Highly cross-linked polyethylene (HXLPE) greatly reduces wear in total hip arthroplasty, compared to conventional polyethylene (CPE). Cross-linking is commonly achieved by irradiation. This study aimed to compare the degree of cross-linking and in vitro wear rates across a cohort of retrieved and unused polyethylene cups/liners from various brands. Polyethylene acetabular cups/liners were collected at one centre from 1 April 2021 to 30 April 2022. The trans-vinylene index (TVI) and oxidation index (OI) were determined by Fourier-transform infrared spectrometry. Wear was measured using a pin-on-disk test.Aims
Methods
Both total joint arthroplasty (TJA) and Alzheimer's Disease (AD) are prevalent in elderly populations. It is the goal of this study to determine if the presence of implant metals originating from TJA correlates with the onset with higher implant metal content in the brain and AD pathology. Tissue samples from four brain regions of 701 (229 with TJA) participants from an ongoing longitudinal cohort study (Rush Memory and Aging Project) was analyzed including the inferior-temporal-cortex (ITC), which is associated with early onset of AD. Implant metal (Co, Cr, Mo, Ti, Al) content was determined by ICP-MS. Comparisons were conducted between the no-TJA-group and a TJA group. Due to the higher likelihood of Co release the TJA group was further differentiated in a THA (N=146) and a TKA/TSA (N=83) group. Diffuse and neuritic amyloid plaques and phosphorylated tau were assessed and summarized as standard measures of AD pathology. We used separate linear regression models adjusted for age, sex, education, and APOɛ4-status for the associations of all metals (log-transformed) with global AD pathology, amyloid plaques, and phosphorylated tau. The THA group had higher cobalt content across all brain regions (p=0.003) and within the ITC (p=0.051) compared to the no-TJA group, whereas the TKA/TSA group did not. Across all tissue samples, Co was associated with higher amyloid load (β=0.35, p=0.027), phosphorylated tau (β=0.47, p=0.011), and global AD pathology (β=0.19, 0.0004) in the ITC. The presence of TJA itself was not associated with AD pathology. We showed that only Co content was higher within the ITC in persons with THA. We found among all tested metals that Co was consistently associated with AD pathology. Although we found an association of cobalt with AD pathology, the cross-sectional nature of this study does not allow the determination of cause and effect.
Ti-6Al-4V is the most common alloy used for orthopaedic implants. Its popularity is due to low density, superior corrosion resistance, good osseointegration and lower elastic modulus when compared to other commonly used alloys such as CoCrMo and stainless steel. In fact, the use of Ti64 has even further increased lately since recent controversy around adverse local tissue reactions and implant failure related to taper corrosion of CoCrMo alloy. However, implants made from Ti64 can fail in some cases due to fatigue fracture, sometimes related to oxide induced stress corrosion cracking or hydrogen embrittlement, or preferential corrosion of the beta phase. Studies performed with Ti-6Al-4V do often not consider that the alloy itself may have a range of characteristics that can vary and could significantly impact the implant properties. These variations are related to the material microstructure which depends not only on chemical composition, but also the manufacturing process and subsequent heat treatments. Different microstructures can occur in implants made form wrought alloys, cast alloys, and more recently, additive manufactured (AM) alloys. Implant alloy microstructure drives mechanical and electrochemical properties. Therefore, this study aims to analyse the microstructure of Ti-6Al-4V alloy of additive manufactured and conventional retrieved orthopaedic implants such as acetabular cups, tibial trays, femoral stem and modular neck by means of electron backscatter diffraction (EBSD). Microstructural features of interest include grains shape and size, phase content and distribution, preferred grain orientation (texture), alloying elements distribution (homogenization) and presence of impurities. Additionally, we demonstrate the direct impact of different microstructural features on hardness. We analysed 17 conventional devices from 6 different manufacturers, 3 additive manufactured devices from 2 different manufactures and 1 control alloy (bar stock). The preliminary results showed that even though all implants have the same chemical composition, their microstructural characteristics vary broadly. Ti64 microstructure of conventional alloys could be categorized in 3 groups: equiaxed grains alloys (Fine and Coarse), bimodal alloys and dendritic alloys. The additive manufactured implants were classified in an additional group on its own which consists of a needle-like microstructures - similar to Widmanstätten patterns, Fig. 1, with a network of β phase along α phase grains. Furthermore, AM alloys exhibited residual grain boundaries from the original β grains from the early stage of the solidification process, Fig. 2. These characteristics may have implication on the fatigue and corrosion behaviour. In addition, it we observed inhomogeneous alloying element distribution in some cases, Fig. 3, especially for the additive manufactured alloys, which also may have consequences on corrosion behaviour. Finally, the hardness testing revealed that the implants with large grain size, such as AM alloys, exhibit low hardness values, as expected, but also the amount of beta phase correlated positively with lower hardness. Grain aspect ratio and beta phase grain size correlated positively with higher hardness. In summary, we found that common Ti64 implants can exhibit a broad variety of different alloy microstructures and the advent of AM alloys introduces an entirely new category. It is imperative to determine the ideal microstructure for specific applications.
Total hip replacement failure due to fretting-corrosion remains a clinical concern. We recently described that damage within CoCrMo femoral heads can occur either by mechanically-dominated fretting processes leading to imprinting (via rough trunnions) and surface fretting (via smooth trunnions), or by a chemically-dominated etching process along preferential corrosion sites, termed “column damage”. These corrosion sites occur due to banding of the alloy microstructure. Banding is likely caused during thermo-mechanical processing of the alloy and is characterized by local molybdenum depletion. It was the objective of this study to quantify material loss from femoral heads with severe corrosion, identify the underlying damage modes, and to correlate the damage to the alloy's microstructure. 105 femoral heads with a Goldberg score 4 were evaluated. Coordinate measuring machine data was used to compute material loss and visualize damage features. Time Introduction
Methods
Little is known about the relationship between head-neck corrosion and its effect on the periprosthetic tissues and distant organs of patients hosting well-functioning devices. The purpose of this study was to investigate in postmortem retrieved specimens the degree and type of taper damage, and the corresponding histologic responses in periprosthetic tissues and distant organs. Fifty postmortem THRs (34 primaries, 16 revisions) retrieved after 0.5 to 26 years were analyzed. Forty-three implants had a CoCrMo stem and seven had a Ti6Al4V stem. All heads were CoCrMo and articulated against polyethylene cups (19 XLPE, 31 UHMWPE). H&E sections of joint pseudocapsules, liver, spleen, kidneys and lymph nodes were graded 1–4 for the intensity of various inflammatory cell infiltrates and tissue characteristics. Corrosion damage of the taper surfaces was assessed using visual scoring and quantitated with an optical coordinate measuring machine. SEM analysis was used to determine the acting corrosion mode. Polyethylene wear was assessed optically. The majority of tapers had minimal to mild damage characterized by local plastic deformation of machining line peaks. Imprinting of the stem topography onto the head taper surface was observed in 18 cases. Column damage on the head taper surface occurred in three cases. All taper surfaces scored moderate or severe exhibited local damage features of fretting and/or pitting corrosion. Moderate or severe corrosion of the head and/or trunnion was present in nine hips. In one asymptomatic patient with bilateral hips, lymphocyte-dominated tissue reactions involving perivascular infiltrates of lymphocytes and plasmacytes were observed. In this patient, mild, focal lymphocytic infiltrates were also present in the liver and kidneys, and there was focal histiocytosis and necrosis of the para-aortic lymph nodes. These two implants, which had been in place for 58.6 and 60.1 months, had severe intergranular corrosion of the CoCrMo trunnion, and column damage and imprinting on the head taper. In the other 41 hips, macrophage responses in the joint pseudocapsule to metallic and/or polyethylene wear particles ranged widely from minimal to marked. Focal necrosis in the pseudocapsules of 12 arthroplasties was related to high concentrations of CoCrMo, TiAl4V, TiO, BaSO4 and polyethylene wear particles. High concentrations of these particles were also detected in para-aortic lymph nodes. Rare to mild macrophages were observed in liver and spleen. This is a comprehensive study of wear and corrosion within well-functioning postmortem retrieved THRs, and the resulting local and distant tissue reactions. One of eight patients with moderate or severe corrosion did have a subclinical inflammatory response dominated by lymphocytes after five years. To what extent such an inflammatory process might progress to become symptomatic is not known. Ionic and particulate products generated by corrosion disseminated systemically. The minor lymphocytic infiltrate in the liver and kidneys of one subject with bilateral severely corroded head-neck junctions might suggest possible metal toxicity. The diagnosis of adverse tissue reactions to corrosion of modular junctions can be challenging. Postmortem retrieval studies add to our understanding of the nature and progression of lymphocyte-dominated adverse local and potentially systemic tissue reactions to corrosion of modular junctions.
The aim of this study was to develop a novel computational model for estimating head/stem taper mechanics during different simulated assembly conditions. Finite element models of generic cobalt-chromium (CoCr) heads on a titanium stem taper were developed and driven using dynamic assembly loads collected from clinicians. To verify contact mechanics at the taper interface, comparisons of deformed microgroove characteristics (height and width of microgrooves) were made between model estimates with those measured from five retrieved implants. Additionally, these models were used to assess the role of assembly technique—one-hit versus three-hits—on the taper interlock mechanical behaviour.Aims
Methods
Total hip arthroplasty (THA) is a commonly performed procedure to relieve arthritis or traumatic injury. However, implant failure can occur from implant loosening or crevice corrosion as a result of inadequate seating of the femoral head onto the stem during implantation. There is no consensus—either by manufacturers or by the surgical community—on what head/stem assembly procedure should be used to maximize modular junction stability. Furthermore, the role of “off-axis” loads—loads not aligned with the stem taper axis—during assembly may significantly affect modular junction stability, but has not been sufficiently evaluated. The objective of this study was to measure the three-dimensional (3D) head/stem assembly loads considering material choice—metal or ceramic—and the surgeon experience level.Introduction
Objective
Improper seating during head/stem assembly can lead to unintended micromotion between the femoral head and stem taper—resulting in fretting corrosion and implant failure. There is no consensus—either by manufacturers or by the surgical community—on what head/stem taper assembly method maximizes modular junction stability in total hip arthroplasty (THA). A 2018 clinical survey found that orthopedic surgeons prefer applying one strike or three, subsequent strikes when assembling head/stem taper. However, it has been suggested that additional strikes may lead to decreased interference. Additionally, the taper surface finish—micro-grooves—has been shown to affect taper interference and may be influenced by assembly method. The objective of this study was to employ a novel, micro-grooved finite element (FEA) model of the hip taper interface and assess the role of head/stem assembly method—one vs three strikes—on modular taper junction stability.Introduction
Objective
The combined incidence of anatomic (aTSA) and reverse total shoulder arthroplasties (rTSA) in the US is 90,000 per annum and rising. There has been little attention given to potential long-term complications due to periprosthetic tissue reactions to implant debris. The shoulder has been felt to be relatively immune to these complications due to lower acting loads compared to other joint arthroplasties. In this study, retrieved aTSAs and rTSAs were examined to determine the extent of implant damage and to characterize the nature of the corresponding periprosthetic tissue responses. TSA components and periprosthetic tissues were retrieved from 23 (eleven aTSA, twelve rTSA). Damage to the implants was characterized using light microscopy. Head/stem taper junction damage was graded 1–4 as minimal, mild, moderate or marked. Damage on polyethylene (PE) and metal bearing surfaces was graded 1–3 (mild, moderate, marked). H&E stained sections of periprosthetic soft tissues were evaluated for the extent and type of cellular response. A semi-quantitative system was used to score (1=rare to 4=marked) the overall number of particle-laden macrophages, foreign body giant cells, lymphocytes, plasma cells, eosinophils, and neutrophils. Implant damage and histopathological patterns were compared between the two TSA groups using the Mann-Whitney and Spearman tests.Introduction
Methods
Wear and corrosion debris generated from total hip replacements (THR) can cause adverse local tissue reactions (ALTR) or osteolysis, often leading to premature implant failure. The tissue response can be best characterized by histopathological analysis, which accurately determines the presence of cell types, but is limited in the characterization of biochemical changes (e.g. protein conformation alteration). Fourier transform infrared micro-spectroscopy imaging (FTIRI) enables rapid analysis of the chemical structure of biological tissue with a high spatial resolution, and minimal additional sample preparation. The data provides the most information through multivariate method carried out by hierarchical clustering analysis (HCA). It is the goal of this study to demonstrate the beneficial use of this multivariate approach in providing pathologist with biochemical information from cellular and subcellular organization within joint capsule tissue retrieved from THR patients. Joint capsule tissue from 2 retrieved THRs was studied.
The results demonstrate that multivariate FTIRI based spectral histopathology is a powerful tool to characterize the chemical structure and foreign body response within periprosthetic tissue, thus providing insights into the biological impact of different types of implant debris. For any figures or tables, please contact the authors directly.
Modular junctions in total hip replacement (THR) have been a primary source of fretting and corrosion which can lead to implant failure. Fretting is a result of unintended micromotion between the femoral head and stem tapers and is suspected to result after improper taper seating during assembly. Two design factors known to influence in-vitro taper assembly mechanics are relative taper alignment—mismatch angle—and the surface finish—micro-grooves. However, these factors have not been systematically evaluated together. The objective of this study was to employ a novel, micro-grooved finite element (FEA) model of the hip taper interface and assess the role of taper mismatch angle and taper surface finish—smooth and rough—on the modular junction mechanics during assembly.Introduction
Objective
Improper seating during head/stem assembly can lead to unintended micromotion between the femoral head and stem taper—resulting in fretting corrosion and implant failure.1 There is no consensus—either by manufacturers or by the surgical community—on what head/stem taper assembly method maximizes modular junction stability in total hip arthroplasty (THA). A 2018 clinical survey2 found that orthopedic surgeons prefer applying one strike or three, subsequent strikes when assembling head/stem taper. However, it has been suggested that additional strikes may lead to decreased interference strength. Additionally, the taper surface finish—micro-grooves—has been shown to affect taper interference strength and may be influenced by assembly method. The objective of this study was to employ a novel, micro-grooved finite element (FEA) model of the hip taper interface and assess the role of head/stem assembly method—one vs three strikes—on modular taper junction stability. A two-dimensional, axisymmetric FEA model representative of a CoCrMo femoral head taper and Ti6Al4V stem taper was created using median geometrical measurements taken from over 100 retrieved implants.3 Surface finish—micro-grooves—of the head/stem taper were modeled using a sinusoidal function with amplitude and period corresponding to retrieval measurements of micro-groove height and spacing, respectively. Two stem taper micro-groove geometries— “rough” and “smooth”—were modeled corresponding to the median and 5th percentile height and spacing measurements from retrievals. All models had a 3' (0.05°), proximal-locked angular mismatch between the tapers. To simulate implant assembly during surgery, multiple dynamic loads (4kN, 8kN, and 12kN) were applied to the femoral head taper in a sequence of one or three strikes. The input load profile (Figure 1) used for both cases was collected from surgeons assembling an experimental setup with a three-dimensional load sensor. Models were assembled and meshed in ABAQUS Standard (v 6.17) using four-node linear hexahedral, reduced integration elements. Friction was modeled between the stem and head taper using surface-to-surface formulation with penalty contact (µ=0.2). A total of 12 implicit, dynamic simulations (3 loads × 2 assembly sequences × 2 stem taper surface finishes) were run, with 2 static simulations at 4kN for evaluating inertial effects. Outcome variables included contact area, contact pressure, equivalent plastic strain, and pull-off force.Introduction
Methods
The lifetime of total hip replacements (THR) is often limited by adverse local tissue reactions to corrosion products generated from modular junctions. Two prominent damage modes are the imprinting of the rougher stem topography into the smoother head taper topography (imprinting) and the occurrence of column-like troughs running parallel to the taper axis (column damage). It was the purpose of this study to identify mechanisms that lead to imprinting and column damage based on a thorough analysis of retrieved implants. 776 femoral heads were studied. Heads were visually inspected for imprinting and column damage. Molds were made of each head taper and scanned with an optical coordinate measuring machine. The resulting intensity images were used to visualize damage on the entire surface. In selected cases, implant surfaces were further analyzed by means of scanning electron microscopy (SEM) and white light interferometry. The alloy microstructure was characterized for designs from different manufactures.INTRODUCTION
METHODS
Little is known about the relationship between head-neck corrosion and its effect on periprosthetic tissues and distant organs in the majority of patients hosting apparently well-functioning devices. We studied the degree and type of taper damage and the histopathologic response in periprosthetic tissue and distant organs. A total of 50 contemporary THRs (34 primary, 16 revision) retrieved postmortem from 40 patients after 0.4–26 years were studied. Forty-three femoral stems were CoCrMo and 7 were Ti6Al4V. In every case, a CoCrMo-alloy head articulated against a cementless polyethylene cup (19 XLPE and 31 UHMWPE). H&E and IHC sections of the joint pseudocapsules and liver were graded 1–4 for the intensity of various inflammatory cell infiltrates and tissue necrosis. The nature of the tissue response in the joint capsule, liver, spleen, kidneys and lymph nodes was assessed. Wear and corrosion products in the tissues were identified using SEM and EDS. Taper surfaces were graded for corrosion damage using modified Goldberg scoring and examined by SEM to determine the acting corrosion mode. Correlations between damage scores and the histologic variables were generated using the Spearman test.Introduction
Methods
Recently, corrosion at the head-neck junction in metal-on-polyethylene bearing total hip arthroplasty (THA) has been recognized as a cause of adverse local tissue reactions (ALTR). Serum metal levels have been advocated as a tool for the diagnosis of ALTR, however no prior studies have specifically examined their utility. The purpose of this study was to determine the optimal cut-off values for serum cobalt and chromium in diagnosing ALTR after metal-on-polyethylene bearing THA. We reviewed 447 consecutive patients with serum metal levels tested at our institution and identified 62 with a metal-on-polyethylene bearing who had axial imaging or underwent reoperation to confirm the presence or absence of ALTR. Receiver operating characteristic curves were produced to identify cut-off thresholds to optimize sensitivity and diagnostic test performance was characterized. 42 Of the 62 patients (66%) were positive for an ALTR. The best test for the diagnosis of ALTR was the serum cobalt level (area under the curve [AUC]=99%). A threshold cut-off of ≥ 1.0 ng/ml had a sensitivity of 100%, specificity of 90%, positive predictive value (PPV) of 96%, and negative predictive value (NPV) of 100%. Serum chromium levels were also diagnostic (AUC=87%). A threshold cut-off of ≥ 0.15 ng/ml had a sensitivity of 100%, specificity of 50%, PPV of 81%, and NPV of 100%. Finally, serum cobalt to chromium ratio was also helpful for diagnosis (AUC=90%). A threshold cut-off of 1.4 for the cobalt to chromium ratio offered a sensitivity of 93%, specificity of 70%, PPV of 87%, and NPV of 82%. Measurement of serum cobalt with a threshold value of 1.0 ng/ml in our experience is the best test for identifying the presence of ALTR in patients with a metal-on-polyethylene THA. Measurement of chromium and the ratio of cobalt to chromium are also of value.
Wear debris from polyethylene tibial inserts has been associated with limited longevity of total knee replacements (TKRs). While material factors were studied extensively and considerable progress has been made, there is little knowledge about surgical factors, particularly on how the wear rate is related to implant positioning. It was the purpose of this study to determine the combined effect of patient and implant positioning factors on the volumetric wear rate of TKRs. Our hypothesis was that implant alignment has a significant impact on the wear rate when controlled for other patient factors. This study included 59 tibial inserts of a cruciate retaining TKR design (Nexgen, Zimmer Inc.). The patients' age, sex, weight, height, and implant size were obtained. All implants were scanned with a coordinate measuring machine. Volumetric wear was determined using an autonomous mathematical reconstruction method (Figure 1). Radiographs were used to determine the anatomic lateral distal femoral angle (aLDFA), anatomic medial proximal tibial angle (aMPTA), femoral tilt angle (FTA) and posterior tibial slope (PTS). Also, the patella position was assessed using the Blackburne-Peel Index (BPI) and the Insall-Salvati Ratio (Figure 2). General linear modeling (SPSS) was conducted in order to determine the most significant patient and implant positioning factors on wear rate.Introduction
Methods
There are increasing reports of total hip replacement (THR) failure due to corrosion within modular taper junctions, and subsequent adverse local tissue reactions (ALTRs) to corrosion products. Modular junction corrosion is a multifactorial problem that depends on material, design, patient and surgical factors. However, the influence of alloy microstructure on corrosion has not been studied sufficiently. Especially for cast CoCrMo, there are concerns regarding microstructure variability with respect to grain size and hard-phase volume fraction. Therefore, it was the goal of this study to (1) identify different types of microstructures in contemporary implants, and (2) determine implications of alloy microstructure on the occurring corrosion modes. Fifteen surgically retrieved femoral stems made from cast CoCrMo alloy were analyzed for this study. Damage on the taper surfaces was investigated by scanning electron microscopy (SEM) and damage was assessed with the Goldberg Score. The alloy microstructure was evaluated by standard metallographic techniques. Alloy samples were sectioned off the femoral stem, and microstructural features were visualized by chemical etching. Cyclic potentio-dynamic polarization tests were carried out with alloy samples from two implants with different commonly occurring types of microstructures. Both had a similar grain size, but type 1 had no hard-phases, where as type 2 exhibited hard-phases along the grain boundaries, as well as intra-granular hard-phase clusters. Tests were performed in bovine serum at 37°C with a saturate calomel reference electrode and a graphite counter electrode. In vitro generated corrosion damage was then compared to in vivo generated damage features on the taper surfaces of the corresponding implants.Introduction
Methods
Failure of total knee replacements due to the generation of polyethylene wear debris remains a crucial issue in orthopedics. Unlike the hip, it is difficult to accurately determine knee implant wear rates from retrieved components. Several studies have relied on thickness measurements to estimate penetration, but the complicated geometry of contemporary tibial liners poses a challenge to accurately assess wear. In this study we address the question whether linear penetration can serve as a surrogate measure for volumetric material loss. Eighty-one retrieved UHMWPE NexGen cruciate-retaining tibial liners (Zimmer, Warsaw, IN) with an average time in situ of 5.27±2.89 years were included in the study. Metrology data for the surfaces of the tibial liners were obtained with a coordinate measuring machine (OGP, Rochester, NY). Using a laser scanner with two micrometer depth accuracy, at least 400,000 measurement points were taken by investigator #1. Areal thickness changes were mapped for the lateral and medial sides with the help of an autonomous mathematical reconstruction algorithm and volume loss was calculated based on wear scar area and local thickness change. Investigator #2, blinded from these results, measured the minimum thickness of the medial and lateral tibial plateau using a dial indicator with a spherical tip radius of 3mm. Twenty-three short term retrievals (3 to 4 per implant size), removed due to infection and without any signs of wear, served as “unused” reference. Linear penetration was then calculated by subtracting the minimum thickness of each plateau from the average thickness of the reference components.Introduction
Methods
Current pre-clinical testing is performed using knee wear simulators with standardized walking profiles. Differences in generated damage patterns to those observed on retrieved liners have been explained with the absence of activities other than walking, less severe loading conditions, and a discrepancy in the simulator's tibiofemoral contact mechanics and in vivo knee excursion. While it has been recognized that rotational alignment of the knee may also drive the location and shape of wear scars, to the best of our knowledge this parameter has not been investigated in knee simulator studies. Here, we use patient specific gait as input to the simulation to approximate the patient specific contact mechanics. Kinematic and kinetic input data was obtained from gait analysis of a patient with a MGII (Zimmer Inc.) prosthesis at 11 years post-op using the point cluster technique for tibiofemoral kinematics, and a mathematical model for internal force calculations. Using the identical type of prosthesis on the simulator, wear tests were conducted in displacement mode on a closed-loop controlled station. Because x-rays of the patient suggested an internal rotation of the tibial tray, it was varied form 0–10° and the effect on location and wear scar dimension was assessed. Results were compared with the retrieved liner (obtained after 13 years in vivo).Introduction
Methods