Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_16 | Pages 31 - 31
17 Nov 2023
Warren J Cowie R Jennings L Wilcox R Fermor H
Full Access

Abstract

Objectives

The aim of this study was to develop an in vitro GAG-depleted patella model and assess the biomechanical effects following treatment with a SAP:CS self-assembling hydrogel.

Methods

Porcine patellae (4–6 month old) were harvested and subject to 0.1% (w/v) sodium dodecyl sulfate (SDS) washes to remove GAGs from the cartilage. Patellae were GAG depleted and then treated by injection with SAP (∼ 6 mM) and CS (10 mg) in Ringer's solution through a 30G needle. Native, GAG depleted and SAP:CS treated patellae were tested through static indentation testing, using 15g load, 5mm indenter over 1hr period. The degree of deformation of each group was assessed and compared (Mann-Whitney, p<0.05). Native, GAG depleted, sham (saline only) and SAP:CS treated paired patellae and femurs were additionally characterized tribologically through sequential wear testing when undergoing a walking gait profile (n=6 per group). The cartilage surfaces were assessed and compared (Mann-Whitney, p<0.05) using the ICRS scoring system, surface damage was illustrated through the application of Indian ink.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_9 | Pages 58 - 58
17 Apr 2023
McCall B Cowie R Jennings L
Full Access

The clinical success of osteochondral autografts is heavily reliant on their mechanical stability, as grafts which protrude above or subside below the native cartilage can have a negative effect on the tribological properties of the joint [1]. Furthermore, high insertion forces have previously been shown to reduce chondrocyte viability [2]. Commercial grafting kits may include a dilation tool to increase the diameter of the recipient site prior to insertion. The aim of this study was to evaluate the influence of dilation on the primary stability of autografts.

Six human cadaveric femurs were studied. For each femur, four 8.5 × 8mm autografts were harvested from the trochlear groove and implanted into the femoral condyles using a Smith & Nephew Osteochondral grafting kit. Two grafts were implanted into dilated recipient sites (n=12) and two were implanted with no dilation (n=12). Insertion force was measured by partially inserting the graft and applying a load at a rate of 1 mm/min, until the graft was flush with the surrounding cartilage. Push-in force was measured by applying the same load, until the graft had subsided 4mm below congruency. Significance was taken as (p<0.05).

Average maximum insertion force of dilated grafts was significantly lower (p<0.001) than their non-dilated equivalent [28.2N & 176.7N respectively]. There was no significant difference between average maximum push-in force between the dilated and non-dilated groups [1062.8N & 1204.2N respectively].

This study demonstrated that significantly less force is required to insert dilated autografts, potentially minimising loss of chondrocyte viability. However, once inserted, the force required to displace the grafts below congruency remained similar, indicating a similar degree of graft stability between both groups.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_6 | Pages 8 - 8
1 Apr 2018
Cowie R Briscoe A Fisher J Jennings L
Full Access

Introduction

PEEK-OPTIMA™ has been considered as an alternative bearing material to cobalt chrome in the femoral component of total knee replacements. To better understand the tribology of UHMWPE-on-PEEK-OPTIMA™ and to find the most appropriate environmental conditions under which to test this novel bearing material combination, a series of tests under different protein lubricant concentrations at rig (∼24°C) and elevated temperature (∼35°C) were carried out in simple geometry wear and friction rigs. Under all conditions, the wear of UHMWPE-on-PEEK-OPTIMA™ was compared to UHMWPE-on-cobalt chrome (CoCr).

Methods

The pins used were GUR1020 UHMWPE (conventional, non-sterile) and the plate material was either polished CoCr (Ra<0.01µm) or PEEK-OPTIMA (Ra∼0.03µm) provided by Invibio Ltd, UK. The wear simulation was carried out in a six station reciprocating rig. The kinematic conditions were consistent for all tests and reflected the average cross shear and contact pressure (3.2MPa) in a total knee replacement. Tests were carried out at either rig running temperature (∼24°C) or at elevated temperature (∼35°C) and in varying protein lubricant concentrations (0, 2, 5, 25 and 90%). Wear of the UHMWPE pins was determined by gravimetric analysis. The pin-on-plate friction rig study was carried out at rig temperature in 0, 2, 5, 25 and 90% serum and reflected the contact pressure used in the wear tests. Measurements were taken using a piezoelectric sensor and the steady state friction derived. At least 3 repeats were taken for each study, statistical analysis carried out using ANOVA with significance taken at p<0.05.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_6 | Pages 7 - 7
1 Apr 2018
Cowie R Briscoe A Fisher J Jennings L
Full Access

Introduction

Experimental wear simulation of an all-polymer knee implant has shown an equivalent rate of wear of UHMWPE tibial components against PEEK-OPTIMA™ and cobalt chrome femoral components of a similar initial geometry and surface topography. However, when the patella is resurfaced with an UHMWPE patella button, it is important to also ascertain the wear of the patella. Wear debris from the patella contributes to the total volume of wear debris produced by the implant which should be minimised to reduce the potential for osteolysis and subsequent implant loosening. The aim of this study was to investigate the wear of the patellofemoral joint in an all-polymer knee implant. The wear of UHMWPE patellae articulating against PEEK-OPTIMA™ femoral components was compared to UHMWPE articulating against cobalt chrome femoral components.

Materials and Methods

Six mid-size (size C) PEEK-OPTIMA™ femoral components (Invibio Knee Ltd., UK) and six cobalt chrome femoral components of similar initial surface topography and geometry were coupled with 28mm all-polyethylene GUR1020 patellae (conventional, EO sterile) (Maxx Orthopaedics, USA). The implants were set up in a ProSim 6 station electromechanical knee simulator (Simulation Solutions, UK) which was modified for testing the patellofemoral joint. 3 million cycles (MC) of wear simulation was carried out under kinematics aiming to replicate a gait cycle adapted for an electromechanical simulator from previous work by Maiti et al. The simulator used has six degrees of freedom of which four were controlled; axial force up to 1200N, flexion/extension 22°, superior-inferior (SI) displacement (22mm) and Abduction-adduction (AA) (4°). The SI and AA were displacement controlled and driven through the patella. The medial-lateral displacement and tilt (internal/external rotation) of the patella were passive so the patella button was free to track the trochlear groove. The lubricant used was 25% bovine serum supplemented with 0.03% sodium azide to retard bacterial growth. The wear of patellae was determined gravimetrically with unloaded soak controls used to compensate for the uptake of moisture by the UHMWPE. The mean wear rate ± 95% confidence limits were calculated and statistical analysis was carried out using ANOVA with significance taken at p<0.05.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 83 - 83
1 Feb 2017
Cowie R Aiken S Cooper J Jennings L
Full Access

Introduction

Calcium sulfate bone void fillers (CS-BVF) are increasingly being used for dead space management in infected arthroplasty revision surgery. The use of loose beads of CS-BVF close to the articulating surfaces of an implant means there is potential for them to migrate between the articulating surfaces acting as a third body particle. The aim of this study was to investigate the influence of CS-BVF on the third body wear of total knee replacements.

Methods

The influence of CS-BVF on wear was investigated using the commercially available CS-BVF ‘Stimulan’ (Biocomposites Ltd., UK) and posterior stabilised U2 total knee replacement system implants (United Orthopaedic Corp., Taiwan). The experimental wear simulation was performed using a six station ProSim electropneumatic knee simulator (Simulation Solutions, UK) running the Leeds intermediate kinematics input profile [1]. To investigate the damage that could be caused by the third body particles, 5cc of CS-BVF beads (excess) were placed on the tibial component of the implant, the simulator was run dry for 60 cycles before adding lubricant (25% bovine serum supplemented with 0.03% sodium azide) and running for an additional 115,000 cycles representative of the 6–8 weeks the CS-BVF are present in the body prior to their resorption. The surface topography of the cobalt chrome femorals was analysed using contacting profilometry to ascertain whether the third body particles of CS-BVF had damaged the surfaces. To investigate the influence of CS-BVF on the third body wear of the UHMWPE tibials, 3 million cycles (MC) of wear simulation was subsequently carried out. The wear of the UHMWPE tibials was assessed gravimetrically and the wear of implants tested with CS-BVF was compared to the wear against negative controls (initial Ra∼0.02µm) and positive controls (initial Ra ∼0.4µm) damaged with a diamond stylus. N=6 was completed for each condition, statistical analysis was carried out using ANOVA with significance taken at p<0.05.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 94 - 94
1 May 2016
Cowie R Briscoe A Fisher J Jennings L
Full Access

Introduction

There is a demand for longer lasting arthroplasty implants driving the investigation of novel material combinations. PEEK has shown promise as an arthroplasty bearing material, with potentially relatively bio inert wear debris [1]. When coupled with an all-polyethylene tibial component this combination shows potential as a metal-free knee. In this study, the suitability of PEEK Optima® as an alternative to cobalt chrome for the femoral component of total knee replacements was assessed using experimental knee wear simulation under two kinematic conditions.

Methods

Three cobalt chrome and three injection moulded PEEK Optima® (Invibio Biomaterial Solutions, UK) femoral components of similar geometry and surface roughness (mean surface roughness (Ra) ∼0.02µm) were coupled with all-polyethylene GUR1020 (conventional, unsterilised) tibial components in a 6 station ProSim knee simulator (Simulation Solutions, UK). 3 million cycles (MC) of wear simulation were carried out under intermediate kinematics (maximum anterior-posterior (AP) displacement 5mm) followed by 3MC under high kinematics (AP 10mm) [2] with 25% serum as the lubricant. The wear of the tibial component was assessed gravimetrically. At each measurement point, the surface roughness of the femoral components was determined using contacting profilometry and throughout testing, the bulk lubricant temperature was monitored close to the articulating surfaces.

Statistical analysis was carried out using ANOVA, with significance at p<0.05.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 100 - 100
1 Jan 2016
Cowie R Briscoe A Fisher J Jennings L
Full Access

Introduction

UHMWPE articulating against PEEK-OPTIMA® has the potential for use as a novel bearing couple in joint arthroplasty due to its potentially low wear rates and the bioinertness of its wear debris. The aim of this study was to investigate the role of protein in the lubricant on the wear of UHMWPE articulating against PEEK at both room and physiological temperature.

Methods

The wear of GUR1020 UHMWPE pins articulating against PEEK plates (Ra ∼0.06µm) was compared to highly polished cobalt chrome plates (Ra <0.01µm) in a 6-station multi-axial pin-on-plate rig using kinematics to replicate those in total knee arthroplasty. Tests were carried out at either ∼20°C or ∼36°C and wear was investigated under varying concentrations of bovine serum (0, 25 or 90%). Studies were carried out for 1 Million cycles with wear of the UHMWPE pins assessed gravimetrically using unloaded soak controls to compensate for moisture uptake.

Statistical analysis was carried out using ANOVA with significance taken at p<0.05.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 99 - 99
1 Jan 2016
Cowie R Carbone S Aiken S Cooper J Fisher J Jennings L
Full Access

Introduction

When third body particles originating from bone cement or bone void fillers become trapped between articulating surfaces of joint replacements, contact surfaces may be damaged leading to accelerated wear and premature failure of the implant. In this study, the damage to cobalt chrome counterfaces by third body particles from PMMA bone cement (GMV, DePuy) and various bone void fillers was investigated; then wear tests of UHMWPE were carried out against these surfaces.

Methods

Third body particles of polymerised GMV bone cement and the bone void fillers; OsteoSet (with tobramycin), Stimulan and Stimulan+ (with vancomycin and tobramycin) (provided by Biocomposites Ltd.) were trapped between an UHMWPE pin and a highly polished cobalt chrome plate. A load of 120N was applied to the pin and using an Instron materials testing machine, the plate was pulled beneath the pin to recreate third body damage [1]. The resulting surface topography of the plate was analysed using white light interferometry (Bruker NPFLEX). Pin on plate wear tests of GUR 1020 UHMWPE pins were carried out against the plates perpendicular to the direction of damage for 500,000 cycles in 25% bovine serum using a 6-station multi-axial reciprocating rig under conditions to replicate the kinematics in total knee replacement. Wear of the pins was determined by gravimetric analysis and results were compared to negative (highly polished) control plates and positive controls scratched with a diamond stylus (lip height 2µm). Statistical analysis was carried out using one-way ANOVA with significance taken at p<0.05.


The Journal of Bone & Joint Surgery British Volume
Vol. 37-B, Issue 1 | Pages 112 - 116
1 Feb 1955
Cowie R