Fracture of a
We analysed 54 alumina ceramic-on-ceramic bearings
from total hip replacements retrieved at one centre after a mean
duration of 3.5 years (0.2 to 10.6) in situ. These
implants were obtained from 54 patients (16 men and 38 women) with
a mean age of 67 years (33 to 88) who underwent revision for a variety
of reasons. Posterior edge loading was found in the majority of
these retrievals (32 out of 54). Anterosuperior edge loading occurred
less often but produced a higher rate of wear. Stripe wear on the
femoral heads had a median volumetric wear rate of 0.2 mm. 3. /year
(0 to 7.2). The wear volume on the femoral heads corresponded to
the width of edge wear on the matching liner. Anteversion of the
acetabular component was found to be a more important determinant
than inclination for wear in
The leading indication for revision total hip
arthroplasty (THA) remains aseptic loosening owing to wear. The younger,
more active patients currently undergoing THA present unprecedented
demands on the bearings. Ceramic-on-ceramic (CoC) bearings have
consistently shown the lowest rates of wear. The recent advances,
especially involving alumina/zirconia composite
Objectives. The use of
Aims. Early evidence has emerged suggesting that ceramic-on-ceramic
articulations induce a different tissue reaction to ceramic-on-polyethylene
and metal-on-metal bearings. Therefore, the aim of this study was
to investigate the tissue reaction and cellular response to ceramic
total hip arthroplasty (THA) materials in vitro,
as well as the tissue reaction in capsular tissue after revision
surgery of ceramic-on-ceramic THAs. Patients and Methods. We investigated tissue collected at revision surgery from nine
ceramic-on-ceramic articulations. we compared our findings with
tissue obtained from five metal-on-metal THA revisions, four ceramic-on-polyethylene
THAs, and four primary osteoarthritis synovial membranes. The latter
were analyzed to assess the amount of tissue fibrosis that might
have been present at the time of implantation to enable evaluation,
in relation to implantation time, of any subsequent response in
the tissues. Results. There was a significant increase in tissue fibrosis with implantation
time for all implant types tested. Interestingly, the tissue fibrosis
in ceramic-on-ceramic THAs was significantly increased compared
with metal-on-metal and ceramic-on-polyethylene. Additionally, we
found
Large
We investigated the wear characteristics and clinical performance of four different total hip joint articulations in 114 patients. Wear and migration was measured by roentgenstereophotogrammetric analysis at five years or at the last follow-up. The mean annual wear was 0.11 mm for a stainless steel/Enduron articulation, 0.34 mm for stainless steel/Hylamer cup, 0.17 mm for zirconium oxide
Aims. Modular dual-mobility constructs reduce the risk of dislocation after revision total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chromium (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the early- to mid-term serum Co and Cr levels after modular dual-mobility components were used in revision and complex primary THAs, and specifically included patients revised for ALTR. Patients and Methods. Serum Co and Cr levels were measured prospectively in 24 patients with a modular dual-mobility construct and a
The practice of removing a well-fixed cementless
femoral component is associated with high morbidity.
Aims. Our aim in this study was to describe a continuing review of
11 total hip arthroplasties using 22.225 mm Alumina
Symptomatic cobalt toxicity from a failed total
hip replacement is a rare but devastating complication. It has been reported
following revision of fractured
It has been suggested that the wear of ultra-high molecular weight polyethylene (UHMWPE) in total hip replacement is substantially reduced when the femoral head is
At yearly intervals we compared the radiological wear characteristics of 81 alumina
We report the results of our continued review of 11 total hip arthroplasties using 22.225 mm alumina
The ideal bearing surface for young patients
undergoing total hip replacement (THR) remains controversial. We report
the five-year results of a randomised controlled trial comparing
the clinical and radiological outcomes of 102 THRs in 91 patients
who were <
65 years of age. These patients were randomised to
receive a cobalt–chrome on ultra-high-molecular-weight polyethylene,
cobalt–chrome on highly cross-linked polyethylene, or a ceramic-on-ceramic
bearing. In all, 97 hip replacements in 87 patients were available
for review at five years. Two hips had been revised, one for infection
and one for peri-prosthetic fracture. At the final follow-up there were no significant differences
between the groups for the mean Western Ontario and McMaster Universities
osteoarthritis index (pain, p = 0.543; function, p = 0.10; stiffness,
p = 0.99), Short Form-12 (physical component, p = 0.878; mental
component, p = 0.818) or Harris hip scores (p = 0.22). Radiological
outcomes revealed no significant wear in the
We prospectively assessed the efficacy of a ceramic-on-metal
(CoM) hip bearing with uncemented acetabular and femoral components
in which cobalt–chrome acetabular liners and alumina
Polyethylene wear debris can cause osteolysis
and the failure of total hip arthroplasty. We present the five-year
wear rates of a highly cross-linked polyethylene (X3) bearing surface
when used in conjunction with a 36 mm
We report two cases of surface deterioration of a zirconia
We report the results of our continued review of 14 hip arthroplasties using alumina
We report catastrophic early failure of a cemented total hip replacement comprising a modular femoral component with a Zirconia
We describe three prostheses with cemented titanium-alloy stems and Al. 2. O. 3.
This systematic review of the literature summarises
the clinical experience with ceramic-on-ceramic hip bearings over
the past 40 years and discusses the concerns that exist in relation
to the bearing combination. Loosening, fracture, liner chipping
on insertion, liner canting and dissociation, edge-loading and squeaking
have all been reported, and the relationship between these issues
and implant design and surgical technique is investigated. New design
concepts are introduced and analysed with respect to previous clinical
experience.
The inherent challenges of total hip replacement
(THR) in children include the choice of implant for the often atypical
anatomical morphology, its fixation to an immature growing skeleton
and the bearing surface employed to achieve a successful long-term
result. We report the medium-term results of 52 consecutive uncemented
THRs undertaken in
35 paediatric patients with juvenile idiopathic arthritis. The mean
age at the time of surgery was 14.4 years (10 to 16). The median
follow-up was 10.5 years (6 to 15). During the study period 13 THRs
underwent revision surgery. With revision as an endpoint, subgroup
analysis revealed 100% survival of the 23 ceramic-on-ceramic THRs
and 55% (16 of 29) of the metal- or ceramic-on-polyethylene. This
resulted in 94% (95% CI 77.8 to 98.4) survivorship of the femoral
component and 62% (95% CI 41.0 to 78.0) of the acetabular component.
Revision of the acetabular component for wear and osteolysis were
the most common reasons for failure accounting for 11 of the 13
revisions. The success seen in patients with a ceramic-on-ceramic articulation
seems to indicate that this implant strategy has the potential to
make a major difference to the long-term outcome in this difficult
group of patients.
We studied 33 third generation, alumina ceramic-on-ceramic bearings retrieved from cementless total hip replacements after more than six months in situ. Wear volume was measured with a Roundtest machine, and acetabular orientation from the anteroposterior pelvic radiograph. The overall median early wear rate was 0.1 mm3/yr for the femoral heads, and 0.04 mm3/yr for the acetabular liners. We then excluded hips where the components had migrated. In this stable subgroup of 22 bearings, those with an acetabular anteversion of <
15° (seven femoral heads) had a median femoral head wear rate of 1.2 mm3/yr, compared with 0 mm3/yr for those with an anteversion of ≥15° (15 femoral heads, p <
0.001). Even under edge loading, wear volumes with ceramic-on-ceramic bearings are small in comparison to other bearing materials. Low acetabular anteversion is associated with greater wear.
Squeaking arising from a ceramic-on-ceramic (CoC)
total hip replacement (THR) may cause patient concern and in some
cases causes patients to seek revision surgery. We performed a meta-analysis
to determine the incidence of squeaking and the incidence of revision
surgery for squeaking. A total of 43 studies including 16 828 CoC
THR that reported squeaking, or revision for squeaking, were entered
into the analysis. The incidence of squeaking was 4.2% and the incidence
of revision for squeaking was 0.2%. The incidence of squeaking in
patients receiving the Accolade femoral stem was 8.3%, and the incidence
of revision for squeaking in these patients was 1.3%. Cite this article:
Noise generation has been reported with ceramic-on-ceramic
articulations in total hip replacement (THR). This study evaluated
208 consecutive Delta Motion THRs at a mean follow-up of 21 months
(12 to 35). There were 141 women and 67 men with a mean age of 59
years (22 to 84). Patients were reviewed clinically and radiologically,
and the incidence of noise was determined using a newly described
assessment method. Noise production was examined against range of
movement, ligamentous laxity, patient-reported outcome scores, activity
level and orientation of the acetabular component. There were 143
silent hips (69%), 22 (11%) with noises other than squeaking, 17
(8%) with unreproducible squeaking and 26 (13%) with reproducible
squeaking. Hips with reproducible squeaking had a greater mean range
of movement (p <
0.001) and mean ligament laxity (p = 0.004), smaller
median head size (p = 0.01) and decreased mean acetabular component
inclination (p = 0.02) and anteversion angle (p = 0.02) compared
with the other groups. There was no relationship between squeaking
and age (p = 0.13), height (p = 0.263), weight (p = 0.333), body
mass index (p = 0.643), gender (p = 0.07) or patient outcome score
(p = 0.422). There were no revisions during follow-up. Despite the
surprisingly high incidence of squeaking, all patients remain satisfied
with their hip replacement. Cite this article:
Long-term clinical outcomes for ceramic-on-ceramic (CoC) bearings
are encouraging. However, there is a risk of squeaking. Guidelines
for the orientation of the acetabular component are defined from
static imaging, but the position of the pelvis and thus the acetabular
component during activities associated with edge-loading are likely
to be very different from those measured when the patient is supine.
We assessed the functional orientation of the acetabular component. A total of 18 patients with reproducible squeaking in their CoC
hips during deep flexion were investigated with a control group
of 36 non-squeaking CoC hips. The two groups were matched for the
type of implant, the orientation of the acetabular component when
supine, the size of the femoral head, ligament laxity, maximum hip
flexion and body mass index. Aims
Patients and Methods
Aims. This study reports the results of 38 total hip arthroplasties (THAs) in 33 patients aged less than 50 years, using the JRI Furlong hydroxyapatite
Aims. We conducted a prospective study of a delta
Large femoral heads have become popular in total
hip replacement (THR) as a method of reducing the risk of dislocation.
However, if large heads are used in ceramic-on-ceramic THR, the
liner must be thinner, which may increase the risk of fracture.
To compare the rates of
Aims. Ceramic-on-ceramic (CoC) bearings in total hip arthroplasty (THA)
are commonly used, but concerns exist regarding
Aims. To investigate the effect of polyethylene manufacturing characteristics and irradiation dose on the survival of cemented and reverse hybrid total hip arthroplasties (THAs). Methods. In this registry study, data from the National Joint Registry of England, Wales, Northern Ireland and the Isle of Man (NJR) were linked with manufacturing data supplied by manufacturers. The primary endpoint was revision of any component. Cox proportional hazard regression was a primary analytic approach adjusting for competing risk of death, patient characteristics, head composition, and stem fixation. Results. A total of 290,770 primary THAs were successfully linked with manufacturing characteristics. Overall 4,708 revisions were analyzed, 1,260 of which were due to aseptic loosening. Total radiation dose was identified as a risk factor and included in the Cox model. For statistical modelling of aseptic loosening, THAs were grouped into three categories: G1 (no radiation); G2 ( > 0 to < 5 Mrad); and G3 ( ≥ 5 Mrad). G1 had the worst survivorship. The Cox regression hazard ratio for revision due to aseptic loosening for G2 was 0.7 (95% confidence interval (CI) 0.58 to 0.83), and for G3 0.4 (95% CI 0.30 to 0.53). Male sex and uncemented stem fixation were associated with higher risk of revision and
Aims. Despite advances in the treatment of paediatric hip disease, adolescent and young adult patients can develop early onset end-stage osteoarthritis. The aims of this study were to address the indications and medium-term outcomes for total hip arthroplasty (THA) with
Alumina–alumina bearings are among the most resistant
to wear in total hip replacement. Examination of their surfaces
is one way of comparing damage caused by wear of hip joints simulated in
vitro to that seen in explanted bearings. The aim of this
study was to determine whether second-generation
Aims. Fretting and corrosion at the modular head/neck junction, known
as trunnionosis, in total hip arthroplasty (THA) is a cause of adverse
reaction to metal debris (ARMD). We describe the outcome of revision
of metal-on-polyethylene (MoP) THA for ARMD due to trunnionosis
with emphasis on the risk of major complications. Patients and Methods. A total of 36 patients with a MoP THA who underwent revision
for ARMD due to trunnionosis were identified. Three were excluded
as their revision had been to another metal head. The remaining
33 were revised to a
Radiostereometric analysis (RSA) is the most accurate radiological method to measure in vivo wear of highly cross-linked polyethylene (XLPE) acetabular components. We have previously reported very low wear rates for a sequentially irradiated and annealed X3 XLPE liner (Stryker Orthopaedics, USA) when used in conjunction with a 32 mm femoral heads at ten-year follow-up. Only two studies have reported the long-term wear rate of X3 liners used in conjunction with larger heads using plain radiographs which have poor sensitivity. The aim of this study was to measure the ten-year wear of thin X3 XLPE liners against larger 36 or 40 mm articulations with RSA. We prospectively reviewed 19 patients who underwent primary cementless THA with the XLPE acetabular liner (X3) and a 36 or 40 mm femoral head with a resultant liner thickness of at least 5.8 mm. RSA radiographs at one week, six months, and one, two, five, and ten years postoperatively and femoral head penetration within the acetabular component were measured with UmRSA software. Of the initial 19 patients, 12 were available at the ten-year time point.Aims
Methods
In 2015, we published the results of our ceramic-on-metal (CoM) total hip arthroplasties (THAs) performed between October 2007 and July 2009 with a mean follow-up of 34 months (23 to 45) and a revision rate of 3.1%. The aim of this paper is to present the longer-term outcomes. A total of 264 patients were reviewed at a mean of 5.8 years (4.6 to 7.2) and 10.1 years (9.2 to 10.6) to determine revision rate, pain, outcome scores, radiological analysis, and blood ion levels. Those who were unwilling or unable to travel were contacted by telephone.Aims
Methods
United Classification System (UCS) B2 and B3 periprosthetic fractures in total hip arthroplasties (THAs) have been commonly managed with modular tapered stems. No study has evaluated the use of monoblock fluted tapered titanium stems for this indication. This study aimed to evaluate the effects of a monoblock stems on implant survivorship, postoperative outcomes, radiological outcomes, and osseointegration following treatment of THA UCS B2 and B3 periprosthetic fractures. A retrospective review was conducted of all patients who underwent revision THA (rTHA) for periprosthetic UCS B2 and B3 periprosthetic fracture who received a single design monoblock fluted tapered titanium stem at two large, tertiary care, academic hospitals. A total of 72 patients met inclusion and exclusion criteria (68 UCS B2, and four UCS B3 fractures). Primary outcomes of interest were radiological stem subsidence (> 5 mm), radiological osseointegration, and fracture union. Sub-analysis was also done for 46 patients with minimum one-year follow-up.Aims
Methods
We retrospectively evaluated 42 hips which had undergone acetabular reconstruction using the Kerboull acetabular reinforcement device between September 1994 and December 1998. We used autogenous bone chips from the ilium and
Contemporary outcomes of primary total hip arthroplasties (THAs) with highly cross-linked polyethylene (HXLPE) liners in patients with inflammatory arthritis have not been well studied. This study examined the implant survivorship, complications, radiological results, and clinical outcomes of THA in patients with inflammatory arthritis. We identified 418 hips (350 patients) with a primary diagnosis of inflammatory arthritis who underwent primary THA with HXLPE liners from January 2000 to December 2017. Of these hips, 68% had rheumatoid arthritis (n = 286), 13% ankylosing spondylitis (n = 53), 7% juvenile rheumatoid arthritis (n = 29), 6% psoriatic arthritis (n = 24), 5% systemic lupus erythematosus (n = 23), and 1% scleroderma (n = 3). Mean age was 58 years (SD 14.8), 66.3% were female (n = 277), and mean BMI was 29 kg/m2 (SD 7). Uncemented femoral components were used in 77% of cases (n = 320). Uncemented acetabular components were used in all patients. Competing risk analysis was used accounting for death. Mean follow-up was 4.5 years (2 to 18).Aims
Methods
In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component. Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). Patients underwent follow-up of minimum two-years (mean 4.6 (2.1 to 8.5) years).Aims
Methods
Excellent outcomes have been reported following CT-based robotic arm-assisted total hip arthroplasty (rTHA) compared with manual THA; however, its superiority over CT-based navigation THA (nTHA) remains unclear. This study aimed to determine whether a CT-based robotic arm-assisted system helps surgeons perform accurate cup placement, minimizes leg length, and offsets discrepancies more than a CT-based navigation system. We studied 60 hips from 54 patients who underwent rTHA between April 2021 and August 2023, and 45 hips from 44 patients who underwent nTHA between January 2020 and March 2021 with the same target cup orientation at the Department of Orthopedic Surgery at Ozu Memorial Hospital, Japan. After propensity score matching, each group had 37 hips. Postoperative acetabular component position and orientation were measured using the planning module of the CT-based navigation system. Postoperative leg length and offset discrepancies were evaluated using postoperative CT in patients who have unilateral hip osteoarthritis.Aims
Methods
We present a retrospective series of 170 cemented titanium straight-stem femoral components combined with two types of femoral head: cobalt-chromium (CoCr) alloy (114 heads) and alumina
This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA). Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up.Aims
Methods
Limited implant survival due to aseptic cup loosening is most commonly responsible for revision total hip arthroplasty (THA). Advances in implant designs and materials have been crucial in addressing those challenges. Vitamin E-infused highly cross-linked polyethylene (VEPE) promises strong wear resistance, high oxidative stability, and superior mechanical strength. Although VEPE monoblock cups have shown good mid-term performance and excellent wear patterns, long-term results remain unclear. This study evaluated migration and wear patterns and clinical and radiological outcomes at a minimum of ten years’ follow-up. This prospective observational study investigated 101 cases of primary THA over a mean duration of 129 months (120 to 149). At last follow-up, 57 cases with complete clinical and radiological outcomes were evaluated. In all cases, the acetabular component comprised an uncemented titanium particle-coated VEPE monoblock cup. Patients were assessed clinically and radiologically using the Harris Hip Score, visual analogue scale (pain and satisfaction), and an anteroposterior radiograph. Cup migration and polyethylene wear were measured using Einzel-Bild-Röntgen-Analyze software. All complications and associated treatments were documented until final follow-up.Aims
Methods
Precise implant positioning, tailored to individual spinopelvic biomechanics and phenotype, is paramount for stability in total hip arthroplasty (THA). Despite a few studies on instability prediction, there is a notable gap in research utilizing artificial intelligence (AI). The objective of our pilot study was to evaluate the feasibility of developing an AI algorithm tailored to individual spinopelvic mechanics and patient phenotype for predicting impingement. This international, multicentre prospective cohort study across two centres encompassed 157 adults undergoing primary robotic arm-assisted THA. Impingement during specific flexion and extension stances was identified using the virtual range of motion (ROM) tool of the robotic software. The primary AI model, the Light Gradient-Boosting Machine (LGBM), used tabular data to predict impingement presence, direction (flexion or extension), and type. A secondary model integrating tabular data with plain anteroposterior pelvis radiographs was evaluated to assess for any potential enhancement in prediction accuracy.Aims
Methods
The aim of this study was to evaluate the suitability of the tapered cone stem in total hip arthroplasty (THA) in patients with excessive femoral anteversion and after femoral osteotomy. We included patients who underwent THA using Wagner Cone due to proximal femur anatomical abnormalities between August 2014 and January 2019 at a single institution. We investigated implant survival time using the endpoint of dislocation and revision, and compared the prevalence of prosthetic impingements between the Wagner Cone, a tapered cone stem, and the Taperloc, a tapered wedge stem, through simulation. We also collected Oxford Hip Score (OHS), visual analogue scale (VAS) satisfaction, and VAS pain by postal survey in August 2023 and explored variables associated with those scores.Aims
Methods
Dislocation is a major concern following total hip arthroplasty (THA) for osteoarthritis (OA). Both dual-mobility components and standard acetabular components with large femoral heads are used to reduce the risk of dislocation. We investigated whether dual-mobility components are superior to standard components in reducing the two-year dislocation and revision risk in a propensity-matched sample from the Danish Hip Arthroplasty Register (DHR). This population-based cohort study analyzed data from the DHR and the Danish National Patient Register. We included all patients undergoing primary THA for OA from January 2010 to December 2019 with either dual-mobility or standard acetabular components with metal-on-polyethylene or ceramic-on-polyethylene articulations with a 36 mm femoral head. The samples were propensity score-matched on patient and implant characteristics. The primary outcome was the difference in the absolute risk of dislocation within two years, with a secondary outcome of the difference in the absolute risk of revision surgery of any cause within the same timeframe. The cumulative incidence of dislocation was calculated using the Aalen-Johansen estimator, while the difference in absolute risk was estimated using absolute risk regression (ARR).Aims
Methods
The Exeter short stem was designed for patients with Dorr type A femora and short-term results are promising. The aim of this study was to evaluate the minimum five-year stem migration pattern of Exeter short stems in comparison with Exeter standard stems. In this case-control study, 25 patients (22 female) at mean age of 78 years (70 to 89) received cemented Exeter short stem (case group). Cases were selected based on Dorr type A femora and matched first by Dorr type A and then age to a control cohort of 21 patients (11 female) at mean age of 74 years (70 to 89) who received with cemented Exeter standard stems (control group). Preoperatively, all patients had primary hip osteoarthritis and no osteoporosis as confirmed by dual X-ray absorptiometry scanning. Patients were followed with radiostereometry for evaluation of stem migration (primary endpoint), evaluation of cement quality, and Oxford Hip Score. Measurements were taken preoperatively, and at three, 12, and 24 months and a minimum five-year follow-up.Aims
Methods
Aims. The Corail stem has good long-term results. After four years
of using this stem, we have detected a small group of patients who
have presented with symptomatic metaphyseal debonding. The aim of
this study was to quantify the incidence of this complication, to
delineate the characteristics of patients presenting with this complication
and to compare these patients with asymptomatic controls to determine
any important predisposing factors. Patients and Methods. Of 855 Corail collarless cementless stems implanted for osteoarthritis,
18 presented with symptomatic metaphyseal debonding. A control group
of 74 randomly selected patients was assembled. Clinical and radiological
parameters were measured and a logistic regression model was created
to evaluate factors associated with metaphyseal debonding. Results. The prevalence of this complication was 2.1% in our series. In
the multivariable model, the presence of a Dorr B-type proximal
femur was associated with metaphyseal debonding (odds ratio (OR)
10.73, 95% confidence interval (CI) 2.31 to 49.97, p = 0.002), as
was a body mass index >
25 kg/m. 2. (OR 6.85, 95% CI 1.06
to 44.28, p = 0.04). Smaller stems and the use of a polyethylene
acetabular liner appeared to be protective when compared with metal
and