Bone & Joint Open
Vol. 2, Issue 5 | Pages 278 - 292
3 May 2021
Miyamoto S Iida S Suzuki C Nakatani T Kawarai Y Nakamura J Orita S Ohtori S

Aims. The main aims were to identify risk factors predictive of a radiolucent line (RLL) around the acetabular component with an interface bioactive bone cement (IBBC) technique in the first year after THA, and evaluate whether these risk factors influence the development of RLLs at five and ten years after THA. Methods. A retrospective review was undertaken of 980 primary cemented THAs in 876 patients using cemented acetabular components with the IBBC technique. The outcome variable was any RLLs that could be observed around the acetabular component at the first year after THA. Univariate analyses with univariate logistic regression and multivariate analyses with exact logistic regression were performed to identify risk factors for any RLLs based on radiological classification of hip osteoarthritis. Results. RLLs were detected in 27.2% of patients one year postoperatively. In multivariate regression analysis controlling for confounders, atrophic osteoarthritis (odds ratio (OR) 2.17 (95% confidence interval (CI), 1.04 to 4.49); p = 0.038) and 26 mm (OR 3.23 (95% CI 1.85 to 5.66); p < 0.001) or 28 mm head diameter (OR 3.64 (95% CI 2.07 to 6.41); p < 0.001) had a significantly greater risk for any RLLs one year after surgery. Structural bone graft (OR 0.19 (95% CI 0.13 to 0.29) p < 0.001) and location of the hip centre within the true acetabular region (OR 0.15 (95% CI 0.09 to 0.24); p < 0.001) were significantly less prognostic. Improvement of the cement-bone interface including complete disappearance and poorly defined RLLs was identified in 15.1% of patients. Kaplan-Meier survival analysis for the acetabular component at ten years with revision of the acetabular component for aseptic loosening as the end point was 100.0% with a RLL and 99.1% without a RLL (95% CI 97.9 to 100). With revision of the acetabular component for any reason as the end point, the survival rate was 99.2% with a RLL (95% CI 97.6 to 100) and 96.5% without a RLL (95% CI 93.4 to 99.7). Conclusion. This study demonstrates that acetabular bone quality, head diameter, structural bone graft, and hip centre position may influence the presence of the any RLL. Cite this article: Bone Joint Open 2021;2(5):278–292


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1604 - 1610
1 Oct 2021
Takaoka Y Goto K Tamura J Okuzu Y Kawai T Kuroda Y Orita K Matsuda S

Aims

We aimed to evaluate the long-term outcome of highly cross-linked polyethylene (HXLPE) cemented acetabular components and assess whether any radiolucent lines (RLLs) which arose were progressive.

Methods

We retrospectively reviewed 170 patients who underwent 187 total hip arthroplasties at two hospitals with a minimum follow-up of ten years. All interventions were performed using the same combination of HXLPE cemented acetabular components with femoral stems made of titanium alloy. Kaplan-Meier survival analysis was performed for the primary endpoint of acetabular component revision surgery for any reason and secondary endpoint of the appearance of RLLs. RLLs that had appeared once were observed over time. We statistically assessed potential relationships between RLLs and a number of factors, including the technique of femoral head autografting and the Japanese Orthopaedic Association score.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1013 - 1018
1 Aug 2008
Goto K Kawanabe K Akiyama H Morimoto T Nakamura T

We reviewed 44 consecutive revision hip replacements in 38 patients performed using the cement-in-cement technique. All were performed for acetabular loosening in the presence of a well-fixed femoral component. The mean follow-up was 5.1 years (2 to 10.1). Radiological analysis at final follow-up indicated no loosening of the femoral component, except for one case with a continuous radiolucent line in all zones and peri-prosthetic fracture which required further revision. Peri-operative complications included nine proximal femoral fractures (20.4%) and perforation of the proximal femur in one hip. In five hips wiring or fixation with a braided suture was undertaken but no additional augmentation was required. There was an improvement in the mean Japanese Orthopaedic Association score from 55.5 (28 to 81) pre-operatively to 77.8 (40 to 95) at final follow-up (p < 0.001). Revision using a cement-in-cement technique allows increased exposure for acetabular revision and is effective in the medium term. Further follow-up is required to assess the long-term results in the light of in vitro studies which have questioned the quality of the cement-in-cement bond


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 586 - 591
1 May 2006
Wang J Fong C Su Y Yu H

Failure of total hip arthroplasty with acetabular deficiency occurred in 55 patients (60 hips) and was treated with acetabular revision using morsellised allograft and a cemented metal-backed component. A total of 50 patients (55 hips) were available for clinical and radiological evaluation at a mean follow-up of 5.8 years (3 to 9.5). No hip required further revision of the acetabular component because of aseptic loosening. All the hips except one had complete incorporation of the allograft demonstrated on the radiographs. A complete radiolucent line of > 1 mm was noted in two hips post-operatively. A good to excellent result occurred in 50 hips (91%). With radiological evidence of aseptic loosening of the acetabular component as the end-point, the survivorship at a mean of 5.8 years after surgery was 96.4%. The use of impacted allograft chips in combination with a cemented metal-backed acetabular component and screw fixation can achieve good medium-term results in patients with acetabular bone deficiency


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 540 - 547
1 Jun 2024
Nandra RS Elnahal WA Mayne A Brash L McBryde CW Treacy RBC

Aims

The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years.

Methods

The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.


Bone & Joint Open
Vol. 4, Issue 7 | Pages 507 - 515
6 Jul 2023
Jørgensen PB Jakobsen SS Vainorius D Homilius M Hansen TB Stilling M

Aims

The Exeter short stem was designed for patients with Dorr type A femora and short-term results are promising. The aim of this study was to evaluate the minimum five-year stem migration pattern of Exeter short stems in comparison with Exeter standard stems.

Methods

In this case-control study, 25 patients (22 female) at mean age of 78 years (70 to 89) received cemented Exeter short stem (case group). Cases were selected based on Dorr type A femora and matched first by Dorr type A and then age to a control cohort of 21 patients (11 female) at mean age of 74 years (70 to 89) who received with cemented Exeter standard stems (control group). Preoperatively, all patients had primary hip osteoarthritis and no osteoporosis as confirmed by dual X-ray absorptiometry scanning. Patients were followed with radiostereometry for evaluation of stem migration (primary endpoint), evaluation of cement quality, and Oxford Hip Score. Measurements were taken preoperatively, and at three, 12, and 24 months and a minimum five-year follow-up.


Bone & Joint Open
Vol. 3, Issue 11 | Pages 867 - 876
10 Nov 2022
Winther SS Petersen M Yilmaz M Kaltoft NS Stürup J Winther NS

Aims

Pelvic discontinuity is a rare but increasingly common complication of total hip arthroplasty (THA). This single-centre study evaluated the performance of custom-made triflange acetabular components in acetabular reconstruction with pelvic discontinuity by determining: 1) revision and overall implant survival rates; 2) discontinuity healing rate; and 3) Harris Hip Score (HHS).

Methods

Retrospectively collected data of 38 patients (39 hips) with pelvic discontinuity treated with revision THA using a custom-made triflange acetabular component were analyzed. Minimum follow-up was two years (mean 5.1 years (2 to 11)).


The Bone & Joint Journal
Vol. 104-B, Issue 9 | Pages 1039 - 1046
1 Sep 2022
Özdemir E Kuijpers MFL Visser J Schreurs BW Rijnen WHC

Aims

The aim of this study is to report the long-term outcomes of instrumented femoral revisions with impaction allograft bone grafting (IBG) using the X-change femoral revision system at 30 years after introduction of the technique.

Methods

We updated the outcomes of our previous study, based on 208 consecutive revisions using IBG and the X-change femoral revision system in combination with a cemented polished stem, performed in our tertiary care institute between 1991 and 2007. Kaplan-Meier survival analyses were used to determine the survival rate of the revisions with endpoint revision for any reason and aseptic loosening. Secondary outcomes were radiological loosening and patient-reported outcome measures.


The Bone & Joint Journal
Vol. 104-B, Issue 9 | Pages 1017 - 1024
1 Sep 2022
Morris WZ Justo PGS Williams KA Kim Y Millis MB Novais EN

Aims

The aims of this study were to characterize the incidence and risk factors associated with stress fractures following periacetabular osteotomy, and to determine their effect on osteotomy union.

Methods

We retrospectively reviewed all periacetabular osteotomies (PAOs) performed for developmental dysplasia of the hip (DDH) at one institution over a six-year period between 2012 and 2017. Perioperative factors were recorded, and included demographic and surgical data. Postoperatively, patients were followed for a minimum of one year with anteroposterior and false profile radiographs of the pelvis to monitor for evidence of stress fracture and union of osteotomies. We characterized the incidence and locations of stress fractures, and used univariate and multivariable analysis to identify factors predictive of stress fracture and the association of stress fracture on osteotomy union.


Bone & Joint Open
Vol. 3, Issue 1 | Pages 20 - 28
10 Jan 2022
Fujii H Hayama T Abe T Takahashi M Amagami A Matsushita Y Otani T Saito M

Aims

Although the short stem concept in hip arthroplasty procedure shows acceptable clinical performance, we sometimes get unexplainable radiological findings. The aim of this retrospective study was to evaluate changes of radiological findings up to three years postoperatively, and to assess any potential contributing factors on such radiological change in a Japanese population.

Methods

This is a retrospective radiological study conducted in Japan. Radiological assessment was done in accordance with predetermined radiological review protocol. A total of 241 hips were included in the study and 118 hips (49.0%) revealed radiological change from immediately after surgery to one year postoperatively; these 118 hips were eligible for further analyses. Each investigator screened whether either radiolucent lines (RLLs), cortical hypertrophy (CH), or atrophy (AT) appeared or not on the one-year radiograph. Further, three-year radiographs of eligible cases were reviewed to determine changes such as, disappeared (D), improved (I), stable (S), and progression (P). Additionally, bone condensation (BC) was assessed on the three-year radiograph.


Bone & Joint Open
Vol. 3, Issue 3 | Pages 229 - 235
11 Mar 2022
Syam K Unnikrishnan PN Lokikere NK Wilson-Theaker W Gambhir A Shah N Porter M

Aims

With increasing burden of revision hip arthroplasty (THA), one of the major challenges is the management of proximal femoral bone loss associated with previous multiple surgeries. Proximal femoral arthroplasty (PFA) has already been popularized for tumour surgeries. Our aim was to describe the outcome of using PFA in these demanding non-neoplastic cases.

Methods

A retrospective review of 25 patients who underwent PFA for non-neoplastic indications between January 2009 and December 2015 was undertaken. Their clinical and radiological outcome, complication rates, and survival were recorded. All patients had the Stanmore Implant – Modular Endo-prosthetic Tumour System (METS).


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 66 - 72
1 Jul 2021
Hernandez NM Hinton ZW Wu CJ Lachiewicz PF Ryan SP Wellman SS

Aims

Modular dual mobility (MDM) acetabular components are often used with the aim of reducing the risk of dislocation in revision total hip arthroplasty (THA). There is, however, little information in the literature about its use in this context. The aim of this study, therefore, was to evaluate the outcomes in a cohort of patients in whom MDM components were used at revision THA, with a mean follow-up of more than five years.

Methods

Using the database of a single academic centre, 126 revision THAs in 117 patients using a single design of an MDM acetabular component were retrospectively reviewed. A total of 94 revision THAs in 88 patients with a mean follow-up of 5.5 years were included in the study. Survivorship was analyzed with the endpoints of dislocation, reoperation for dislocation, acetabular revision for aseptic loosening, and acetabular revision for any reason. The secondary endpoints were surgical complications and the radiological outcome.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 122 - 128
1 Jul 2021
Tibbo ME Limberg AK Gausden EB Huang P Perry KI Yuan BJ Berry DJ Abdel MP

Aims

The prevalence of ipsilateral total hip arthroplasty (THA) and total knee arthroplasty (TKA) is rising in concert with life expectancy, putting more patients at risk for interprosthetic femur fractures (IPFFs). Our study aimed to assess treatment methodologies, implant survivorship, and IPFF clinical outcomes.

Methods

A total of 76 patients treated for an IPFF from February 1985 to April 2018 were reviewed. Prior to fracture, at the hip/knee sites respectively, 46 femora had primary/primary, 21 had revision/primary, three had primary/revision, and six had revision/revision components. Mean age and BMI were 74 years (33 to 99) and 30 kg/m2 (21 to 46), respectively. Mean follow-up after fracture treatment was seven years (2 to 24).


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 78 - 83
1 Jul 2021
Roedel GG Kildow BJ Sveom DS Garvin KL

Aims

Highly cross-linked polyethylene (HXLPE) has greatly improved the durability of total hip arthroplasty (THA) in young patients because of its improved wear characteristics. Few studies have followed this population into the second decade, and therefore the purpose of this investigation was to evaluate the clinical outcome for THA patients 50 years of age and younger at a minimum of 15 years postoperatively. The second purpose was to evaluate the radiological findings secondary to wear or mechanical failure of the implant.

Methods

Between October 1999 and December 2005, 105 THAs were performed in 95 patients (53 female, 42 male) aged 50 years and younger (mean 42 years (20 to 50)). There were 87 patients (96 hips) that were followed for a minimum of 15 years (mean 17.3 years (15 to 21)) for analysis. Posterior approach was used with cementless fixation with a median head size of 28 mm. HXLPE was the acetabular bearing for all hips. Radiographs were evaluated for polyethylene wear, radiolucent lines, and osteolysis.


The Bone & Joint Journal
Vol. 102-B, Issue 6 | Pages 709 - 715
1 Jun 2020
Abdelsamie KR Elhawary I Ali H Ali M EL-Shafie M Dominic Meek RM

Aims

Femoral revision component subsidence has been identified as predicting early failure in revision hip surgery. This comparative cohort study assessed the potential risk factors of subsidence in two commonly used femoral implant designs.

Methods

A comparative cohort study was undertaken, analyzing a consecutive series of patients following revision total hip arthroplasties using either a tapered-modular (TM) fluted titanium or a porous-coated cylindrical modular (PCM) titanium femoral component, between April 2006 and May 2018. Clinical and radiological assessment was compared for both treatment cohorts. Risk factors for subsidence were assessed and compared.


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 760 - 767
1 Jul 2019
Galea VP Rojanasopondist P Laursen M Muratoglu OK Malchau H Bragdon C

Aims

Vitamin E-diffused, highly crosslinked polyethylene (VEPE) and porous titanium-coated (PTC) shells were introduced in total hip arthroplasty (THA) to reduce the risk of aseptic loosening. The purpose of this study was: 1) to compare the wear properties of VEPE to moderately crosslinked polyethylene; 2) to assess the stability of PTC shells; and 3) to report their clinical outcomes at seven years.

Patients and Methods

A total of 89 patients were enrolled into a prospective study. All patients received a PTC shell and were randomized to receive a VEPE liner (n = 44) or a moderately crosslinked polyethylene (ModXLPE) liner (n = 45). Radiostereometric analysis (RSA) was used to measure polyethylene wear and component migration. Differences in wear were assessed while adjusting for body mass index, activity level, acetabular inclination, anteversion, and head size. Plain radiographs were assessed for radiolucency and patient-reported outcome measures (PROMs) were administered at each follow-up.


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 458 - 464
1 Apr 2017
Abrahams JM Kim YS Callary SA De Ieso C Costi K Howie DW Solomon LB

Aims

This study aimed to determine the diagnostic performance of radiographic criteria to detect aseptic acetabular loosening after revision total hip arthroplasty (THA). Secondary aims were to determine the predictive values of different thresholds of migration and to determine the predictive values of radiolucency criteria.

Patients and Methods

Acetabular component migration to re-revision was measured retrospectively using Ein-Bild-Rontgen-Analyse (EBRA-Cup) and manual measurements (Sutherland method) in two groups: Group A, 52 components (48 patients) found not loose at re-revision and Group B, 42 components (36 patients) found loose at re-revision between 1980 and 2015. The presence and extent of radiolucent lines was also assessed.


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 787 - 792
1 Jul 2019
Goto K Kuroda Y Kawai T Kawanabe K Matsuda S

Aims

In the 1990s, a bioactive bone cement (BABC) containing apatite-wollastonite glass-ceramic (AW-GC) powder and bisphenol-a-glycidyl methacrylate resin was developed at our hospital. In 1996, we used BABC to fix the acetabular component in primary total hip arthroplasty (THA) in 20 patients as part of a clinical trial. The purpose of this study was to investigate the long-term results of primary THA using BABC.

Patients and Methods

A total of 20 patients (three men and 17 women) with a mean age of 57.4 years (40 to 71), a mean body weight of 52.3 kg (39 to 64), and a mean body mass index (BMI) of 23.0 kg/m2 (19.8 to 28.6) were evaluated clinically and radiologically. Survival analyses were undertaken, and wear analyses were carried out using a computer-aided method.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 601 - 606
1 May 2017
Narkbunnam R Amanatullah DF Electricwala AJ Huddleston III JI Maloney WJ Goodman SB

Aims

The stability of cementless acetabular components is an important factor for surgical planning in the treatment of patients with pelvic osteolysis after total hip arthroplasty (THA). However, the methods for determining the stability of the acetabular component from pre-operative radiographs remain controversial. Our aim was to develop a scoring system to help in the assessment of the stability of the acetabular component under these circumstances.

Patients and Methods

The new scoring system is based on the mechanism of failure of these components and the location of the osteolytic lesion, according to the DeLee and Charnley classification. Each zone is evaluated and scored separately. The sum of the individual scores from the three zones is reported as a total score with a maximum of 10 points. The study involved 96 revision procedures which were undertaken for wear or osteolysis in 91 patients between July 2002 and December 2012. Pre-operative anteroposterior pelvic radiographs and Judet views were reviewed. The stability of the acetabular component was confirmed intra-operatively.


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1002 - 1009
1 Aug 2018
Westerman RW Whitehouse SL Hubble MJW Timperley AJ Howell JR Wilson MJ

Aims

The aim of this study was to report the initial results of the Exeter V40 stem, which became available in 2000.

Patients and Methods