromium as the main metallic element of corrosion. Conclusion. Patients are exposed to the risk of implant-related osteolysis of unclear short- and long-term clinical consequences. The authors advocate in favour of an early implant removal after osseous consolidation. Cite this article: Bone Joint Res 2021;10(7):425–436


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1168 - 1176
1 Sep 2019
Calder PR McKay JE Timms AJ Roskrow T Fugazzotto S Edel P Goodier WD

Aims. The Precice intramedullary limb-lengthening system has demonstrated significant benefits over external fixation lengthening methods, leading to a paradigm shift in limb lengthening. This study compares outcomes following antegrade and retrograde femoral lengthening in both adolescent and adult patients. Patients and Methods. A retrospective review of prospectively collected data was undertaken of a consecutive series of 107 femoral lengthening operations in 92 patients. In total, 73 antegrade nails and 34 retrograde nails were inserted. Outcome was assessed by the regenerate healing index (HI), hip and knee range of movement (ROM), and the presence of any complications. Results. The mean lengthening was 4.65 cm (1.5 to 8) in the antegrade group and 4.64 cm (1.6 to 8) in the retrograde group. Of the 107 lengthenings, 100 had sufficient datapoints to calculate the mean HI. This was 31.6 days/cm (15 to 108). There was a trend toward a lower (better) HI with an antegrade nail and better outcomes in adolescent patients, but these were not statistically significant. Hip and knee ROM was maintained and/or improved following commencement of femoral lengthening in 44 patients (60%) of antegrade nails and 13 patients (38%) of retrograde nails. In female patients, loss of movement occurred both earlier and following less total length achieved. Minor implant complications included locking bolt migration and in one patient deformity of the nail, but no implant failed to lengthen and there were no deep infections. Three patients had delayed union, five patients required surgical intervention for joint contracture. Conclusion. This study confirms excellent results in femoral lengthening with antegrade and retrograde Precice nails. There is a trend for better healing and less restriction in hip and knee movement following antegrade nails. There are clinical scenarios, that mandate the use of a retrograde nail. However, when these are not present, we recommend the use of antegrade nailing. Cite this article: Bone Joint J 2019;101-B:1168–1176


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 962 - 967
1 Jul 2009
Aston WJS Calder PR Baker D Hartley J Hill RA

We present a retrospective review of a single-surgeon series of 30 consecutive lengthenings in 27 patients with congenital short femur using the Ilizarov technique performed between 1994 and 2005. The mean increase in length was 5.8 cm/18.65% (3.3 to 10.4, 9.7% to 48.8%), with a mean time in the frame of 223 days (75 to 363). By changing from a distal to a proximal osteotomy for lengthening, the mean range of knee movement was significantly increased from 98.1° to 124.2° (p = 0.041) and there was a trend towards a reduced requirement for quadricepsplasty, although this was not statistically significant (p = 0.07). The overall incidence of regenerate deformation or fracture requiring open reduction and internal fixation was similar in the distal and proximal osteotomy groups (56.7% and 53.8%, respectively). However, in the proximal osteotomy group, pre-placement of a Rush nail reduced this rate from 100% without a nail to 0% with a nail (p < 0.001). When comparing a distal osteotomy with a proximal one over a Rush nail for lengthening, there was a significant decrease in fracture rate from 58.8% to 0% (p = 0.043). We recommend that in this group of patients lengthening of the femur with an Ilizarov construct be carried out through a proximal osteotomy over a Rush nail. Lengthening should also be limited to a maximum of 6 cm during one treatment, or 20% of the original length of the femur, in order to reduce the risk of complications


Bone & Joint Open
Vol. 4, Issue 8 | Pages 643 - 651
24 Aug 2023
Langit MB Tay KS Al-Omar HK Barlow G Bates J Chuo CB Muir R Sharma H

Aims

The standard of wide tumour-like resection for chronic osteomyelitis (COM) has been challenged recently by adequate debridement. This paper reviews the evolution of surgical debridement for long bone COM, and presents the outcome of adequate debridement in a tertiary bone infection unit.

Methods

We analyzed the retrospective record review from 2014 to 2020 of patients with long bone COM. All were managed by multidisciplinary infection team (MDT) protocol. Adequate debridement was employed for all cases, and no case of wide resection was included.


Bone & Joint Open
Vol. 3, Issue 11 | Pages 907 - 912
23 Nov 2022
Hurley RJ McCabe FJ Turley L Maguire D Lucey J Hurson CJ

Aims

The use of fluoroscopy in orthopaedic surgery creates risk of radiation exposure to surgeons. Appropriate personal protective equipment (PPE) can help mitigate this. The primary aim of this study was to assess if current radiation protection in orthopaedic trauma is safe. The secondary aims were to describe normative data of radiation exposure during common orthopaedic procedures, evaluate ways to improve any deficits in protection, and validate the use of electronic personal dosimeters (EPDs) in assessing radiation dose in orthopaedic surgery.

Methods

Radiation exposure to surgeons during common orthopaedic trauma operations was prospectively assessed using EPDs and thermoluminescent dosimeters (TLDs). Normative data for each operation type were calculated and compared to recommended guidelines.


Bone & Joint 360
Vol. 12, Issue 3 | Pages 43 - 43
1 Jun 2023
Das A

This edition of Cochrane Corner looks at some of the work published by the Cochrane Collaboration, covering interventions for treating distal femur fractures in adults; ultrasound and shockwave therapy for acute fractures in adults; and local corticosteroid injection versus placebo for carpal tunnel syndrome.


Bone & Joint Open
Vol. 3, Issue 5 | Pages 359 - 366
1 May 2022
Sadekar V Watts AT Moulder E Souroullas P Hadland Y Barron E Muir R Sharma HK

Aims

The timing of when to remove a circular frame is crucial; early removal results in refracture or deformity, while late removal increases the patient morbidity and delay in return to work. This study was designed to assess the effectiveness of a staged reloading protocol. We report the incidence of mechanical failure following both single-stage and two stage reloading protocols and analyze the associated risk factors.

Methods

We identified consecutive patients from our departmental database. Both trauma and elective cases were included, of all ages, frame types, and pathologies who underwent circular frame treatment. Our protocol is either a single-stage or two-stage process implemented by defunctioning the frame, in order to progressively increase the weightbearing load through the bone, and promote full loading prior to frame removal. Before progression, through the process we monitor patients for any increase in pain and assess radiographs for deformity or refracture.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 938 - 942
1 Jul 2006
Singh S Lahiri A Iqbal M

Limb lengthening by callus distraction and external fixation has a high rate of complications. We describe our experience using an intramedullary nail (Fitbone) which contains a motorised and programmable sliding mechanism for limb lengthening and bone transport. Between 2001 and 2004 we lengthened 13 femora and 11 tibiae in ten patients (seven men and three women) with a mean age of 32 years (21 to 47) using this nail. The indications for operation were short stature in six patients and developmental or acquired disorders in the rest. The mean lengthening achieved was 40 mm (27 to 60). The mean length of stay in hospital was seven days (5 to 9). The mean healing index was 35 days/cm (18.8 to 70.9). There were no cases of implant-related infection or malunion


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 955 - 961
1 Jul 2009
Simpson AHWR Shalaby H Keenan G

Femoral lengthening using the Intramedullary Skeletal Kinetic Distractor is a new technique. However, with intramedullary distraction the surgeon has less control over the lengthening process. Therefore, 33 femora lengthened with this device were assessed to evaluate the effect of operative variables under the surgeon’s control on the course of lengthening. The desired lengthening was achieved in 32 of 33 limbs. Problems encountered included difficulty in achieving length in eight femora (24%) and uncontrolled lengthening in seven (21%). Uncontrolled lengthening was more likely if the osteotomy was placed with less than 80 mm of the thick portion of the nail in the distal fragment (p = 0.052), and a failure to lengthen was more likely if there was over 125 mm in the distal fragment (p = 0.008). The latter problem was reduced with over-reaming by 2.5 mm to 3 mm. Previous intramedullary nailing also predisposed to uncontrolled lengthening (p = 0.042), and these patients required less reaming. Using the Intramedullary Skeletal Kinetic Distractor, good outcomes were obtained; problems were minimised by optimising the position of the osteotomy and the amount of over-reaming performed


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1394 - 1399
1 Oct 2009
Oh C Song H Kim J Choi J Min W Park B

Ten patients, who were unsuitable for limb lengthening over an intramedullary nail, underwent lengthening with a submuscular locking plate. Their mean age at operation was 18.5 years (11 to 40). After fixing a locking plate submuscularly on the proximal segment, an external fixator was applied to lengthen the bone after corticotomy. Lengthening was at 1 mm/day and on reaching the target length, three or four screws were placed in the plate in the distal segment and the external fixator was removed. All patients achieved the pre-operative target length at a mean of 4.0 cm (3.2 to 5.5). The mean duration of external fixation was 61.6 days (45 to 113) and the mean external fixation index was 15.1 days/cm (13.2 to 20.5), which was less than one-third of the mean healing index (48 days/cm (41.3 to 55). There were only minor complications. Lengthening with a submuscular locking plate can successfully permit early removal of the fixator with fewer complications and is a useful alternative in children or when nailing is difficult


Bone & Joint Open
Vol. 2, Issue 9 | Pages 705 - 709
1 Sep 2021
Wright J Timms A Fugazzotto S Goodier D Calder P

Aims

Patients undergoing limb reconstruction surgery often face a challenging and lengthy process to complete their treatment journey. The majority of existing outcome measures do not adequately capture the patient-reported outcomes relevant to this patient group in a single measure. Following a previous systematic review, the Stanmore Limb Reconstruction Score (SLRS) was designed with the intent to address this need for an effective instrument to measure patient-reported outcomes in limb reconstruction patients. We aim to assess the face validity of this score in a pilot study.

Methods

The SLRS was designed following structured interviews with several groups including patients who have undergone limb reconstruction surgery, limb reconstruction surgeons, specialist nurses, and physiotherapists. This has subsequently undergone further adjustment for language and clarity. The score was then trialled on ten patients who had undergone limb reconstruction surgery, with subsequent structured questioning to understand the perceived suitability of the score.


Bone & Joint Open
Vol. 1, Issue 6 | Pages 229 - 235
9 Jun 2020
Lazizi M Marusza CJ Sexton SA Middleton RG

Aims

Elective surgery has been severely curtailed as a result of the COVID-19 pandemic. There is little evidence to guide surgeons in assessing what processes should be put in place to restart elective surgery safely in a time of endemic COVID-19 in the community.

Methods

We used data from a stand-alone hospital admitting and operating on 91 trauma patients. All patients were screened on admission and 100% of patients have been followed-up after discharge to assess outcome.


Bone & Joint Open
Vol. 1, Issue 10 | Pages 645 - 652
19 Oct 2020
Sheridan GA Hughes AJ Quinlan JF Sheehan E O'Byrne JM

Aims

We aim to objectively assess the impact of COVID-19 on mean total operative cases for all indicative procedures (as outlined by the Joint Committee on Surgical Training (JCST)) experienced by orthopaedic trainees in the deanery of the Republic of Ireland. Subjective experiences were reported for each trainee using questionnaires.

Methods

During the first four weeks of the nationwide lockdown due to COVID-19, the objective impact of the pandemic on each trainee’s surgical caseload exposure was assessed using data from individual trainee logbook profiles in the deanery of the Republic of Ireland. Independent predictor variables included the trainee grade (ST 3 to 8), the individual trainee, the unit that the logbook was reported from, and the year in which the logbook was recorded. We used the analysis of variance (ANOVA) test to assess for any statistically significant predictor variables. The subjective experience of each trainee was captured using an electronic questionnaire.