Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 77 - 77
1 Feb 2020
Ramirez-Martinez I Smith S Trail I Joyce T
Full Access

Introduction

Despite the positive outcomes in shoulder joint replacements in the last two decades, polyethylene wear debris in metal-on-polyethylene artificial shoulder joints is well-known as a limitation in the long-term survival of shoulder arthroplasties systems. Consequently, there is an interest in the use of novel materials as an alternative to hard bearing surfaces such as pyrolytic carbon layer (PyroCarbon).

Materials and Methods

In the present study, the unique Newcastle Shoulder Wear Simulator was used (Smith et al., 2015; Smith et al., 2016) to evaluate the wear behavior of four commercially available PyroCarbon humeral heads 43 mm diameter, articulating against conventional ultra-high molecular weight polyethylene (UHMWPE) glenoid inserts with a radius of curvature of 17.5 mm to form an anatomic total shoulder arthroplasty.

A physiological combined cycled “Repeat-motion-load” (RML) (Ramirez-Martinez et al., 2019) obtained from the typical activities of daily life of patients with shoulder implants was applied as a simulator input. A fifth sample of the same size and design was used as a soak control and subjected to dynamic loading without motion during the wear test. The mean volumetric wear rate of PyroCarbon-on-polyethylene was evaluated over 5 million cycles gravimetrically and calculated on the basis of linear regression, as well as the change in surface roughness (Sa) of the components using a non-contacting white light profilometer throughout the test.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 149 - 149
1 Feb 2020
Kandemir G Smith S Joyce T
Full Access

Introduction

Total hip replacement with metal-on-polymer (MoP) hip prostheses is a successful treatment for late-stage osteoarthritis. However, the wear debris generated from the polymer acetabular liners remains a problem as it can be associated with osteolysis and aseptic loosening of the implant. This has led to the investigation of more wear resistant polymers in orthopaedics. Cross-linked polyethylene (XLPE) is now the gold-standard acetabular liner material. However, we asked if carbon fibre reinforced polyether ether ketone (CFR-PEEK) might be a lower wear material. In addition, we sought to understand the influence of contact stress on the wear of both XLPE and CFR-PEEK as this has not previously been reported.

Materials and Methods

A 50-station circularly translating pin-on-disc (SuperCTPOD) machine was used to wear test both XLPE and CFR-PEEK pins against cobalt chromium (CoCr) discs to investigate the influence of contact stress on their wear rates. Fifty XLPE and 50 CFR-PEEK pins were articulated against CoCr discs. The pins, 9 mm in outer diameter and 12 mm in height, were drilled with different diameter holes to generate different sized annuli and thus, different contact areas. The pins were tested at 1.10, 1.38, 1.61, 2.00 and 5.30 MPa, which are typical contact stresses observed in the natural hip joint. An additional pin for every test group was used as a control to track the lubricant uptake. The discs were polished to 0.015 μm Sa prior to testing. The test stations contained 16 ml of diluted newborn calf serum (protein concentration: 22 g/L). Wear was measured gravimetrically with a balance (resolution: 10 μm) every 500,000 cycles. A standardised cleaning and weighing protocol was followed.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 15 - 15
1 Apr 2019
Ramirez-Martinez I Smith S Joyce T
Full Access

INTRODUCTION

Shoulder joint prostheses have become the most commonly replaced after knee and hip artificial implants. Reverse shoulder arthroplasty (RSA) is the treatment option for patients with severe osteoarthritis, rotator cuff arthropathy or a massive rotator cuff tear with pseudoparalysis. Though successful, the long-term survival of such implants are limited by wear of the materials in contact [1, 2]. The aim of this study was to investigate RSA wear in vitro using a clinically relevant activities of daily living (ADLs).

MATERIALS AND METHODS

Four new JRI Orthopaedics Reverse Shoulder 42 mm diameter VAIOS with cobalt-chromium (CoCr) glenospheres and ultra-high molecular weight polyethylene (UHMWPE) humeral components were tested. A five million cycles wear test was undertaken using the unique Newcastle Shoulder Wear Simulator with dilute bovine serum as a lubricant. “Mug to mouth” was performed as the ADL to the test prostheses in intervals of 100 cycles, following by 5 seconds of high load (450N) with no motion simulating an ADL such as “lifting an object”. This combined load cycle was then repeated. A fifth reverse shoulder prosthesis was subject to dynamic loading only in a soak control station. Wear was assessed gravimetrically and roughness (Sa) of the articulating surfaces was measured with a non-contacting profilometer.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 17 - 17
1 Apr 2019
Bhalekar R Smith S Joyce T
Full Access

Introduction

Metal-on-polyethylene (MoP) is the most commonly used bearing couple in total hip replacements (THRs). Retrieval studies (Cooper et al, 2012, JBJS, Lindgren et al, 2011, JBJS) report adverse reactions to metal debris (ARMD) due to debris produced from the taper-trunnion junction of the modular MoP THRs. A recent retrospective observational study (Matharu et al, 2016, BMC Musc Dis) showed that the risk of ARMD revision surgery is increasing in MoP THRs. To the authors' best knowledge, no hip simulator tests have investigated material loss from the taper-trunnion junction of contemporary MoP THRs.

Methods

A 6-station anatomical hip joint simulator was used to investigate material loss at the articulating and taper-trunnion surfaces of 32mm diameter metal-on-cross-linked polyethylene (MoXLPE) joints for 5 million cycles (Mc) with a sixth joint serving as a dynamically loaded soak control. Commercially available cobalt-chromium-molybdenum (CoCrMo) femoral heads articulating against XLPE acetabular liners (7.5Mrad) were used with a diluted new-born-calf-serum lubricant. Each CoCrMo femoral head was mounted on a 12/14 titanium alloy trunnion. The test was stopped every 0.5Mc, components were cleaned and gravimetric measurements performed following ISO 14242-2 and the lubricant was changed. Weight loss (mg) obtained from gravimetric measurements was converted into volume loss (mm3) and wear rates were calculated from the slopes of the linear regression lines in the volumetric loss versus number of cycles plot for heads, liners and trunnions. Additionally, volumetric measurements of the head tapers were obtained using a coordinate measuring machine (CMM) post-test. The surface roughness (Sa) of all heads and liners was measured pre and post-test. At the end of the test, the femoral heads were cut and the roughness of the worn and unworn area was measured. Statistical analysis was performed using a paired-t-test (for roughness measurements) and an independent sample t-test (for wear rates).


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 16 - 16
1 Apr 2019
Bhalekar R Smith S Joyce T
Full Access

Introduction

The bearing surfaces of ceramic-on-ceramic (CoC) total hip replacements (THR) show a substantially lower wear rate than metal-on-polyethylene (MoP) THR in-vitro. However, revision rates for CoC THR are comparable with MoP. Our hypothesis that an explanation could be adverse reaction to metal debris (ARMD) from the trunnion led us to investigate the wear at both the bearing surfaces and the taper-trunnion interface of a contemporary CoC THR in an in-vitro study.

Methods

Three 36mm CoC hips were tested in a hip simulator for 5 million cycles (Mc). BIOLOX®delta ceramic femoral heads were mounted on 12/14 titanium (Ti6Al4V) trunnions. Wear of femoral heads, acetabular liners and trunnions was determined gravimetrically using the analytical balance. Roughness measurements (Sa) were taken on the articulating surfaces (pre and post-test) and on the trunnion surfaces (worn and unworn). Furthermore, Energy Dispersive X-ray Spectroscopy (EDX) was used to identify and quantify the wear debris present in the lubricant using scanning electron microscope (SEM).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 129 - 129
1 Sep 2012
Scharfenberger A Verma S Beaupre L Kemp KA Smith S
Full Access

Purpose

Management of compound fractures, which have a higher infection risk than closed fractures, currently depends on surgeon training and past practice rather than evidence based practice. Some centres use delayed closure involving a second surgery with repeat debridement and wound closure 48 hours after initial debridement and fixation. Other centres use primary closure in the absence of gross contamination or major soft tissue deficits, where debridement, fixation and wound closure occur during the initial surgery. Delayed closure was used at our centre until January 2009 when the standard of care evolved to primary closure where appropriate. Primary closure allows more efficient OR utilization due to fewer OR visits, but it is unknown if primary closure increases the risk of infection, which can, in turn, lead to fracture non-union. The purpose of this pilot study was to complete a safety analysis of infection rates in the first 40 patients undergoing primary closure of a compound fracture; enrolment is ongoing and updated results will be presented.

Method

Patients admitted in 2010 with a long bone(femur, tibia/fibula, humerus, radius/ulna) Gustilo grade I-IIIA compound fracture, without the following: gross organic contamination, compartment or crush syndrome, amputation, or gunshot wound, were eligible for primary closure at fracture fixation, and thus for study inclusion.

The analysis compared primary closure subjects with matched delayed closure subjects taken from a previous prospective cohort study of >700 subjects. Subjects were matched at a one:two ratio(i.e. one primary closure:two delayed closure patients) on fracture location, Gustilo grade of fracture, age(within five years), significant comorbidities(diabetes, kidney disease and osteoporosis) and social factors(smoking and alcohol abuse). The outcomes were 1) any infection and 2) deep infection within six weeks of surgery. Time on antibiotics and length of hospital stay(LOS) was also recorded.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_II | Pages 13 - 13
1 Feb 2012
Steffen R Smith S Gill H Beard D McLardy-Smith P Urban J Murray D
Full Access

This study aims to investigate femoral blood flow during Metal-on-Metal Hip Resurfacing (MMHR) by monitoring oxygen concentration during the operative procedure.

Patients undergoing MMHR using the posterior approach were evaluated. Following division of fascia lata, a calibrated gas-measuring electrode was inserted into the femoral neck, aiming for the supero-lateral quadrant of the head. Baseline oxygen concentration levels were detected after electrode insertion 2-3cm below the femoral head surface and all intra-operative measures were referenced against these. Oxygen levels were continuously monitored throughout the operation. Data from ten patients are presented.

Oxygen concentration dropped most noticeably during the surgical approach and was reduced by 62% (Std.dev +/-26%) following dislocation and capsulectomy. Insertion of implants resulted in a further oxygenation decrease by 18% (Std.dev +/-28%). The last obtained measure before wound closure detected 22% (Std.dev +/-31%) of initial baseline oxygen levels. Variation between subjects was observed and three patients demonstrated a limited recovery of oxygen levels during implant insertion and hip relocation.

Intra-operative measurement of oxygen concentration in blood perfusing the femoral head is feasible. Results in ten patients undergoing MMHR showed a dramatic effect on the oxygenation in the femoral head during surgical approach and implant fixation. This may increase the risk of avascular necrosis and subsequent femoral neck fracture. Future experiments will determine if less invasive procedures or specific positioning of the limb can protect the blood supply to femoral neck and head.