Advertisement for orthosearch.org.uk
Results 1 - 20 of 174
Results per page:
Bone & Joint Open
Vol. 4, Issue 8 | Pages 551 - 558
1 Aug 2023
Thomas J Shichman I Ohanisian L Stoops TK Lawrence KW Ashkenazi I Watson DT Schwarzkopf R

Aims. United Classification System (UCS) B2 and B3 periprosthetic fractures in total hip arthroplasties (THAs) have been commonly managed with modular tapered stems. No study has evaluated the use of monoblock fluted tapered titanium stems for this indication. This study aimed to evaluate the effects of a monoblock stems on implant survivorship, postoperative outcomes, radiological outcomes, and osseointegration following treatment of THA UCS B2 and B3 periprosthetic fractures. Methods. A retrospective review was conducted of all patients who underwent revision THA (rTHA) for periprosthetic UCS B2 and B3 periprosthetic fracture who received a single design monoblock fluted tapered titanium stem at two large, tertiary care, academic hospitals. A total of 72 patients met inclusion and exclusion criteria (68 UCS B2, and four UCS B3 fractures). Primary outcomes of interest were radiological stem subsidence (> 5 mm), radiological osseointegration, and fracture union. Sub-analysis was also done for 46 patients with minimum one-year follow-up. Results. For the total cohort, stem osseointegration, fracture union, and stem subsidence were 98.6%, 98.6%, and 6.9%, respectively, at latest follow-up (mean follow-up 27.0 months (SD 22.4)). For patients with minimum one-year of follow-up, stem osseointegration, fracture union, and stem subsidence were 97.8%, 97.8%, and 6.5%, respectively. Conclusion. Monoblock fluted stems can be an acceptable modality for the management of UCS B2 periprosthetic fractures in rTHAs due to high rates of stem osseointegration and survival, and the low rates of stem subsidence, and revision. Further research on the use of this stem for UCS B3 periprosthetic fractures is warranted to determine if the same conclusion can be made for this fracture pattern. Cite this article: Bone Jt Open 2023;4(8):551–558


Bone & Joint Research
Vol. 6, Issue 5 | Pages 351 - 357
1 May 2017
Takahashi E Kaneuji A Tsuda R Numata Y Ichiseki T Fukui K Kawahara N

Objectives. Favourable results for collarless polished tapered stems have been reported, and cement creep due to taper slip may be a contributing factor. However, the ideal cement thickness around polished stems remains unknown. We investigated the influence of cement thickness on stem subsidence and cement creep. Methods. We cemented six collarless polished tapered (CPT) stems (two stems each of small, medium and large sizes) into composite femurs that had been reamed with a large CPT rasp to achieve various thicknesses of the cement mantle. Two or three tantalum balls were implanted in the proximal cement in each femur. A cyclic loading test was then performed for each stem. The migration of the balls was measured three-dimensionally, using a micro-computed tomography (CT) scanner, before and after loading. A digital displacement gauge was positioned at the stem shoulder, and stem subsidence was measured continuously by the gauge. Final stem subsidence was measured at the balls at the end of each stem. Results. A strong positive correlation was observed between mean cement thickness and stem subsidence in the CT slices on the balls. In the small stems, the balls moved downward to almost the same extent as the stem. There was a significant negative correlation between cement thickness and the horizontal:downward ratio of ball movement. Conclusion. Collarless polished tapered stems with thicker cement mantles resulted in greater subsidence of both stem and cement. This suggests that excessive thickness of the cement mantle may interfere with effective radial cement creep. Cite this article: E. Takahashi, A. Kaneuji, R. Tsuda, Y. Numata, T. Ichiseki, K. Fukui, N. Kawahara. The influence of cement thickness on stem subsidence and cement creep in a collarless polished tapered stem: When are thick cement mantles detrimental? Bone Joint Res 2017;6:–357. DOI: 10.1302/2046-3758.65.BJR-2017-0028.R1


The Bone & Joint Journal
Vol. 105-B, Issue 5 | Pages 481 - 486
1 May 2023
Scott CEH Jain S Moran M Haddad FS

The Unified Classification System (UCS), or Vancouver system, is a validated and widely used classification system to guide the management of periprosthetic femoral fractures. It suggests that well-fixed stems (type B1) can be treated with fixation but that loose stems (types B2 and B3) should be revised. Determining whether a stem is loose can be difficult and some authors have questioned how to apply this classification system to polished taper slip stems which are, by definition, loose within their cement mantle. Recent evidence has challenged the common perception that revision surgery is preferable to fixation surgery for UCS-B periprosthetic fractures around cemented polished taper slip stems. Indications for fixation include an anatomically reducible fracture and cement mantle, a well-fixed femoral bone-cement interface, and a well-functioning acetabular component. However, not all type B fractures can or should be managed with fixation due to the risk of early failure. This annotation details specific fracture patterns that should not be managed with fixation alone. Cite this article: Bone Joint J 2023;105-B(5):481–486


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 16 - 22
1 Jan 2008
Yates PJ Burston BJ Whitley E Bannister GC

We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known. At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%). The collarless polished tapered stem has an excellent clinical and radiological outcome at a minimum of ten years’ follow-up. The pattern and magnitude of subsidence of the stem within the cement mantle occurred in a predictable pattern, consistent with the design philosophy


The Bone & Joint Journal
Vol. 106-B, Issue 3 | Pages 220 - 223
1 Mar 2024
Kayani B Luo TD Haddad FS


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 644 - 649
1 Apr 2021
Alsousou J Oragu E Martin A Strickland L Newman S Kendrick B Taylor A Glyn-Jones S

Aims. The aim of this prospective cohort study was to evaluate the early migration of the TriFit cementless proximally coated tapered femoral stem using radiostereometric analysis (RSA). Methods. A total of 21 patients (eight men and 13 women) undergoing primary total hip arthroplasty (THA) for osteoarthritis of the hip were recruited in this study and followed up for two years. Two patients were lost to follow-up. All patients received a TriFit stem and Trinity Cup with a vitamin E-infused highly cross-linked ultra-high molecular weight polyethylene liner. Radiographs for RSA were taken postoperatively and then at three, 12, and 24 months. Oxford Hip Score (OHS), EuroQol five-dimension questionnaire (EQ-5D), and adverse events were reported. Results. At two years, the mean subsidence of the head and tip for the TriFit stem was 0.38 mm (SD 0.32) and 0.52 mm (SD 0.36), respectively. The total migration of the head and tip was 0.55 mm (SD 0.32) and 0.71 mm (SD 0.38), respectively. There were no statistically significant differences between the three to 12 months' migration (p = 0.105) and 12 to 24 months' migration (p = 0.694). The OHS and EQ-5D showed significant improvements at two years. Conclusion. The results of this study suggest that the TriFit femoral stem achieves initial stability and is likely to be stable in the mid and long term. A long-term outcome study is required to assess late mechanisms of failure and the effects of bone mineral density (BMD) related changes. Cite this article: Bone Joint J 2021;103-B(4):644–649


Aims. The aims of this study were to evaluate the incidence of reoperation (all cause and specifically for periprosthetic femoral fracture (PFF)) and mortality, and associated risk factors, following a hemiarthroplasty incorporating a cemented collarless polished taper slip stem (PTS) for management of an intracapsular hip fracture. Methods. This retrospective study included hip fracture patients aged 50 years and older treated with Exeter (PTS) bipolar hemiarthroplasty between 2019 and 2022. Patient demographics, place of domicile, fracture type, delirium status, American Society of Anesthesiologists (ASA) grade, length of stay, and mortality were collected. Reoperation and mortality were recorded up to a median follow-up of 29.5 months (interquartile range 12 to 51.4). Cox regression was performed to evaluate independent risk factors associated with reoperation and mortality. Results. The cohort consisted of 1,619 patients with a mean age of 82.2 years (50 to 104), of whom 1,100 (67.9%) were female. In total, 29 patients (1.8%) underwent a reoperation; 12 patients (0.7%) sustained a PFF during the observation period (United Classification System (UCS)-A n = 2; UCS-B n = 5; UCS-C n = 5), of whom ten underwent surgical management. Perioperative delirium was independently associated with the occurrence of PFF (hazard ratio (HR) 5.92; p = 0.013) and surgery for UCS-B PFF (HR 21.7; p = 0.022). Neither all-cause reoperation nor PFF-related surgery was independently associated with mortality (HR 0.66; p = 0.217 and HR 0.38; p = 0.170, respectively). Perioperative delirium, male sex, older age, higher ASA grade, and pre-fracture residential status were independently associated with increased mortality risk following hemiarthroplasty (p < 0.001). Conclusion. The cumulative incidence of PFF at four years was 1.1% in elderly patients following cemented PTS hemiarthroplasty for a hip fracture. Perioperative delirium was independently associated with a PFF. However, reoperation for PPF was not independently associated with patient mortality after adjusting for patient-specific factors. Cite this article: Bone Jt Open 2024;5(4):269–276


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 889 - 894
1 Jul 2012
Burston BJ Barnett AJ Amirfeyz R Yates PJ Bannister GC

We prospectively followed 191 consecutive collarless polished tapered (CPT) femoral stems, implanted in 175 patients who had a mean age at operation of 64.5 years (21 to 85). At a mean follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were still alive. The fate of all original stems is known. The 16-year survivorship with re-operation for any reason was 80.7% (95% confidence interval 72 to 89.4). There was no loss to follow-up, with clinical data available on all 95 hips and radiological assessment performed on 90 hips (95%). At latest follow-up, the mean Harris hip score was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48). Stems subsided within the cement mantle, with a mean subsidence of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem (0.5%) has been revised due to aseptic loosening. In total seven stems were revised for any cause, of which four revisions were required for infection following revision of the acetabular component. A total of 21 patients (11%) required some sort of revision procedure; all except three of these resulted from failure of the acetabular component. Cemented acetabular components had a significantly lower revision burden (three hips, 2.7%) than Harris Galante uncemented components (17 hips, 21.8%) (p < 0.001). The CPT stem continues to provide excellent radiological and clinical outcomes at 15 years following implantation. Its results are consistent with other polished tapered stem designs


Bone & Joint Open
Vol. 5, Issue 10 | Pages 858 - 867
11 Oct 2024
Yamate S Hamai S Konishi T Nakao Y Kawahara S Hara D Motomura G Nakashima Y

Aims. The aim of this study was to evaluate the suitability of the tapered cone stem in total hip arthroplasty (THA) in patients with excessive femoral anteversion and after femoral osteotomy. Methods. We included patients who underwent THA using Wagner Cone due to proximal femur anatomical abnormalities between August 2014 and January 2019 at a single institution. We investigated implant survival time using the endpoint of dislocation and revision, and compared the prevalence of prosthetic impingements between the Wagner Cone, a tapered cone stem, and the Taperloc, a tapered wedge stem, through simulation. We also collected Oxford Hip Score (OHS), visual analogue scale (VAS) satisfaction, and VAS pain by postal survey in August 2023 and explored variables associated with those scores. Results. Of the 58 patients (62 hips), two (two hips) presented with dislocation or reoperation, and Kaplan-Meier analysis indicated a five-year survival rate of 96.7% (95% CI 92.4 to 100). Mean stem anteversion was 35.2° (SD 18.2°) for the Taperloc stem and 29.8° (SD 7.9°) for the Wagner Cone stem; mean reduction from Taperloc to Wagner Cone was 5.4° (SD 18.8°). Overall, 55 hips (52 patients) were simulated, and the prevalence of prosthetic impingement was lower for the Wagner Cone (5.5%, 3/55) compared with the Taperloc (20.0%, 11/55) stem, with an odds ratio of 0.20 (p = 0.038). Among the 33 respondents to the postal survey (36 hips), the mean scores were VAS pain 10.9, VAS satisfaction 86.9, and OHS 44.7. A multivariable analysis revealed that reduction of stem anteversion from Taperloc to Wagner Cone was more favourable for VAS pain (p = 0.029) and VAS satisfaction (p = 0.002). Conclusion. The mid-term survival rate for THA using the Wagner Cone stem was high, which may be supported by a reduction in prosthetic impingement. The reduction in excessive stem anteversion by using a tapered cone stem was associated with reduced pain and increased patient satisfaction. Cite this article: Bone Jt Open 2024;5(10):858–867


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 125 - 132
1 May 2024
Carender CN Perry KI Sierra RJ Trousdale RT Berry DJ Abdel MP

Aims

Uncemented implants are now commonly used at reimplantation of a two-stage revision total hip arthoplasty (THA) following periprosthetic joint infection (PJI). However, there is a paucity of data on the performance of the most commonly used uncemented femoral implants – modular fluted tapered (MFT) femoral components – in this setting. This study evaluated implant survival, radiological results, and clinical outcomes in a large cohort of reimplantation THAs using MFT components.

Methods

We identified 236 reimplantation THAs from a single tertiary care academic institution from September 2000 to September 2020. Two designs of MFT femoral components were used as part of an established two-stage exchange protocol for the treatment of PJI. Mean age at reimplantation was 65 years (SD 11), mean BMI was 32 kg/m2 (SD 7), and 46% (n = 109) were female. Mean follow-up was seven years (SD 4). A competing risk model accounting for death was used.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1000 - 1004
1 Aug 2008
Wraighte PJ Howard PW

Femoral impaction bone allografting has been developed as a means of restoring bone stock in revision total hip replacement. We report the results of 75 consecutive patients (75 hips) with a mean age of 68 years (35 to 87) who underwent impaction grafting using the Exeter collarless, polished, tapered femoral stem between 1992 and 1998. The mean follow-up period was 10.5 years (6.3 to 14.1). The median pre-operative bone defect score was 3 (interquartile range (IQR) 2 to 3) using the Endo-Klinik classification. The median subsidence at one year post-operatively was 2 mm (IQR 1 to 3). At the final review the median Harris hip score was 80.6 (IQR 67.6 to 88.9) and the median subsidence 2 mm (IQR 1 to 4). Incorporation of the allograft into trabecular bone and secondary remodelling were noted radiologically at the final follow-up in 87% (393 of 452 zones) and 40% (181 of 452 zones), respectively. Subsidence of the Exeter stem correlated with the pre-operative Endo-Klinik bone loss score (p = 0.037). The degree of subsidence at one year had a strong association with long-term subsidence (p < 0.001). There was a significant correlation between previous revision surgery and a poor Harris Hip score (p = 0.028), and those who had undergone previous revision surgery for infection had a higher risk of complications (p = 0.048). Survivorship at 10.5 years with any further femoral operation as the end-point was 92% (95% confidence interval 82 to 97)


Bone & Joint Open
Vol. 4, Issue 7 | Pages 472 - 477
1 Jul 2023
Xiang W Tarity TD Gkiatas I Lee H Boettner F Rodriguez JA Wright TM Sculco PK

Aims. When performing revision total hip arthroplasty using diaphyseal-engaging titanium tapered stems (TTS), the recommended 3 to 4 cm of stem-cortical diaphyseal contact may not be available. In challenging cases such as these with only 2 cm of contact, can sufficient axial stability be achieved and what is the benefit of a prophylactic cable? This study sought to determine, first, whether a prophylactic cable allows for sufficient axial stability when the contact length is 2 cm, and second, if differing TTS taper angles (2° vs 3.5°) impact these results. Methods. A biomechanical matched-pair cadaveric study was designed using six matched pairs of human fresh cadaveric femora prepared so that 2 cm of diaphyseal bone engaged with 2° (right femora) or 3.5° (left femora) TTS. Before impaction, three matched pairs received a single 100 lb-tensioned prophylactic beaded cable; the remaining three matched pairs received no cable adjuncts. Specimens underwent stepwise axial loading to 2600 N or until failure, defined as stem subsidence > 5 mm. Results. All specimens without cable adjuncts (6/6 femora) failed during axial testing, while all specimens with a prophylactic cable (6/6) successfully resisted axial load, regardless of taper angle. In total, four of the failed specimens experienced proximal longitudinal fractures, three of which occurred with the higher 3.5° TTS. One fracture occurred in a 3.5° TTS with a prophylactic cable yet passed axial testing, subsiding < 5 mm. Among specimens with a prophylactic cable, the 3.5° TTS resulted in lower mean subsidence (0.5 mm (SD 0.8)) compared with the 2° TTS (2.4 mm (SD 1.8)). Conclusion. A single prophylactic beaded cable dramatically improved initial axial stability when stem-cortex contact length was 2 cm. All implants failed secondary to fracture or subsidence > 5 mm when a prophylactic cable was not used. A higher taper angle appears to decrease the magnitude of subsidence but increased the fracture risk. The fracture risk was mitigated by the use of a prophylactic cable. Cite this article: Bone Jt Open 2023;4(7):472–477


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1210 - 1215
1 Sep 2005
Pospischill M Knahr K

We carried out a clinical and radiological review of 103 cementless primary hip arthroplasties with a tapered rectangular grit-blasted titanium press-fit femoral component and a threaded conical titanium acetabular component at a mean follow-up of 14.4 years (10.2 to 17.1).

The mean Harris hip score at the last follow-up was 89.2 (32 to 100). No early loosening and no fracture of the implant were found. One patient needed revision surgery because of a late deep infection. In 11 hips (10.7%), the reason for revision was progressive wear of the polyethylene liner. Exchange of the acetabular component because of aseptic loosening without detectable liner wear was carried out in three hips (2.9%).

After 15 years the survivorship with aseptic loosening as the definition for failure was 95.6% for the acetabular component and 100% for the femoral component.


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 755 - 761
1 Jun 2015
Flatøy B Röhrl SM Rydinge J Dahl J Diep LM Nordsletten L

Cemented femoral stems with force closed fixation designs have shown good clinical results despite high early subsidence. A new triple-tapered stem in this category (C-stem AMT) was introduced in 2005. This study compares this new stem with an established stem of similar design (Exeter) in terms of migration (as measured using radiostereometric analysis), peri-prosthetic bone remodelling (measured using dual energy x-ray densitometry, DXA), Oxford Hip Score, and plain radiographs.

A total of 70 patients (70 hips) with a mean age of 66 years (53 to 78) were followed for two years. Owing to missing data of miscellaneous reasons, the final analysis represents data from 51 (RSA) and 65 (DXA) patients. Both stems showed a typical pattern of migration: Subsidence and retroversion that primarily occurred during the first three months. C-stem AMT subsided less during the first three months (p = 0.01), before stabilising at a subsidence rate similar to the Exeter stem from years one to two. The rate of migration into retroversion was slightly higher for C-stem AMT during the second year (p = 0.03). Whilst there were slight differences in movement patterns between the stems, the C-stem AMT exhibits good early clinical outcomes and displays a pattern of migration and bone remodelling that predicts good clinical performance.

Cite this article: Bone Joint J 2015;97-B:755–61.


The Bone & Joint Journal
Vol. 97-B, Issue 8 | Pages 1031 - 1037
1 Aug 2015
da Assunção RE Pollard TCB Hrycaiczuk A Curry J Glyn-Jones S Taylor A

Periprosthetic femoral fracture (PFF) is a potentially devastating complication after total hip arthroplasty, with historically high rates of complication and failure because of the technical challenges of surgery, as well as the prevalence of advanced age and comorbidity in the patients at risk. This study describes the short-term outcome after revision arthroplasty using a modular, titanium, tapered, conical stem for PFF in a series of 38 fractures in 37 patients. The mean age of the cohort was 77 years (47 to 96). A total of 27 patients had an American Society of Anesthesiologists grade of at least 3. At a mean follow-up of 35 months (4 to 66) the mean Oxford Hip Score (OHS) was 35 (15 to 48) and comorbidity was significantly associated with a poorer OHS. All fractures united and no stem needed to be revised. Three hips in three patients required further surgery for infection, recurrent PFF and recurrent dislocation and three other patients required closed manipulation for a single dislocation. One stem subsided more than 5 mm but then stabilised and required no further intervention. . In this series, a modular, tapered, conical stem provided a versatile reconstruction solution with a low rate of complications. Cite this article: Bone Joint J 2015;97-B:1031–7


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 17 - 20
1 Nov 2013
Munro JT Masri BA Garbuz DS Duncan CP

Tapered, fluted, modular, titanium stems are increasingly popular in the operative management of Vancouver B2 and selected B3 peri-prosthetic femoral fractures. We have reviewed the results at our institution looking at stem survival and clinical outcomes and compared this with reported outcomes in the literature. Stem survival at a mean of 54 months was 96% in our series and 97% for combined published cases. Review of radiology showed maintenance or improvement of bone stock in 89% of cases with high rates of femoral union. Favourable clinical outcome scores have reported by several authors. No difference in survival or clinical scores was observed between B2 and B3 fractures. Tapered stems are a useful option in revision for femoral fracture across the spectrum of femoral bone deficiency.

Cite this article: Bone Joint J 2013;95-B, Supple A:17–20.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 46 - 52
1 Jul 2021
McGoldrick NP Fischman D Nicol GM Kreviazuk C Grammatopoulos G Beaulé PE

Aims

The aim of this study was to radiologically evaluate the quality of cement mantle and alignment achieved with a polished tapered cemented femoral stem inserted through the anterior approach and compared with the posterior approach.

Methods

A comparative retrospective study of 115 consecutive hybrid total hip arthroplasties or cemented hemiarthroplasties in 110 patients, performed through anterior (n = 58) or posterior approach (n = 57) using a collarless polished taper-slip femoral stem, was conducted. Cement mantle quality and thickness were assessed in both planes. Radiological outcomes were compared between groups.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 50 - 53
1 Jan 2016
Konan S Garbuz DS Masri BA Duncan CP

Tapered fluted titanium stems are increasingly used for femoral revision arthroplasty. They are available in modular and non-modular forms. Modularity has advantages when the bone loss is severe, the proximal femur is mis shapen or the surgeon is unfamiliar with the implant, but it introduces the risk of fracture of the stem at the junction between it and the proximal body segment. For that reason, and while awaiting intermediate-term results of more recently introduced designs of this junction, non-modularity has attracted attention, at least for straightforward revision cases.

We review the risks and causes of fracture of tapered titanium modular revision stems and present an argument in favour of the more selective use of modular designs.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):50–3.


Aims

To evaluate the hypothesis that failed osteosynthesis of periprosthetic Vancouver type B1 fractures can be treated successfully with stem revision using a transfemoral approach and a cementless, modular, tapered revision stem with reproducible rates of fracture healing, stability of the revision stem, and clinically good results.

Patients and Methods

A total of 14 patients (11 women, three men) with a mean age of 72.4 years (65 to 90) undergoing revision hip arthroplasty after failed osteosynthesis of periprosthetic fractures of Vancouver type B1 were treated using a transfemoral approach to remove the well-fixed stem before insertion of a modular, fluted titanium stem which obtained distal fixation. These patients were clinically and radiologically followed up for a mean 52.2 months (24 to 144).


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1565 - 1571
1 Dec 2018
Kristensen TB Dybvik E Furnes O Engesæter LB Gjertsen J

Aims. The aim of this large registry-based study was to compare mid-term survival rates of cemented femoral stems of different designs used in hemiarthroplasty for a fracture of the femoral neck. Patients and Methods. From the Norwegian Hip Fracture Register (NHFR), 20 532 primary cemented bipolar hemiarthroplasties, which were undertaken in patients aged > 70 years with a femoral neck fracture between 2005 and 2016, were included. Polished tapered stems (n = 12 065) (Exeter and CPT), straight stems (n = 5545) (Charnley, Charnley Modular, and Spectron EF), and anatomical stems (n = 2922) (Lubinus SP2) were included. The survival of the implant with any reoperation as the endpoint was calculated using the Kaplan–Meier method and hazard ratios (HRs), and the different indications for reoperation were calculated using Cox regression analysis. Results. The one-year survival was 96.0% (95% confidence interval (CI) 95.6 to 96.4) for the Exeter stem, 97.0% (95% CI 96.4 to 97.6) for the Lubinus SP2 stem, 97.6% (95% CI 97.0 to 98.2) for the Charnley stem, 98.1% (95% CI 97.3 to 98.9) for the Spectron EF stem, and 96.4% (95% CI 95.6 to 97.2) for the Charnley Modular stem, respectively. The hazard ratio for reoperation after one year was lower for Lubinus SP2 (HR 0.77, 95% CI 0.60 to 0.97), Charnley (HR 0.64, 95% CI 0.48 to 0.86), and Spectron EF stems (HR 0.44, 95% CI 0.29 to 0.67) compared with the Exeter stem. Reoperation for periprosthetic fracture occurred almost exclusively after the use of polished tapered stems. Conclusion. We were able to confirm that implant survival after cemented hemiarthroplasty for a hip fracture is high. Differences in rates of reoperation seem to favour anatomical and straight stems compared with polished tapered stems, which had a higher risk of periprosthetic fracture