The modification of bearing surfaces with poly(2-methacryloyloxyethyl phosphorylcholine) (PMPC) is known to increase the hydration of the surfaces and decrease the wear of the substrates. PMPC grafting to acetabular liner of total hip arthroplasty showed a drastic reduction of cross-linked polyethylene (CLPE) wear in a long-term hip simulator test and achieved a good short-term clinical result. To apply this technique to other joint prostheses, the wear resistance under various conditions needs to be evaluated because every joint has a different wear mode. ASTM F732 gives a method that disk shaped polymer specimen is loaded with hemispherical pin using
Aims. Highly cross-linked polyethylene (HXLPE) greatly reduces wear in total hip arthroplasty, compared to conventional polyethylene (CPE). Cross-linking is commonly achieved by irradiation. This study aimed to compare the degree of cross-linking and in vitro wear rates across a cohort of retrieved and unused polyethylene cups/liners from various brands. Methods. Polyethylene acetabular cups/liners were collected at one centre from 1 April 2021 to 30 April 2022. The trans-vinylene index (TVI) and oxidation index (OI) were determined by Fourier-transform infrared spectrometry. Wear was measured using a
The main objective of joint arthroplasty is to improve activities of daily living of the patient. However, normal daily activities may lead to separation of articular surfaces of an artificial joint, possibly as a result of a combined impact and sliding motion. Therefore, the properties of articular surfaces define the durability of implant materials. Modification of bearing surfaces with poly(2-methacryloyloxyethyl phosphorylcholine) (PMPC) increases the hydration of the surfaces and decreases the wear of the substrates. Hence, a PMPC layer can potentially cushion the impact and improve the resistance of cross-linked polyethylene (CLPE). This study aimed to explore the fatigue and wear resistance of PMPC-grafted hydrated CLPE under impact-to-wear conditions using a
Cross-shear has been shown to increase ultra-high molecular weight polyethylene (UHMWPE) wear in
Introduction. Fretting crevice-corrosion (tribocorrosion) of metallic biomaterials is a major concern in orthopedic, spinal, dental and cardiovascular devices. 1. Stainless steel (i.e., 316L SS) is one alloy that sees extensive use in applications where fretting, crevices and corrosion may be present. While fretting-corrosion of this alloy has been somewhat studied, the concept of fretting-initiating crevice corrosion (FICC), where an initial fretting corrosion process leads to ongoing crevice-corrosion without continued fretting, is less understood. This study investigated the susceptibility of 316L SS to FICC and the role of applied potential on the process. The hypothesis is crevice-corrosion can be induced in 316L SS at potentials well below the pitting potential. Materials and Methods. A
Generic walking profiles applied to mechanical knee simulators are the gold standard in wear testing of total knee replacements (TKRs). Recently, there was a change in the international standard (ISO) for knee wear testing (ISO 14243-3): the direction of motion in the anterior/posterior (AP) and internal/external (IE) directions were reversed. The effects of this change have not been investigated, therefore it is not known whether results generated by following this new standard can be compared to historical wear tests which used the old standard. Using a finite element analysis (FEA) model of a TKR in parallel with an energy based wear model and adaptive remeshing, we investigated differences in wear between the newest ISO standard developed in 2014, and the previous ISO standard developed in 2004. CAD models of a left sided NexGen Cruciate Retaining (CR) TKR (Zimmer, Warsaw, IN) were used to create the FEA model (Figure 1). The loads and motions specified by simulator standards ISO 14243-3(2004) and ISO 14243-3(2014) were applied to the model. Analyses were run using ABAQUS v6.13-2 Standard (Dassault Systèmes, Waltham, MA). 8 node hexahedral elements were used to model the UHMWPE component. The contact was modeled as penalty contact, with the friction coefficient set to 0.04 on the articular surface. The cobalt chromium molybdenum femoral component was modeled as a rigid surface, utilizing a mix of 2. nd. order quadrilaterals and tetrahedrons. Wear of the polyethylene (PE) component was predicted to 1,000,000 cycles using a previously published frictional energy-based wear model. The wear model, developed from data generated in wheel-on-flat tests, utilizes two parameters defining the frictional energy required to remove a unit volume of material both parallel (3.86E8 J/mm. 3. ) and perpendicular (3.55E7 J/mm. 3. ) to the primary polyethylene fibril direction. Primary fibril direction for the analysis was set to the AP direction. Wear for each simulation of a gait cycle was scaled to 500,000 cycles. Two gait cycles were simulated representing 1,000,000 cycles in total. Adaptive remeshing was driven by the wear model, with the mesh being updated every time increment to simulate material ablation. The time step size was variable with a maximum of 0.01s. The FEA predicted higher wear rates for the newest ISO standard (7.34mg/million cycles) compared to the previous standard (6.04mg/million cycles) (Figure 2). Comparing the predicted wear scars (Figure 3), the new version of the standard covered a larger percentage of the total articular surface, with wear being more spread out as opposed to localized. This is more similar to what is seen in patient retrievals. The results of the study suggest that major differences between the old and the new ISO standard exist and therefore historical wear results are not comparable to newly obtained results. In addition, this study demonstrates the utility of FEA in wear analysis, though the wear model needs further work and validation before it can be used as a supplement to simulator testing. Validation of the wear model against simulator tests and
Background. Wear and fatigue damage to polyethylene components remain major factors leading to complications after total knee and unicompartmental arthroplasty. A number of wear simulations have been reported using mechanical test equipment as well as computer models. Computational models of knee wear have generally not replicated experimental wear under diverse conditions. This is partly because of the complexity of quantifying the effect of cross-shear at the articular interface and partly because the results of
Today's aging society is seeing an increase of patients with rheumatoid arthritis and osteoarthritis, as well as an increase in joint replacement surgery. The artificial joints used in this surgery frequently uses ultra-high molecular weight polyethylene (UHMWPE) as a bearing material. However, UHMWPE wear particles are considered to be a major factor in long-term osteolysis, and implant loosening. Many researchers have reported that the volume and size of particles are critical factors in macrophage activation, with particles in the size range of 0.1 – 1.0 μm being the most biologically active. The micro slurry-jet erosion (MSE) apparatus was introduced to minimize the amount of wear, and increase the size of UHMWPE wear particles by texturing the surfaces of Co-Cr-Mo alloy implants. The MSE apparatus uses a slurry of alumina particles (WA#8000: average diameter 1.2 μm) mixed with water. The slurry and compressed air are mixed within an injection nozzle, which is then applied to the Co-Cr-Mo alloy at high speed to achieve a desired nano-textured surface. In this study, four Co-Cr-Mo alloy surface profiles were prepared. The MSE injection nozzle was fed 40.0 mm in alternating directions across each surface with an orthogonal step of 0.5 mm. The surface M-1 was processed with an injection nozzle feed rate of 1.0 mm/s, and obtained a surface roughness of 5.7 nm. M-2 was processed with a feed rate of 2.0 mm/s, and had a surface roughness of 2.3 nm. The M-4 surface used a 40.0 mm alternating directions surface feed, but with a 1.0 mm orthogonal step, and an injection nozzle feed rate of 0.5 mm/s. It obtained a surface roughness of 4.0 nm. The G-1 surface, with a roughness of 10.0 nm, was processed with the typical lapping method, which is used in conventional artificial joints [Fig. 1]. A
Introduction. Ultra high molecular weight polyethylene (UHMWPE) has been used successfully as a bearing material in hip, knee, and shoulder joint replacements. However, there are problems to cause a failure in UHMWPE component, which are wear behavior and creep deformation. Continuous bearing motion and dynamic load have occurred to UHMWPE wear debris caused osteolysis in periprosthetic tissue and to plastic deformation of joint component, and subsequent aseptic loosening of components. Therefore, many studies have being carried out in order to reduce wear debris and to improve mechanical strength from UHMWPE, and there is tremendous improvement of mechanical property in UHMWPE from gamma irradiated conventional UHMWPE (GIPE), highly crosslinked PE (XLPE), and XLPE with vitamin E1, 2. Friction has a significant one of the factors effect on the wear and creep deformation. In this study, the short-term frictional behaviors of three typical types of GIPE, remelted XLPE (R-XLPE), and s annealed XLPE (A-XLPE), and XLPE with Vitamin E against Co-Cr alloy were compared under three levels of contact pressures which occured in hip, knee, and shoulder joints. Methods. Friction tests were conducted with UHMWPE against Co-Cr alloy by using
It have been reported that the wear volume of vitamin E-containing UHMWPE tested with a knee joint simulator was approximately 30% lower than that of virgin UHMWPE at 5 million cycles. However, the wear resistance mechanism of vitamin E-containing UHMWPE has not yet been clarified. The present study examines the effects of the addition of vitamin E on the frictional properties of ultra-high molecular weight polyethylene (UHMWPE) under several different load and serum conditions. Friction tests were carried out using a computer-controlled
Introduction. First-generation annealed HXLPE has been clinically successful at reducing both clinical wear rates and the incidence of osteolysis in total hip arthroplasty. However, studies have observed oxidative and mechanical degradation occurring in annealed HXLPE. Thus, it is unclear whether the favorable clinical performance of 1st generation HXLPE is due to the preservation of bearing surface tribological properties or, at least partially, to the reduction in patient activity. The purpose of this study was to evaluate the in vitro wear performance (assessed using multidirectional
Statement of Purpose:. The wear rate of Ultra High Molecular Weight Polyethylene (UHMWPE) in joint replacements has been correlated to both contact area and contact stress in the literature, [1], [2]. In both publications and our experiment, UHMWPE articulated with a polished surface of cobalt-chromium alloy was evaluated using a
Third-body wear is believed to be one trigger for adverse results
with metal-on-metal (MOM) bearings. Impingement and subluxation
may release metal particles from MOM replacements. We therefore
challenged MOM bearings with relevant debris types of cobalt–chrome
alloy (CoCr), titanium alloy (Ti6Al4V) and polymethylmethacrylate
bone cement (PMMA). Cement flakes (PMMA), CoCr and Ti6Al4V particles (size range
5 µm to 400 µm) were run in a MOM wear simulation. Debris allotments
(5 mg) were inserted at ten intervals during the five million cycle
(5 Mc) test. Objectives
Methods
Wear of polyethylene is associated with aseptic loosening of orthopaedic implants and has been observed in hip and knee prostheses and anatomical implants for the shoulder. The reversed shoulder prostheses have not been assessed as yet. We investigated the volumetric polyethylene wear of the reversed and anatomical Aequalis shoulder prostheses using a mathematical musculoskeletal model. Movement and joint stability were achieved by EMG-controlled activation of the muscles. A non-constant wear factor was considered. Simulated activities of daily living were estimated from After one year of use, the volumetric wear was 8.4 mm3 for the anatomical prosthesis, but 44.6 mm3 for the reversed version. For the anatomical prosthesis the predictions for contact pressure and wear were consistent with biomechanical and clinical data. The abrasive wear of the polyethylene in reversed prostheses should not be underestimated, and further analysis, both experimental and clinical, is required.