Advertisement for orthosearch.org.uk
Results 1 - 50 of 50
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1356 - 1362
1 Oct 2010
Simpson DJ Kendrick BJL Hughes M Glyn-Jones S Gill HS Rushforth GF Murray DW

We have evaluated the difference in the migration patterns over two years of two cementless stems in a randomised, controlled trial using radiostereophotogrammetric analysis (RSA). The implants studied were the Furlong HAC stem, which has good long-term results and the Furlong Active stem, which is a modified version of the former designed to minimise stress concentrations between the implant and bone, and thus to improve fixation. A total of 23 Furlong HAC and 20 Furlong Active stems were implanted in 43 patients. RSA examinations were carried out immediately post-operatively and at six, 12 and 24 months post-operatively. The subsidence during the first year in the Furlong HAC stem, was approximately one-third that of the Furlong Active stem, the measured mean subsidence of the femoral head at six months being 0.27 mm (95% confidence interval (CI) 0.03 to 0.51) and 0.99 mm (95% CI 0.38 to 1.60), respectively (p = 0.03). One Active stem continued to subside during the second year. All hips, regardless of the type of stem were clinically successful as judged by the Oxford hip score and a derived pain score without any distinction between the two types of stem. The initial stability of the Furlong Active stem was not as good as the established stem which might compromise osseo-integration to the detriment of long-term success. The changes in the geometry of the stem, to minimise stress have affected the attainment of initial stability


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 749 - 754
1 Jun 2015
Syed MA Hutt NJ Shah N Edge AJ

This study reports the results of 38 total hip arthroplasties (THAs) in 33 patients aged <  50 years, using the JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component. This represents an update of previous reports of the same cohort at ten and 16 years, which were reported in 2004 and 2009, respectively. We describe the survival, radiological and functional outcomes at a mean follow-up of 21 years (17 to 25). Of the surviving 34 THAs, one underwent femoral revision for peri-prosthetic fracture after 21 years, and one patient (one hip) was lost to follow-up. Using aseptic loosening as the end-point, 12 hips (31.5%) needed acetabular revision but none needed femoral revision, demonstrating 100% survival (95% confidence interval 89 to 100). In young patients with high demands, the Furlong HAC–coated femoral component gives excellent long-term results. Cite this article: Bone Joint J 2015;97-B:749–54


Bone & Joint Open
Vol. 5, Issue 4 | Pages 286 - 293
9 Apr 2024
Upadhyay PK Kumar V Mirza SB Shah N

Aims. This study reports the results of 38 total hip arthroplasties (THAs) in 33 patients aged less than 50 years, using the JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component. Methods. We describe the survival, radiological, and functional outcomes of 33 patients (38 THAs) at a mean follow-up of 27 years (25 to 32) between 1988 and 2018. Results. Of the surviving 30 patients (34 THAs), there were four periprosthetic fractures: one underwent femoral revision after 21 years, two had surgical fixation as the stem was deemed stable, and one was treated nonoperatively due to the patient’s comorbidities. The periprosthetic fracture patients showed radiological evidence of change in bone stock around the femoral stem, which may have contributed to the fractures; this was reflected in change of the canal flare index at the proximal femur. Two patients (two hips) were lost to follow-up. Using aseptic loosening as the endpoint, 16 patients (18 hips; 48%) needed acetabular revision. None of the femoral components were revised for aseptic loosening, demonstrating 100% survival. The estimate of the cumulative proportion surviving for revisions due to any cause was 0.97 (standard error 0.03). Conclusion. In young patients with high demands, the Furlong HAC-coated femoral component gives excellent long-term results. Cite this article: Bone Jt Open 2024;5(4):286–293



The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 865 - 869
1 Jul 2009
Shah NN Edge AJ Clark DW

In 2004 we described the ten-year prospective results of 38 total hip replacements using the Furlong hydroxyapatite-ceramic-coated femoral component in 35 patients < 50 years old. We have now reviewed the surviving 35 arthroplasties in 33 patients at a mean of 16 years (10.3 to 19.9). The mean age of the surviving patients at the time of operation was 41.3 years (26.0 to 49.0). Of these, eight have undergone revision of their acetabular component for aseptic loosening. None of the femoral components has had revision for aseptic loosening giving a survival rate of 100% at 16 years (95% confidence interval 89% to 100%). The Furlong hydroxyapatite-ceramic-coated femoral component gives excellent long-term survival in young and active patients



The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 3 | Pages 461 - 461
1 Apr 2001
FURLONG R


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 4 | Pages 745 - 746
1 Jul 1998
FURLONG R






The Bone & Joint Journal
Vol. 102-B, Issue 1 | Pages 11 - 16
1 Jan 2020
Parker MJ Cawley S

Aims. Debate continues about whether it is better to use a cemented or uncemented hemiarthroplasty to treat a displaced intracapsular fracture of the hip. The aim of this study was to attempt to resolve this issue for contemporary prostheses. Methods. A total of 400 patients with a displaced intracapsular fracture of the hip were randomized to receive either a cemented polished tapered stem hemiarthroplasty or an uncemented Furlong hydroxyapatite-coated hemiarthroplasty. Follow-up was conducted by a nurse blinded to the implant at set intervals for up to one year from surgery. Results. A total of 115 patients died in the year after surgery. There was a tendency towards a slightly higher mortality in those treated with the uncemented prosthesis after one year (64 vs 51; p = 0.18). For the survivors, there was no significant difference in pain score at any of the time intervals. Patients treated using the cemented hemiarthroplasty recovered mobility better than those treated with the uncemented hemiarthroplasty (mean decrease in mobility score at one year: 1.7 vs 1.1, SD 1.9; p = 0.008). There was a tendency to more periprosthetic fractures in the uncemented group (five vs two cases; p = 0.45), but overall the need for further surgery was similar in both groups (nine vs seven cases). There were four perioperative deaths in the cemented group. Conclusion. These results indicate that a contemporary cemented hemiarthroplasty gives better results than an uncemented hemiarthroplasty for patients with a displaced intracapsular fracture of the hip. When the condition of the patient permits, a cemented hemiarthroplasty should be used. Cite this article: Bone Joint J. 2020;102-B(1):11–16


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 27 - 30
1 Jan 2008
Rajaratnam SS Jack C Tavakkolizadeh A George MD Fletcher RJ Hankins M Shepperd JAN

Between 1986 and 1991 we implanted 331 consecutive Furlong hydroxyapatite-coated femoral components of a total hip replacement in 291 patients. A cemented acetabular prosthesis was used in 217 hips and a hydroxyapatite-coated component in 114. We describe the long-term clinical and radiological survival of the femoral component at a mean follow-up of 17.5 years (15 to 21). Only two patients (0.68%) were lost to follow-up. With revision of the femoral component for any reason as the endpoint, the survival at a mean of 17 years was 97.4% (95% confidence interval 94.1 to 99.5), and with revision for aseptic loosening as the endpoint it was 100%. The survival at a maximum of 21 years with revision of the femoral component for any reason as the endpoint was 97.4% (95% confidence interval 81.0 or 99.5). These results compare favourably with the best long-term results of cemented or uncemented femoral components used in total hip replacement


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 8 | Pages 1118 - 1123
1 Nov 2004
Singh S Trikha SP Edge AJ

We describe the clinical and radiological results of 38 total hip replacements (THR) using the JRI Furlong hydroxyapatite-ceramic (HAC)-coated femoral component in patients younger than 50 years. The mean age at the time of operation was 42 years (22 to 49) and the mean length of follow-up was ten years (63 to 170 months). All patients receiving a Furlong HAC THR were entered into the study regardless of the primary pathology including patients who had undergone previous hip surgery. The mean Harris hip score improved from 44 before operation to 92 at the last postoperative review. After 12 years the cumulative surivival for the stem was 100% (95% confidence interval 89 to 100). No femoral component was revised. Our results show that the Furlong HAC implant gives excellent long-term results in young patients with high demands


The Journal of Bone & Joint Surgery British Volume
Vol. 73-B, Issue 5 | Pages 741 - 745
1 Sep 1991
Furlong R Osborn J

We report the histological findings in post-mortem specimens obtained ten days, 17 days and seven weeks after implantation of hydroxyapatite-coated femoral components of hip arthroplasties. There was early deposition of woven bone on the hydroxyapatite ceramic, identical to that deposited on surviving cancellous trabeculae. The space between these deposits became bridged from both sides by new trabeculae, and there was no evidence of an inflammatory reaction or of fibrous tissue formation. The use of an hydroxyapatite coating seems to allow early, sound, secondary fixation of implants.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1055 - 1060
1 Aug 2005
Trikha SP Singh S Raynham OW Lewis JC Mitchell PA Edge AJ

We describe the clinical and radiological results of 120 consecutive revision hip replacements in 107 patients, using the JRI Furlong hydroxyapatite-ceramic-coated femoral component. The mean age of the patients at operation was 71 years (36 to 92) and the mean length of follow-up 8.0 years (5.0 to 12.4). We included patients on whom previous revision hip surgery had taken place. The patients were independently reviewed and scored using the Harris hip score, the Western Ontario and McMaster Universities osteoarthritis index (WOMAC) and the Charnley modification of the Merle d’Aubigné and Postel score. Radiographs were assessed by three reviewers for the formation of new bone, osteolysis, osseointegration and radiolucent lines in each Gruen zone. The mean Harris hip score was 85.8 (42 to 100) at the latest post-operative review. The mean WOMAC and Merle d’Aubigné and Postel scores were 34.5 and 14.8, respectively. The mean visual analogue score for pain (possible range 0 to 10) was 1.2 overall, but 0.5 specifically for mid-thigh pain. There were no revisions of the femoral component for aseptic loosening. There were four re-revisions, three for infection and one for recurrent dislocation. Radiological review of all the femoral components, including the four re-revisions showed stable bony ingrowth and no new radiolucent lines in any zone. Using revision or impending revision for aseptic loosening as an end-point, the cumulative survival of the femoral component at ten years was 100% (95% confidence interval 94 to 100). We present excellent medium- to long-term clinical, radiological and survivorship results with the fully hydroxyapatite-ceramic-coated femoral component in revision hip surgery


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1050 - 1054
1 Aug 2005
Shetty AA Slack R Tindall A James KD Rand C

We describe the survival of 134 consecutive JRI Furlong hydroxyapatite-coated uncemented total hip replacements. The mean follow-up was for 14.2 years (13 to 15). Patients were assessed clinically, using the Merle d’Aubigné and Postel score. Radiographs were evaluated using Gruen zones for the stem and DeLee and Charnley zones for the cup. Signs of subsidence, radiolucent lines, endosteal bone formation (spot welds) and pedestal formation were used to assess fixation and stability of the stem according to Engh’s criteria. Cup angle, migration and radiolucency were used to assess loosening of the cup. The criteria for failure were revision, or impending revision because of pain or loosening. Survival analysis was performed using a life table and the Kaplan-Meier curve. The mean total Merle d’Aubigné and Postel score was 7.4 pre-operatively and 15.9 at follow-up. During the study period 22 patients died and six were lost to follow-up. None of the cups was revised. One stem was revised for a periprosthetic fracture following a fall but none was revised for loosening, giving a 99% survival at 13 years. Our findings suggest that the long-term results of these hydroxyapatite-coated prostheses are more than satisfactory


The Bone & Joint Journal
Vol. 105-B, Issue 11 | Pages 1196 - 1200
1 Nov 2023
Parker MJ Chatterjee R Onsa M Cawley S Gurusamy K

Aims

The aim of this study was to report the three-year follow-up for a series of 400 patients with a displaced intracapsular fracture of the hip, who were randomized to be treated with either a cemented polished tapered hemiarthroplasty or an uncemented hydroxyapatite-coated hemiarthroplasty.

Methods

The mean age of the patients was 85 years (58 to 102) and 273 (68%) were female. Follow-up was undertaken by a nurse who was blinded to the hemiarthroplasty that was used, at intervals for up to three years from surgery. The short-term follow-up of these patients at a mean of one year has previously been reported.


Bone & Joint Open
Vol. 3, Issue 5 | Pages 423 - 431
1 May 2022
Leong JWY Singhal R Whitehouse MR Howell JR Hamer A Khanduja V Board TN

Aims

The aim of this modified Delphi process was to create a structured Revision Hip Complexity Classification (RHCC) which can be used as a tool to help direct multidisciplinary team (MDT) discussions of complex cases in local or regional revision networks.

Methods

The RHCC was developed with the help of a steering group and an invitation through the British Hip Society (BHS) to members to apply, forming an expert panel of 35. We ran a mixed-method modified Delphi process (three rounds of questionnaires and one virtual meeting). Round 1 consisted of identifying the factors that govern the decision-making and complexities, with weighting given to factors considered most important by experts. Participants were asked to identify classification systems where relevant. Rounds 2 and 3 focused on grouping each factor into H1, H2, or H3, creating a hierarchy of complexity. This was followed by a virtual meeting in an attempt to achieve consensus on the factors which had not achieved consensus in preceding rounds.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 7 | Pages 974 - 978
1 Sep 2004
Thomas SR Shukla D Latham PD

Cemented titanium stems in hip arthroplasty are associated with proximal cement-stem debonding and early failure. This was well publicised with the 3M Capital hip. However, corrosion in this setting has been reported with only one stem design and is less widely accepted. We present a series of 12 cemented titanium Furlong Straight Stems which required revision at a mean of 78 months for thigh pain. At revision the stems were severely corroded in a pattern which was typical of crevice corrosion. Symptoms were eliminated after revision to an all-stainless steel femoral prosthesis of the same design. We discuss the likely causes for the corrosion. The combination of a titanium stem and cement appears to facilitate crevice corrosion


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 1 | Pages 27 - 30
1 Jan 2004
Hallam P Haddad F Cobb J

We have investigated nine patients with cemented Furlong (JRI, London, UK) titanium hip replacements who presented with early pain despite a well-fixed, aseptic prosthesis. All were followed up clinically and radiologically at regular intervals. Pain was located in the thigh and was worse at night. Radiographs showed cortical hypertrophy of the femur around the tip of the stem. Eight of the nine patients subsequently required single-stage revision using an uncemented prosthesis, which relieved the pain. At revision, the pH of the tip of the stem was found to be highly acidic with macroscopic evidence of corrosion consisting of multiple layers of titanium oxides when studied by X-ray dispersive analysis. Cemented titanium implants have a potential for crevice corrosion leading to cortical hypertrophy and intractable pain


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 3 | Pages 378 - 382
1 Apr 2000
McNally SA Shepperd JAN Mann CV Walczak JP

We describe the clinical and radiological outcome of 100 consecutive total hip replacements in 86 patients using the JRI Furlong hydroxyapatite-coated femoral component. The follow-up was 100% at a mean of ten years (9.0 to 12.0). All 54 living patients (62 hips) were examined annually. Their mean Merle d’Aubigné score was 5.8/5.6/4.6. A total of 32 patients had died. Their notes at the last annual assessment recorded a similar outcome to that of the living patients. Radiographs showed no radiolucent lines around any femoral component and there were no signs of impending failure. One patient required excision arthroplasty, but the femoral component had been bonded satisfactorily. There have been no other revisions of femoral components. Our findings suggest that the hydroxyapatite coating gives a satisfactory prosthesis-bone interface which is preferable to any other system


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 5 | Pages 655 - 658
1 Jul 2000
Giannoudis PV MacDonald DA Matthews SJ Smith RM Furlong AJ De Boer P

We assessed factors which may affect union in 32 patients with nonunion of a fracture of the diaphysis of the femur and 67 comparable patients whose fracture had united. These included gender, age, smoking habit, the use of non-steroidal anti-inflammatory drugs (NSAIDs) the type of fracture (AO classification), soft-tissue injury (open or closed), the type of nail, the mode of locking, reaming v non-reaming, infection, failure of the implant, distraction at the fracture site, and the time to full weight-bearing. Patients with severe head injuries were excluded. Both groups were comparable with regard to gender, Injury Severity Score and soft-tissue injury.

There was no relationship between the rate of union and the type of implant, mode of locking, reaming, distraction or smoking. There were fewer cases of nonunion in more comminuted fractures (type C) and in patients who were able to bear weight early. There was a marked association between nonunion and the use of NSAIDs after injury (p = 0.000001) and delayed healing was noted in patients who took NSAIDs and whose fractures had united.


The Bone & Joint Journal
Vol. 102-B, Issue 9 | Pages 1113 - 1121
14 Sep 2020
Nantha Kumar N Kunutsor SK Fernandez MA Dominguez E Parsons N Costa ML Whitehouse MR

Aims

We conducted a systematic review and meta-analysis to compare the mortality, morbidity, and functional outcomes of cemented versus uncemented hemiarthroplasty in the treatment of intracapsular hip fractures, analyzing contemporary and non-contemporary implants separately.

Methods

PubMed, Medline, EMBASE, CINAHL, and Cochrane Library were searched to 2 February 2020 for randomized controlled trials (RCTs) comparing the primary outcome, mortality, and secondary outcomes of function, quality of life, reoperation, postoperative complications, perioperative outcomes, pain, and length of hospital stay. Relative risks (RRs) and mean differences (with 95% confidence intervals (CIs)) were used as summary association measures.


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1618 - 1625
1 Dec 2018
Gill JR Kiliyanpilakkill B Parker MJ

Aims

This study describes and compares the operative management and outcomes in a consecutive case series of patients with dislocated hemiarthroplasties of the hip, and compares outcomes with those of patients not sustaining a dislocation.

Patients and Methods

Of 3326 consecutive patients treated with hemiarthroplasty for fractured neck of femur, 46 (1.4%) sustained dislocations. Of the 46 dislocations, there were 37 female patients (80.4%) and nine male patients (19.6%) with a mean age of 83.8 years (66 to 100). Operative intervention for each, and subsequent dislocations, were recorded. The following outcome measures were recorded: dislocation; mortality up to one-year post-injury; additional surgery; residential status; mobility; and pain score at one year.


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 50 - 59
1 Jan 2017
Carli AV Negus JJ Haddad FS

Aims

Periprosthetic femoral fractures (PFF) following total hip arthroplasty (THA) are devastating complications that are associated with functional limitations and increased overall mortality. Although cementless implants have been associated with an increased risk of PFF, the precise contribution of implant geometry and design on the risk of both intra-operative and post-operative PFF remains poorly investigated. A systematic review was performed to aggregate all of the PFF literature with specific attention to the femoral implant used.

Patients and Methods

A systematic search strategy of several journal databases and recent proceedings from the American Academy of Orthopaedic Surgeons was performed. Clinical articles were included for analysis if sufficient implant description was provided. All articles were reviewed by two reviewers. A review of fundamental investigations of implant load-to-failure was performed, with the intent of identifying similar conclusions from the clinical and fundamental literature.


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 917 - 924
1 Jul 2016
Whittaker RK Hothi HS Meswania JM Berber R Blunn GW Skinner JA Hart AJ

Aims

Surgeons have commonly used modular femoral heads and stems from different manufacturers, although this is not recommended by orthopaedic companies due to the different manufacturing processes.

We compared the rate of corrosion and rate of wear at the trunnion/head taper junction in two groups of retrieved hips; those with mixed manufacturers (MM) and those from the same manufacturer (SM).

Materials and Methods

We identified 151 retrieved hips with large-diameter cobalt-chromium heads; 51 of two designs that had been paired with stems from different manufacturers (MM) and 100 of seven designs paired with stems from the same manufacturer (SM). We determined the severity of corrosion with the Goldberg corrosion score and the volume of material loss at the head/stem junction. We used multivariable statistical analysis to determine if there was a significant difference between the two groups.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1618 - 1624
1 Dec 2012
Daurka JS Malik AK Robin DA Witt JD

The inherent challenges of total hip replacement (THR) in children include the choice of implant for the often atypical anatomical morphology, its fixation to an immature growing skeleton and the bearing surface employed to achieve a successful long-term result. We report the medium-term results of 52 consecutive uncemented THRs undertaken in 35 paediatric patients with juvenile idiopathic arthritis. The mean age at the time of surgery was 14.4 years (10 to 16). The median follow-up was 10.5 years (6 to 15). During the study period 13 THRs underwent revision surgery. With revision as an endpoint, subgroup analysis revealed 100% survival of the 23 ceramic-on-ceramic THRs and 55% (16 of 29) of the metal- or ceramic-on-polyethylene. This resulted in 94% (95% CI 77.8 to 98.4) survivorship of the femoral component and 62% (95% CI 41.0 to 78.0) of the acetabular component. Revision of the acetabular component for wear and osteolysis were the most common reasons for failure accounting for 11 of the 13 revisions.

The success seen in patients with a ceramic-on-ceramic articulation seems to indicate that this implant strategy has the potential to make a major difference to the long-term outcome in this difficult group of patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1017 - 1020
1 Aug 2011
Stafford GH Islam SU Witt JD

Ceramic-on-ceramic bearings in hip replacement have low rates of wear and are increasingly being used in young adults. Our aim was to determine the incidence of audible phenomena or other bearing-related complications.

We retrospectively analysed 250 ceramic-on-ceramic hip replacements in 224 patients which had been implanted between April 2000 and December 2007. The mean age of the patients at operation was 44 years (14 to 83) and all the operations were performed using the same surgical technique at a single centre.

At a mean follow-up of 59 months (24 to 94), the mean Oxford hip score was 40.89 (11 to 48). There were six revisions, three of which were for impingement-related complications. No patient reported squeaking, but six described grinding or clicking, which was usually associated with deep flexion. No radiological evidence of osteolysis or migration of the components was observed in any hip.

The early to mid-term results of contemporary ceramic-on-ceramic hip replacement show promising results with few concerns in terms of noise and squeaking. Positioning of the acetabular component remains critical in regard to the reduction of other impingement-related complications.


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 467 - 471
1 Apr 2013
Sandiford N Doctor C Rajaratnam SS Ahmed S East DJ Miles K Butler-Manuel A Shepperd JAN

We present the extended follow-up (≥ 20 years) of a series of fully hydroxyapatite-coated femoral components used in 72 primary total hip replacements (THRs). Earlier results of this cohort have been previously published. All procedures were performed between 1986 and 1991. The series involved 45 women and 15 men with 12 bilateral procedures. Their mean age at the time of surgery was 60 years (46 to 80) and the mean duration of follow-up was 22.5 years (20 to 25). At final follow-up, the mean Merle d’Aubigné and Postel hip scores were 5.5 (4.5 to 6), 3.8 (3.5 to 5) and 3.3 (3.0 to 5.0) for pain, mobility and function, respectively. Of the patients 92% were very satisfied at the time of final follow-up.

There were seven revisions: six of the acetabular component for aseptic loosening and one of both the stem and the acetabular component for loosening due to deep infection. The survival of this prosthesis at 22.5 years with revision for any reason as the endpoint was 91.7% (95% confidence interval (CI) 84 to 99). Survival with aseptic loosening of the stem as the endpoint was 100% (95% CI 90 to 100).

This prosthesis provides pain relief in the long term. Survival of this component is comparable to the best results for primary THR with any means of fixation.

Cite this article: Bone Joint J 2013;95-B:467–71.


The Bone & Joint Journal
Vol. 98-B, Issue 11 | Pages 1441 - 1449
1 Nov 2016
Petheram TG Whitehouse SL Kazi HA Hubble MJW Timperley AJ Wilson MJ Howell JR

Aims

We present a minimum 20-year follow-up study of 382 cemented Exeter Universal total hip arthroplasties (350 patients) operated on at a mean age of 66.3 years (17 to 94).

Patients and Methods

All patients received the same design of femoral component, regardless of the original diagnosis. Previous surgery had been undertaken for 33 hips (8.6%). During the study period 218 patients with 236 hips (62%) died, 42 hips (11%) were revised and 110 hips (29%) in 96 patients were available for review. The acetabular components were varied and some designs are now obsolete, however they were all cemented.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 619 - 623
1 May 2012
Vanhegan IS Malik AK Jayakumar P Ul Islam S Haddad FS

Revision arthroplasty of the hip is expensive owing to the increased cost of pre-operative investigations, surgical implants and instrumentation, protracted hospital stay and drugs. We compared the costs of performing this surgery for aseptic loosening, dislocation, deep infection and peri-prosthetic fracture. Clinical, demographic and economic data were obtained for 305 consecutive revision total hip replacements in 286 patients performed at a tertiary referral centre between 1999 and 2008. The mean total costs for revision surgery in aseptic cases (n = 194) were £11 897 (sd 4629), for septic revision (n = 76) £21 937 (sd 10 965), for peri-prosthetic fracture (n = 24) £18 185 (sd 9124), and for dislocation (n = 11) £10 893 (sd 5476). Surgery for deep infection and peri-prosthetic fracture was associated with longer operating times, increased blood loss and an increase in complications compared to revisions for aseptic loosening. Total inpatient stay was also significantly longer on average (p < 0.001). Financial costs vary significantly by indication, which is not reflected in current National Health Service tariffs.


Bone & Joint 360
Vol. 4, Issue 1 | Pages 2 - 5
1 Feb 2015
Wright GM Porteous MJ


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 6 | Pages 742 - 745
1 Jun 2007
Harvie P Haroon M Henderson N El-Guindi M

We describe three cases of fracture of the titanium JRI-Furlong hydroxyapatite-ceramic (HAC)-coated femoral component. We have examined previous case reports of failure of this stem and conclude that fracture may occur in two places, namely at the neck-shoulder junction and at the conical-distal cylindrical junction. These breakages are the result of fatigue in a metallurgically-proven normal femoral component. All the cases of failure of the femoral component have occurred in patients with a body mass index of more than 25 in whom a small component, either size 9 or 10, had been used.

In patients with a body mass index above normal size 9 components should be avoided and the femoral canal should be reamed sufficiently to accept a large femoral component to ensure that there is adequate metaphyseal fixation.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1046 - 1049
1 Aug 2005
Shepperd JAN Apthorp H


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 665 - 677
1 May 2011
Sköldenberg OG Salemyr MO Bodén HS Lundberg A Ahl TE Adolphson PY

Our aim in this pilot study was to evaluate the fixation of, the bone remodelling around, and the clinical outcome after surgery of a new, uncemented, fully hydroxyapatite-coated, collared and tapered femoral component, designed specifically for elderly patients with a fracture of the femoral neck.

We enrolled 50 patients, of at least 70 years of age, with an acute displaced fracture of the femoral neck in this prospective single-series study. They received a total hip replacement using the new component and were followed up regularly for two years.

Fixation was evaluated by radiostereometric analysis and bone remodelling by dual-energy x-ray absorptiometry. Hip function and the health-related quality of life were assessed using the Harris hip score and the EuroQol-5D.

Up to six weeks post-operatively there was a mean subsidence of 0.2 mm (−2.1 to +0.5) and a retroversion of a mean of 1.2° (−8.2° to +1.5°). No component migrated after three months. The patients had a continuous loss of peri-prosthetic bone which amounted to a mean of 16% (−49% to +10%) at two years. The mean Harris hip score was 82 (51 to 100) after two years.

The two-year results from this pilot study indicate that this new, uncemented femoral component can be used for elderly patients with osteoporotic fractures of the femoral neck.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1578 - 1585
1 Dec 2014
Rankin KS Sprowson AP McNamara I Akiyama T Buchbinder R Costa ML Rasmussen S Nathan SS Kumta S Rangan A

Trauma and orthopaedics is the largest of the surgical specialties and yet attracts a disproportionately small fraction of available national and international funding for health research. With the burden of musculoskeletal disease increasing, high-quality research is required to improve the evidence base for orthopaedic practice. Using the current research landscape in the United Kingdom as an example, but also addressing the international perspective, we highlight the issues surrounding poor levels of research funding in trauma and orthopaedics and indicate avenues for improving the impact and success of surgical musculoskeletal research.

Cite this article: Bone Joint J 2014; 96-B:1578–85.


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 874 - 876
1 Jul 2013
Kallala R Anderson P Morris S Haddad FS

In a time of limited resources, the debate continues over which types of hip prosthesis are clinically superior and more cost-effective. Orthopaedic surgeons increasingly need robust economic evidence to understand the full value of the operation, and to aid decision making on the ‘package’ of procedures that are available and to justify their practice beyond traditional clinical preference.

In this paper we explore the current economic debate about the merits of cemented and cementless total hip replacement, an issue that continues to divide the orthopaedic community.

Cite this article: Bone Joint J 2013;95-B:874–6.


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 299 - 305
1 Mar 2014
Bell KR Clement ND Jenkins PJ Keating JF

We performed a case–control study to compare the rates of further surgery, revision and complications, operating time and survival in patients who were treated with either an uncemented hydroxyapatite-coated Corail bipolar femoral stem or a cemented Exeter stem for a displaced intracapsular fracture of the hip. The mean age of the patients in the uncemented group was 82.5 years (53 to 97) and in the cemented group was 82.7 years (51 to 99) We used propensity score matching, adjusting for age, gender and the presence or absence of dementia and comorbidities, to produce a matched cohort receiving an Exeter stem (n = 69) with which to compare the outcome of patients receiving a Corail stem (n = 69). The Corail had a significantly lower all-cause rate of further surgery (p = 0.016; odds ratio (OR) 0.18, 95% CI 0.04 to 0.84) and number of hips undergoing major further surgery (p = 0.029; OR 0.13, 95% CI 0.01 to 1.09). The mean operating time was significantly less for the Corail group than for the cemented Exeter group (59 min [12 to 136] vs 70 min [40 to 175], p = 0.001). The Corail group also had a lower risk of a peri-prosthetic fracture (p = 0.042; OR 0.19, 95% CI 0.01 to 1.42) . There was no difference in the mortality rate between the groups. There were significantly fewer complications in the uncemented group, suggesting that the use of this stem would result in a decreased rate of morbidity in these frail patients. Whether this relates to an improved functional outcome remains unknown.

Cite this article: Bone Joint J 2014;96-B:299–305.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1561 - 1565
1 Dec 2009
Blakey CM Eswaramoorthy VK Hamilton LC Biant LC Field RE

We report the minimum five-year follow-up of 352 primary total hip replacements using the uncemented hydroxyapatite-coated ANCA-Fit femoral component with a modular neck and head. The series comprised 319 patients (212 men, 107 women) with a mean age at operation of 64.4 years (28 to 97). The principal diagnosis was osteoarthritis. A total of 18 patients (21 hips) died before their follow-up at five years, nine patients (11 hips) were lost to follow-up, and four (four hips) declined further follow-up. Patient-reported outcomes have been recorded for 288 patients (316 hips).

Their mean Oxford Hip Score improved significantly from 41 points (16 to 57) pre-operatively to 20 points (12 to 44) at five-year follow-up. Radiological assessment showed good bony stability in 98% of implants. There were two cases of aseptic loosening of the femoral component. There were no clinical or radiological complications related to modularity. In our series we did not see the high rate of intra-operative fracture previously reported for this implant.

This medium-term follow-up study demonstrates that the clinical outcome of the ANCA-Fit femoral component is, to date, comparable with that of other metaphyseal loading femoral components.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 7 | Pages 890 - 896
1 Jul 2011
Bajwa AS Villar RN

Arthroscopy of the native hip is an established diagnostic and therapeutic procedure. Its application in the symptomatic replaced hip is still being explored. We describe the use of arthroscopy of the hip in 24 symptomatic patients following total hip replacement, resurfacing arthroplasty of the hip and partial resurfacing (study group), and compared it with arthroscopy of the native hip in 24 patients (control group). A diagnosis was made or confirmed at arthroscopy in 23 of the study group and a therapeutic arthroscopic intervention resulted in relief of symptoms in ten of these. In a further seven patients it led to revision hip replacement. In contrast, arthroscopy in the control group was diagnostic in all 24 patients and the resulting arthroscopic therapeutic intervention provided symptomatic relief in 21.

The mean operative time in the study group (59.7 minutes (35 to 93)) was less than in the control group (71 minutes (40 to 100), p = 0.04) but the arthroscopic approach was more difficult in the arthroplasty group. We suggest that arthroscopy has a role in the management of patients with a symptomatic arthroplasty when other investigations have failed to provide a diagnosis.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1167 - 1167
1 Aug 2005
Horan F


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 4 | Pages 480 - 485
1 Apr 2010
Mannan K Freeman MAR Scott G

The outcome at ten years of 100 Freeman hip stems (Finsbury Orthopaedics, Leatherhead, United Kingdom) retaining the neck with a proximal hydroxyapatite coating in a series of 52 men (six bilateral) and 40 women (two bilateral), has been described previously. None required revision for aseptic loosening. We have extended the follow-up to 20 years with a minimum of 17 years. The mean age of the patients at total hip replacement was 58.9 years (19 to 84).

Six patients were lost to follow-up, but were included up to their last clinical review. A total of 22 patients (22 hips) had died, all from causes unrelated to their surgery. There have been 43 re-operations for failure of the acetabular component. However, in 38 of these the stem was not revised since it remained stable and there was no associated osteolysis. Two of the revisions were for damage to the trunnion after fracture of a modular ceramic head, and in another two, removal of the femoral component was because of the preference of the surgeon. In all cases the femoral component was well fixed, but could be extracted at the time of acetabular revision. In one case both components were revised for deep infection. There has been one case of aseptic loosening of the stem which occurred at 14 years. This stem had migrated distally by 7.6 mm in ten years and 8.4 mm at the time of revision at which stage it was found to be rotationally loose. With hindsight this component had been undersized at implantation.

The survivorship for the stem at 17 years with aseptic loosening as the endpoint was 98.6% (95% confidence interval 95.9 to 100) when 62 hips were at risk. All remaining stems had a satisfactory clinical and radiological outcome. The Freeman proximally hydroxyapatite-coated femoral component is therefore a dependable implant and its continued use can be recommended.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 236 - 239
1 Feb 2008
Gupta R Jambhekar N Sanghvi D

Giant-cell tumour of the synovium is known to affect the fingers or toes of adults. It has seldom been described in the spine and rarely in the thoracic vertebrae or in a child. The lesions of giant-cell tumour of the synovium have a classical radiological appearance, but require a high index of suspicion for correct recognition. Unlike giant-cell tumour of the synovium at other well-known sites, spinal lesions lack the characteristic papillary architecture, thereby raising other diagnostic possibilities. We describe a giant-cell tumour of the synovium of the left facet joint of a thoracic vertebra in a nine-year-old girl. The tumour was treated successfully by surgical excision.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 5 | Pages 577 - 582
1 May 2009
Duncan WW Hubble MJW Howell JR Whitehouse SL Timperley AJ Gie GA

The removal of well-fixed bone cement from the femoral canal during revision of a total hip replacement (THR) can be difficult and risks the loss of excessive bone stock and perforation or fracture of the femoral shaft. Retaining the cement mantle is attractive, yet the technique of cement-in-cement revision is not widely practised. We have used this procedure at our hospital since 1989. The stems were removed to gain a better exposure for acetabular revision, to alter version or leg length, or for component incompatibility.

We studied 136 hips in 134 patients and followed them up for a mean of eight years (5 to 15). A further revision was required in 35 hips (25.7%), for acetabular loosening in 26 (19.1%), sepsis in four, instability in three, femoral fracture in one and stem fracture in one. No femoral stem needed to be re-revised for aseptic loosening.

A cement-in-cement revision of the femoral stem is a reliable technique in the medium term. It also reduces the risk of perforation or fracture of the femoral shaft.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 116 - 122
1 Jan 2010
Parker MI Pryor G Gurusamy K

We undertook a prospective randomised controlled trial involving 400 patients with a displaced intracapsular fracture of the hip to determine whether there was any difference in outcome between treatment with a cemented Thompson hemiarthroplasty and an uncemented Austin-Moore prosthesis. The surviving patients were followed up for between two and five years by a nurse blinded to the type of prosthesis used.

The mean age of the patients was 83 years (61 to 104) and 308 (77%) were women. The degree of residual pain was less in those treated with a cemented prosthesis (p < 0.0001) three months after surgery. Regaining mobility was better in those treated with a cemented implant (p = 0.005) at six months after operation. No statistically significant difference was found between the two groups with regard to mortality, implant-related complications, re-operations or post-operative medical complications.

The use of a cemented Thompson hemiarthroplasty resulted in less pain and less deterioration in mobility than an uncemented Austin-Moore prosthesis with no increase in complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 1 | Pages 12 - 15
1 Jan 2005
Robertson A Lavalette D Morgan S Angus PD

We present the clinical and radiographic outcome of 68 consecutive primary total hip replacements performed in 54 patients under the age of 55 years using a hydroxyapatite-coated femoral component and threaded cup with a modular ceramic head (JRI-Furlong). We reviewed 62 (91%) hips at a median follow-up of 8.8 years (5 to 13.8) after implantation; six (9%) were lost to follow-up. At review there had been four (6%) revisions but only one for aseptic loosening (acetabulum). Radiographic review of the remaining hips did not identify any evidence of femoral or acetabular loosening. The median Harris and Merle d’Aubigné and Postel hip scores were 95.9 (42.7 to 100) and 17 (3 to 18) respectively. The JRI-Furlong hip gives promising functional and radiographic results in young patients in the medium term.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 879 - 886
1 Jul 2007
Howie DW Wimhurst JA McGee MA Carbone TA Badaruddin BS

We investigated 219 revisions of total hip replacement (THR) in 211 patients using a collarless double-taper cemented femoral component. The mean age of the patients was 72 years (30 to 90). The 137 long and 82 standard length stems were analysed separately. The mean follow-up was six years (2 to 18), and no patient was lost to follow-up.

Survival of the long stems to re-revision for aseptic loosening at nine years was 98% (95% confidence interval (CI) 94 to 100), and for the standard stems was 93% (95% CI 85 to 100). At five years, one long stem was definitely loose radiologically and one standard stem was probably loose. Pre-operative femoral bone deficiency did not influence the results for the long stems, and corrective femoral osteotomy was avoided, as were significant subsidence, major stress shielding and persistent thigh pain.

Because of these reliable results, cemented long collarless double-taper femoral components are recommended for routine revision THR in older patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1061 - 1067
1 Aug 2005
Raman R Kamath RP Parikh A Angus PD

We report the clinical and radiological outcome of 86 revisions of cemented hip arthroplasties using JRI-Furlong hydroxyapatite-ceramic-coated acetabular and femoral components. The acetabular component was revised in 62 hips and the femoral component in all hips. The mean follow-up was 12.6 years and no patient was lost to follow-up.

The mean age of the patients was 71.2 years. The mean Harris hip and Oxford scores were 82 (59 to 96) and 23.4 (14 to 40), respectively. The mean Charnley modification of the Merle d’Aubigné and Postel score was 5 (3 to 6) for pain, 4.9 (3 to 6) for movement and 4.4 (3 to 6) for mobility. Migration of the acetabular component was seen in two hips and the mean acetabular inclination was 42.6°. The mean linear polyethylene wear was 0.05 mm/year. The mean subsidence of the femoral component was 1.9 mm and stress shielding was seen in 23 (28%) with bony ingrowth in 76 (94%). Heterotopic ossification was seen in 12 hips (15%). There were three re-revisions, two for deep sepsis and one for recurrent dislocation and there were no re-revisions for aseptic loosening. The mean EuroQol EQ-5D description scores and health thermometer scores were 0.69 (0.51 to 0.89) and 79 (54 to 95), respectively. With an end-point of definite or probable loosening, the probability of survival at 12 years was 93.9% and 95.6% for the acetabular and femoral components, respectively. Overall survival at 12 years, with removal or further revision of either component for any reason as the end-point, was 92.3%.

Our study supports the continued use of this arthroplasty and documents the durability of hydroxyapatite-ceramic-coated components.