Advertisement for orthosearch.org.uk
Results 1 - 20 of 31
Results per page:
Bone & Joint Open
Vol. 5, Issue 1 | Pages 46 - 52
19 Jan 2024
Assink N ten Duis K de Vries JPM Witjes MJH Kraeima J Doornberg JN IJpma FFA

Aims. Proper preoperative planning benefits fracture reduction, fixation, and stability in tibial plateau fracture surgery. We developed and clinically implemented a novel workflow for 3D surgical planning including patient-specific drilling guides in tibial plateau fracture surgery. Methods. A prospective feasibility study was performed in which consecutive tibial plateau fracture patients were treated with 3D surgical planning, including patient-specific drilling guides applied to standard off-the-shelf plates. A postoperative CT scan was obtained to assess whether the screw directions, screw lengths, and plate position were performed according the preoperative planning. Quality of the fracture reduction was assessed by measuring residual intra-articular incongruence (maximum gap and step-off) and compared to a historical matched control group. Results. A total of 15 patients were treated with 3D surgical planning in which 83 screws were placed by using drilling guides. The median deviation of the achieved screw trajectory from the planned trajectory was 3.4° (interquartile range (IQR) 2.5 to 5.4) and the difference in entry points (i.e. plate position) was 3.0 mm (IQR 2.0 to 5.5) compared to the 3D preoperative planning. The length of 72 screws (86.7%) were according to the planning. Compared to the historical cohort, 3D-guided surgery showed an improved surgical reduction in terms of median gap (3.1 vs 4.7 mm; p = 0.126) and step-off (2.9 vs 4.0 mm; p = 0.026). Conclusion. The use of 3D surgical planning including drilling guides was feasible, and facilitated accurate screw directions, screw lengths, and plate positioning. Moreover, the personalized approach improved fracture reduction as compared to a historical cohort. Cite this article: Bone Jt Open 2024;5(1):46–52


Aims. This study aimed to analyze the accuracy and errors associated with 3D-printed, patient-specific resection guides (3DP-PSRGs) used for bone tumour resection. Methods. We retrospectively reviewed 29 bone tumour resections that used 3DP-PSRGs based on 3D CT and 3D MRI. We evaluated the resection amount errors and resection margin errors relative to the preoperative plans. Guide-fitting errors and guide distortion were evaluated intraoperatively and one month postoperatively, respectively. We categorized each of these error types into three grades (grade 1, < 1 mm; grade 2, 1 to 3 mm; and grade 3, > 3 mm) to evaluate the overall accuracy. Results. The maximum resection amount error was 2 mm. Out of 29 resection amount errors, 15 (51.7%) were grade 1 errors and 14 (48.3%) were grade 2 errors. Complex resections were associated with higher-grade resection amount errors (p < 0.001). The actual resection margins correlated significantly with the planned margins; however, there were some discrepancies. The maximum guide-fitting error was 3 mm. There were 22 (75.9%), five (17.2%), and two (6.9%) grade 1, 2, and 3 guide-fitting errors, respectively. There was no significant association between complex resection and fitting error grades. The guide distortion after one month in all patients was rated as grade 1. Conclusion. In terms of the accurate resection amount according to the preoperative planning, 3DP-PSRGs can be a viable option for bone tumour resection. However, 3DP-PSRG use may be associated with resection margin length discrepancies relative to the planned margins. Such discrepancies should be considered when determining surgical margins. Therefore, a thorough evaluation of the preoperative imaging and surgical planning is still required, even if 3DP-PSRGs are to be used. Cite this article: Bone Joint J 2023;105-B(2):190–197


The Bone & Joint Journal
Vol. 98-B, Issue 8 | Pages 1080 - 1085
1 Aug 2016
Gauci MO Boileau P Baba M Chaoui J Walch G

Aims. Patient-specific glenoid guides (PSGs) claim an improvement in accuracy and reproducibility of the positioning of components in total shoulder arthroplasty (TSA). The results have not yet been confirmed in a prospective clinical trial. Our aim was to assess whether the use of PSGs in patients with osteoarthritis of the shoulder would allow accurate and reliable implantation of the glenoid component. Patients and Methods. A total of 17 patients (three men and 14 women) with a mean age of 71 years (53 to 81) awaiting TSA were enrolled in the study. Pre- and post-operative version and inclination of the glenoid were measured on CT scans, using 3D planning automatic software. During surgery, a congruent 3D-printed PSG was applied onto the glenoid surface, thus determining the entry point and orientation of the central guide wire used for reaming the glenoid and the introduction of the component. Manual segmentation was performed on post-operative CT scans to compare the planned and the actual position of the entry point (mm) and orientation of the component (°). Results. The mean error in the accuracy of the entry point was -0.1 mm (standard deviation (. sd. ) 1.4) in the horizontal plane, and 0.8 mm (. sd. 1.3) in the vertical plane. The mean error in the orientation of the glenoid component was 3.4° (. sd. 5.1°) for version and 1.8° (. sd. 5.3°) for inclination. Conclusion. Pre-operative planning with automatic software and the use of PSGs provides accurate and reproducible positioning and orientation of the glenoid component in anatomical TSA. Cite this article: Bone Joint J 2016;98-B:1080–5


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 354 - 359
1 Mar 2013
Chareancholvanich K Narkbunnam R Pornrattanamaneewong C

Patient-specific cutting guides (PSCGs) are designed to improve the accuracy of alignment of total knee replacement (TKR). We compared the accuracy of limb alignment and component positioning after TKR performed using PSCGs or conventional instrumentation. A total of 80 patients were randomised to undergo TKR with either of the different forms of instrumentation, and radiological outcomes and peri-operative factors such as operating time were assessed. No significant difference was observed between the groups in terms of tibiofemoral angle or femoral component alignment. Although the tibial component in the PSCGs group was measurably closer to neutral alignment than in the conventional group, the size of the difference was very small (89.8° (. sd. 1.2) vs 90.5° (. sd. 1.6); p = 0.030). This new technology slightly shortened the bone-cutting time by a mean of 3.6 minutes (p < 0.001) and the operating time by a mean 5.1 minutes (p = 0.019), without tangible differences in post-operative blood loss (p = 0.528) or need for blood transfusion (p = 0.789). This study demonstrated that both PSCGs and conventional instrumentation restore limb alignment and place the components with the similar accuracy. The minimal advantages of PSCGs in terms of consistency of alignment or operative time are unlikely to be clinically relevant. Cite this article: Bone Joint J 2013;95-B:354–9


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 939 - 944
1 Jul 2016
Boonen B Schotanus MGM Kerens B van der Weegen W Hoekstra HJ Kort NP

Aims

We wished to compare the clinical outcome, as assessed by questionnaires and the rate of complications, in total knee arthroplasty (TKA) undertaken with patient-matched positioning guides (PMPGs) or conventional instruments.

Patients and Methods

A total of 180 patients (74 men, 106 women; mean age 67 years) were included in a multicentre, adequately powered, double-blind, randomised controlled trial. The mean follow-up was 44 months (24 to 57).


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 786 - 792
1 Jun 2016
Schotanus MGM Sollie R van Haaren EH Hendrickx RPM Jansen EJP Kort NP

Aims

This prospective randomised controlled trial was designed to evaluate the outcome of both the MRI- and CT-based patient-specific matched guides (PSG) from the same manufacturer.

Patients and Methods

A total of 137 knees in 137 patients (50 men, 87 women) were included, 67 in the MRI- and 70 in the CT-based PSG group. Their mean age was 68.4 years (47.0 to 88.9). Outcome was expressed as the biomechanical limb alignment (centre hip-knee-ankle: HKA-axis) achieved post-operatively, the position of the individual components within 3° of the pre-operatively planned alignment, correct planned implant size and operative data (e.g. operating time and blood loss).


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1342 - 1346
1 Oct 2016
Spencer-Gardner L Pierrepont J Topham M Baré J McMahon S Shimmin AJ

Aims. Accurate placement of the acetabular component during total hip arthroplasty (THA) is an important factor in the success of the procedure. However, the reported accuracy varies greatly and is dependent upon whether free hand or navigated techniques are used. The aim of this study was to assess the accuracy of an instrument system that incorporates 3D printed, patient-specific guides designed to optimise the placement of the acetabular component. Patients and Methods. A total of 100 consecutive patients were prospectively enrolled and the accuracy of placement of the acetabular component was measured using post-operative CT scans. Results. The mean absolute deviation from the planned inclination and anteversion was 3.9° (0.0° to 13.6°) and 3.6° (0.0° to 12.9°), respectively. In 91% of cases the planned target of +/-10° was achieved for both inclination and anteversion. Conclusion. Accurate placement of the acetabular component can be achieved using patient-specific guides and is superior to free hand techniques and comparable to navigated and robotic techniques. Cite this article: Bone Joint J 2016;98-B:1342–6


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 907 - 913
1 Jul 2014
Dossett HG Estrada NA Swartz GJ LeFevre GW Kwasman BG

We have previously reported the short-term radiological results of a randomised controlled trial comparing kinematically aligned total knee replacement (TKR) and mechanically aligned TKR, along with early pain and function scores. In this study we report the two-year clinical results from this trial. A total of 88 patients (88 knees) were randomly allocated to undergo either kinematically aligned TKR using patient-specific guides, or mechanically aligned TKR using conventional instruments. They were analysed on an intention-to-treat basis. The patients and the clinical evaluator were blinded to the method of alignment. At a minimum of two years, all outcomes were better for the kinematically aligned group, as determined by the mean Oxford knee score (40 (15 to 48) versus 33 (13 to 48); p = 0.005), the mean Western Ontario McMaster Universities Arthritis index (WOMAC) (15 (0 to 63) versus 26 (0 to 73); p = 0.005), mean combined Knee Society score (160 (93 to 200) versus 137 (64 to 200); p= 0.005) and mean flexion of 121° (100 to 150) versus 113° (80 to 130) (p = 0.002). The odds ratio of having a pain-free knee at two years with the kinematically aligned technique (Oxford and WOMAC pain scores) was 3.2 (p = 0.020) and 4.9 (p = 0.001), respectively, compared with the mechanically aligned technique. Patients in the kinematically aligned group walked a mean of 50 feet further in hospital prior to discharge compared with the mechanically aligned group (p = 0.044). In this study, the use of a kinematic alignment technique performed with patient-specific guides provided better pain relief and restored better function and range of movement than the mechanical alignment technique performed with conventional instruments. Cite this article: Bone Joint J 2014;96-B:907–13


Bone & Joint Open
Vol. 5, Issue 4 | Pages 317 - 323
18 Apr 2024
Zhu X Hu J Lin J Song G Xu H Lu J Tang Q Wang J

Aims

The aim of this study was to investigate the safety and efficacy of 3D-printed modular prostheses in patients who underwent joint-sparing limb salvage surgery (JSLSS) for malignant femoral diaphyseal bone tumours.

Methods

We retrospectively reviewed 17 patients (13 males and four females) with femoral diaphyseal tumours who underwent JSLSS in our hospital.


The Bone & Joint Journal
Vol. 106-B, Issue 11 | Pages 1284 - 1292
1 Nov 2024
Moroder P Poltaretskyi S Raiss P Denard PJ Werner BC Erickson BJ Griffin JW Metcalfe N Siegert P

Aims

The objective of this study was to compare simulated range of motion (ROM) for reverse total shoulder arthroplasty (rTSA) with and without adjustment for scapulothoracic orientation in a global reference system. We hypothesized that values for simulated ROM in preoperative planning software with and without adjustment for scapulothoracic orientation would be significantly different.

Methods

A statistical shape model of the entire humerus and scapula was fitted into ten shoulder CT scans randomly selected from 162 patients who underwent rTSA. Six shoulder surgeons independently planned a rTSA in each model using prototype development software with the ability to adjust for scapulothoracic orientation, the starting position of the humerus, as well as kinematic planes in a global reference system simulating previously described posture types A, B, and C. ROM with and without posture adjustment was calculated and compared in all movement planes.


Bone & Joint 360
Vol. 11, Issue 4 | Pages 29 - 32
1 Aug 2022


Aims

For rare cases when a tumour infiltrates into the hip joint, extra-articular resection is required to obtain a safe margin. Endoprosthetic reconstruction following tumour resection can effectively ensure local control and improve postoperative function. However, maximizing bone preservation without compromising surgical margin remains a challenge for surgeons due to the complexity of the procedure. The purpose of the current study was to report clinical outcomes of patients who underwent extra-articular resection of the hip joint using a custom-made osteotomy guide and 3D-printed endoprosthesis.

Methods

We reviewed 15 patients over a five-year period (January 2017 to December 2022) who had undergone extra-articular resection of the hip joint due to malignant tumour using a custom-made osteotomy guide and 3D-printed endoprosthesis. Each of the 15 patients had a single lesion, with six originating from the acetabulum side and nine from the proximal femur. All patients had their posterior column preserved according to the surgical plan.


The Bone & Joint Journal
Vol. 105-B, Issue 4 | Pages 465 - 465
15 Mar 2023


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 892 - 897
1 Sep 2024
Mancino F Fontalis A Kayani B Magan A Plastow R Haddad FS

Advanced 3D imaging and CT-based navigation have emerged as valuable tools to use in total knee arthroplasty (TKA), for both preoperative planning and the intraoperative execution of different philosophies of alignment. Preoperative planning using CT-based 3D imaging enables more accurate prediction of the size of components, enhancing surgical workflow and optimizing the precision of the positioning of components. Surgeons can assess alignment, osteophytes, and arthritic changes better. These scans provide improved insights into the patellofemoral joint and facilitate tibial sizing and the evaluation of implant-bone contact area in cementless TKA. Preoperative CT imaging is also required for the development of patient-specific instrumentation cutting guides, aiming to reduce intraoperative blood loss and improve the surgical technique in complex cases. Intraoperative CT-based navigation and haptic guidance facilitates precise execution of the preoperative plan, aiming for optimal positioning of the components and accurate alignment, as determined by the surgeon’s philosophy. It also helps reduce iatrogenic injury to the periarticular soft-tissue structures with subsequent reduction in the local and systemic inflammatory response, enhancing early outcomes. Despite the increased costs and radiation exposure associated with CT-based navigation, these many benefits have facilitated the adoption of imaged based robotic surgery into routine practice. Further research on ultra-low-dose CT scans and exploration of the possible translation of the use of 3D imaging into improved clinical outcomes are required to justify its broader implementation.

Cite this article: Bone Joint J 2024;106-B(9):892–897.


Bone & Joint 360
Vol. 12, Issue 5 | Pages 18 - 21
1 Oct 2023

The October 2023 Knee Roundup360 looks at: Cementless total knee arthroplasty is associated with more revisions within a year; Kinematically and mechanically aligned total knee arthroplasties: long-term follow-up; Aspirin thromboprophylaxis following primary total knee arthroplasty is associated with a lower rate of early periprosthetic joint infection compared with other agents; The impact of a revision arthroplasty network on patient outcomes; Re-revision knee arthroplasty in a tertiary centre: how does infection impact on outcomes?; Does the knee joint have its own microbiome?; Revision knee surgery provision in Scotland; Aspirin is a safe and effective thromboembolic prophylaxis after total knee arthroplasty: a systematic review and meta-analysis; Patellar resurfacing and kneeling ability after total knee arthroplasty: a systematic review.


Bone & Joint Open
Vol. 4, Issue 2 | Pages 53 - 61
1 Feb 2023
Faraj S de Windt TS van Hooff ML van Hellemondt GG Spruit M

Aims

The aim of this study was to assess the clinical and radiological results of patients who were revised using a custom-made triflange acetabular component (CTAC) for component loosening and pelvic discontinuity (PD) after previous total hip arthroplasty (THA).

Methods

Data were extracted from a single centre prospective database of patients with PD who were treated with a CTAC. Patients were included if they had a follow-up of two years. The Hip Disability and Osteoarthritis Outcome Score (HOOS), modified Oxford Hip Score (mOHS), EurQol EuroQoL five-dimension three-level (EQ-5D-3L) utility, and Numeric Rating Scale (NRS), including visual analogue score (VAS) for pain, were gathered at baseline, and at one- and two-year follow-up. Reasons for revision, and radiological and clinical complications were registered. Trends over time are described and tested for significance and clinical relevance.


The Bone & Joint Journal
Vol. 105-B, Issue 2 | Pages 102 - 108
1 Feb 2023
MacDessi SJ Oussedik S Abdel MP Victor J Pagnano MW Haddad FS

Orthopaedic surgeons are currently faced with an overwhelming number of choices surrounding total knee arthroplasty (TKA), not only with the latest technologies and prostheses, but also fundamental decisions on alignment philosophies. From ‘mechanical’ to ‘adjusted mechanical’ to ‘restricted kinematic’ to ‘unrestricted kinematic’ — and how constitutional alignment relates to these — there is potential for ambiguity when thinking about and discussing such concepts. This annotation summarizes the various alignment strategies currently employed in TKA. It provides a clear framework and consistent language that will assist surgeons to compare confidently and contrast the concepts, while also discussing the latest opinions about alignment in TKA. Finally, it provides suggestions for applying consistent nomenclature to future research, especially as we explore the implications of 3D alignment patterns on patient outcomes.

Cite this article: Bone Joint J 2023;105-B(2):102–108.


Bone & Joint Open
Vol. 5, Issue 8 | Pages 688 - 696
22 Aug 2024
Hanusrichter Y Gebert C Steinbeck M Dudda M Hardes J Frieler S Jeys LM Wessling M

Aims

Custom-made partial pelvis replacements (PPRs) are increasingly used in the reconstruction of large acetabular defects and have mainly been designed using a triflange approach, requiring extensive soft-tissue dissection. The monoflange design, where primary intramedullary fixation within the ilium combined with a monoflange for rotational stability, was anticipated to overcome this obstacle. The aim of this study was to evaluate the design with regard to functional outcome, complications, and acetabular reconstruction.

Methods

Between 2014 and 2023, 79 patients with a mean follow-up of 33 months (SD 22; 9 to 103) were included. Functional outcome was measured using the Harris Hip Score and EuroQol five-dimension questionnaire (EQ-5D). PPR revisions were defined as an endpoint, and subgroups were analyzed to determine risk factors.


Bone & Joint Open
Vol. 5, Issue 4 | Pages 260 - 268
1 Apr 2024
Broekhuis D Meurs WMH Kaptein BL Karunaratne S Carey Smith RL Sommerville S Boyle R Nelissen RGHH

Aims

Custom triflange acetabular components (CTACs) play an important role in reconstructive orthopaedic surgery, particularly in revision total hip arthroplasty (rTHA) and pelvic tumour resection procedures. Accurate CTAC positioning is essential to successful surgical outcomes. While prior studies have explored CTAC positioning in rTHA, research focusing on tumour cases and implant flange positioning precision remains limited. Additionally, the impact of intraoperative navigation on positioning accuracy warrants further investigation. This study assesses CTAC positioning accuracy in tumour resection and rTHA cases, focusing on the differences between preoperative planning and postoperative implant positions.

Methods

A multicentre observational cohort study in Australia between February 2017 and March 2021 included consecutive patients undergoing acetabular reconstruction with CTACs in rTHA (Paprosky 3A/3B defects) or tumour resection (including Enneking P2 peri-acetabular area). Of 103 eligible patients (104 hips), 34 patients (35 hips) were analyzed.


The Bone & Joint Journal
Vol. 104-B, Issue 5 | Pages 604 - 612
1 May 2022
MacDessi SJ Wood JA Diwan A Harris IA

Aims

Intraoperative pressure sensors allow surgeons to quantify soft-tissue balance during total knee arthroplasty (TKA). The aim of this study was to determine whether using sensors to achieve soft-tissue balance was more effective than manual balancing in improving outcomes in TKA.

Methods

A multicentre randomized trial compared the outcomes of sensor balancing (SB) with manual balancing (MB) in 250 patients (285 TKAs). The primary outcome measure was the mean difference in the four Knee injury and Osteoarthritis Outcome Score subscales (ΔKOOS4) in the two groups, comparing the preoperative and two-year scores. Secondary outcomes included intraoperative balance data, additional patient-reported outcome measures (PROMs), and functional measures.