In 2004 we described the ten-year prospective results of 38 total hip replacements using the Furlong hydroxyapatite-ceramic-coated femoral component in 35 patients <
50 years old. We have now reviewed the surviving 35 arthroplasties in 33 patients at a mean of 16 years (10.3 to 19.9). The mean age of the surviving patients at the time of operation was 41.3 years (26.0 to 49.0). Of these, eight have undergone revision of their acetabular component for aseptic loosening. None of the femoral components has had revision for aseptic loosening giving a survival rate of 100% at 16 years (95% confidence interval 89% to 100%). The Furlong hydroxyapatite-ceramic-coated femoral component gives excellent long-term survival in young and active patients.
We describe a case of symptomatic focal femoral osteolysis around a screw hole distal to the hydroxyapatite-coated portion of a cannulated femoral component in a revision hip replacement. No locking screw had been inserted into this, the most proximal of the three distal holes for locking screws. The presence of polyethylene wear debris in the tissue excised from the lesion suggested that it had passed through the cannulated portion of the stem and out of the proximal unfilled distal locking hole, initiating an osteolytic reaction in an otherwise well-fixed stem. This case highlights an important design characteristic of such cannulated, uncemented femoral components. We recommend that the proximal aperture of these cannulated stems be occluded at implantation.