Long-term clinical outcomes for ceramic-on-ceramic (CoC) bearings
are encouraging. However, there is a risk of squeaking. Guidelines
for the orientation of the acetabular component are defined from
static imaging, but the position of the pelvis and thus the acetabular
component during activities associated with edge-loading are likely
to be very different from those measured when the patient is supine.
We assessed the functional orientation of the acetabular component. A total of 18 patients with reproducible squeaking in their CoC
hips during deep flexion were investigated with a control group
of 36 non-squeaking CoC hips. The two groups were matched for the
type of implant, the orientation of the acetabular component when
supine, the size of the femoral head, ligament laxity, maximum hip
flexion and body mass index. Aims
Patients and Methods
We analysed 54 alumina ceramic-on-ceramic bearings
from total hip replacements retrieved at one centre after a mean
duration of 3.5 years (0.2 to 10.6)
We evaluated the short-term functional outcome and prevalence
of bearing-specific generation of audible noise in 301 patients
(336 hips) operated on with fourth generation (Delta) medium diameter
head, ceramic-on-ceramic (CoC) total hip arthroplasties (THAs). There were 191 female (63%) and 110 male patients (37%) with
a mean age of 61 years (29 to 78) and mean follow-up of 2.1 years
(1.3 to 3.4). Patients completed three questionnaires: Oxford Hip
Score (OHS), Research and Development 36-item health survey (RAND-36)
and a noise-specific symptom questionnaire. Plain radiographs were also
analysed. A total of three hips (0.9%) were revised. Aims
Patients and Methods
Aims. This study reports the mid-term results of total hip arthroplasty (THA) performed using a monoblock acetabular component with a large-diameter head (LDH) ceramic-on-ceramic (CoC) bearing. Patients and Methods. Of the 276 hips (246 patients) included in this study, 264 (96%) were reviewed at a mean of 67 months (48 to 79) postoperatively. Procedures were performed with a mini posterior approach. Clinical and radiological outcomes were recorded at regular intervals. A noise assessment questionnaire was completed at last follow-up. Results. There were four re-operations (1%) including one early revision for insufficient primary fixation (0.4%). No hip dislocation was reported. The mean University of California, Los Angeles (UCLA) activity score, 12-Item Short-Form Health Survey (SF-12) Mental Component Summary (MCS) score, SF-12 Physical Component Summary (PCS) score, Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) score, and Forgotten Joint Score (FJS) were 6.6 (2 to 10), 52.8 (25.5 to 65.7), 53.0 (27.2 to 66.5), 7.7 (0 to 63), and 88.5 (23 to 100), respectively. No signs of loosening or osteolysis were observed on radiological review. The incidence of squeaking was 23% (n = 51/225).
We investigated factors that were thought to be associated with an increased incidence of squeaking of ceramic-on-ceramic total hip replacements. Between June 1997 and December 2008 the three senior authors implanted 2406 primary total hip replacements with a ceramic-on-ceramic bearing surface. The mean follow-up was 10.6 years. The diagnosis was primary osteoarthritis in each case, and no patient had undergone previous surgery to the hip. We identified 74 squeaking hips (73 patients) giving an incidence of 3.1% at a mean follow-up of 9.5 years (4.1 to 13.3). Taller, heavier and younger patients were significantly more likely to have hips that squeaked.
The benefit of a dual-mobility acetabular component (DMC) for primary total hip arthroplasties (THAs) is controversial. This study aimed to compare the dislocation and complication rates when using a DMC compared to single-mobility (SM) acetabular component in primary elective THA using data collected at a single centre, and compare the revision rates and survival outcomes in these two groups. Between 2010 and 2019, 2,075 primary THAs using either a cementless DM or SM acetabular component were included. Indications for DMC were patients aged older than 70 years or with high risk of dislocation. All other patients received a SM acetabular component. Exclusion criteria were cemented implants, patients treated for femoral neck fracture, and follow-up of less than one year. In total, 1,940 THAs were analyzed: 1,149 DMC (59.2%) and 791 SM (40.8%). The mean age was 73 years (SD 9.2) in the DMC group and 57 years (SD 12) in the SM group. Complications and revisions have been analyzed retrospectively.Aims
Methods
This systematic review of the literature summarises
the clinical experience with ceramic-on-ceramic hip bearings over
the past 40 years and discusses the concerns that exist in relation
to the bearing combination. Loosening, fracture, liner chipping
on insertion, liner canting and dissociation, edge-loading and squeaking
have all been reported, and the relationship between these issues
and implant design and surgical technique is investigated. New design
concepts are introduced and analysed with respect to previous clinical
experience.
Noise generation has been reported with ceramic-on-ceramic
articulations in total hip replacement (THR). This study evaluated
208 consecutive Delta Motion THRs at a mean follow-up of 21 months
(12 to 35). There were 141 women and 67 men with a mean age of 59
years (22 to 84). Patients were reviewed clinically and radiologically,
and the incidence of noise was determined using a newly described
assessment method. Noise production was examined against range of
movement, ligamentous laxity, patient-reported outcome scores, activity
level and orientation of the acetabular component. There were 143
silent hips (69%), 22 (11%) with noises other than squeaking, 17
(8%) with unreproducible squeaking and 26 (13%) with reproducible
squeaking. Hips with reproducible squeaking had a greater mean range
of movement (p <
0.001) and mean ligament laxity (p = 0.004), smaller
median head size (p = 0.01) and decreased mean acetabular component
inclination (p = 0.02) and anteversion angle (p = 0.02) compared
with the other groups. There was no relationship between squeaking
and age (p = 0.13), height (p = 0.263), weight (p = 0.333), body
mass index (p = 0.643), gender (p = 0.07) or patient outcome score
(p = 0.422). There were no revisions during follow-up. Despite the
surprisingly high incidence of squeaking, all patients remain satisfied
with their hip replacement. Cite this article:
We report the clinical and radiological outcomes
of a series of contemporary cementless ceramic-on-ceramic total hip
replacements (THRs) at ten years in patients aged ≤ 55 years of
age. Pre- and post-operative activity levels are described. A total
of 120 consecutive ceramic cementless THRs were performed at a single
centre in 110 patients from 1997 to 1999. The mean age of the patients
at operation was 45 years (20 to 55). At ten years, four patients
had died and six were lost to follow-up, comprising ten hips. The
mean post-operative Harris hip score was 94.7 (55 to 100). Radiological
analysis was undertaken in 90 available THRs of the surviving 106
hips at final review: all had evidence of stable bony ingrowth,
with no cases of osteolysis. Wear was undetectable. There were four
revisions. The survival for both components with revision for any
cause as an endpoint was 96.5% (95% confidence interval 94.5 to
98.7). The mean modified University of California, Los Angeles activity
level rose from a mean of 6.4 (4 to 10) pre-operatively to 9.0 (6
to 10) at the ten-year post-operative period. Alumina ceramic-on-ceramic bearings in cementless primary THR
in this series have resulted in good clinical and radiological outcomes
with undetectable rates of wear and excellent function in the demanding
younger patient group at ten years. Cite this article:
We reviewed the literature on the currently available
choices of bearing surface in total hip replacement (THR). We present
a detailed description of the properties of articulating surfaces
review the understanding of the advantages and disadvantages of
existing bearing couples. Recent technological developments in the
field of polyethylene and ceramics have altered the risk of fracture
and the rate of wear, although the use of metal-on-metal bearings has
largely fallen out of favour, owing to concerns about reactions
to metal debris. As expected, all bearing surface combinations have
advantages and disadvantages. A patient-based approach is recommended,
balancing the risks of different options against an individual’s
functional demands. Cite this article:
This multicentre study analysed 12 alumina ceramic-on-ceramic
components retrieved from squeaking total hip replacements after
a mean of 23 months
Large femoral heads have become popular in total
hip replacement (THR) as a method of reducing the risk of dislocation.
However, if large heads are used in ceramic-on-ceramic THR, the
liner must be thinner, which may increase the risk of fracture.
To compare the rates of ceramic fracture and dislocation between
28 mm and 32 mm ceramic heads, 120 hips in 109 patients (51 men
and 58 women, mean age 49.2 years) were randomised to THR with either
a 28 mm or a 32 mm ceramic articulation. A total of 57/60 hips assigned
to the 28 mm group and 55/60 hips assigned to the 32 mm group were
followed for at least five years. No ceramic component fractures
occured in any patient in either group. There was one dislocation
in the 32 mm group and none in the 28 mm group (p = 0.464). No hip
had detectable wear, focal osteolysis or prosthetic loosening. In
our small study the 32 mm ceramic articulation appeared to be safe
in terms of ceramic liner fracture. Cite this article:
A total of 219 hips in 192 patients aged between
18 and 65 years were randomised to 28-mm metal-on-metal uncemented
total hip replacements (THRs, 107 hips) or hybrid hip resurfacing
(HR, 112 hips). At a mean follow-up of eight years (6.6 to 9.3)
there was no significant difference between the THR and HR groups
regarding rate of revision (4.0% (4 of 99) Cite this article:
Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce cementless fixation, which had been used successfully in the fixation of acetabular and tibial components. This modification appeared to have resulted in metallosis due to abrasive two-body wear. The component has subsequently been recalled and is no longer in use. The presentation, investigation, and findings at revision are described and a possible aetiology and its implications are discussed.