Objectives. Intramedullary fixation is considered the most stable treatment for pertrochanteric fractures of the proximal femur and
Objectives. Because of the contradictory body of evidence related to the
potential benefits of helical blades in trochanteric fracture fixation,
we studied the effect of bone compaction resulting from the insertion
of a proximal femoral nail anti-rotation (PFNA). . Methods. We developed a subject-specific computational model of a trochanteric
fracture (31-A2 in the AO classification) with lack of medial support
and varied the bone density to account for variability in bone properties
among hip fracture patients. Results. We show that for a bone density corresponding to 100% of the
bone density of the cadaveric femur, there does not seem to be any
advantage in using a PFNA with respect to the risk of blade
Aims. Proximal femur fractures treatment can involve anterograde nailing with a single or double cephalic screw. An undesirable failure for this fixation is screw
Aims. Fixation of osteoporotic proximal humerus fractures remains challenging even with state-of-the-art locking plates. Despite the demonstrated biomechanical benefit of screw tip augmentation with bone cement, the clinical findings have remained unclear, potentially as the optimal augmentation combinations are unknown. The aim of this study was to systematically evaluate the biomechanical benefits of the augmentation options in a humeral locking plate using finite element analysis (FEA). Methods. A total of 64 cement augmentation configurations were analyzed using six screws of a locking plate to virtually fix unstable three-part fractures in 24 low-density proximal humerus models under three physiological loading cases (4,608 simulations). The biomechanical benefit of augmentation was evaluated through an established FEA methodology using the average peri-screw bone strain as a validated predictor of cyclic
The purpose of this study was to identify factors
that predict implant
The February 2023 Trauma Roundup. 360. looks at: Masquelet versus bone transport in infected nonunion of tibia; Hyperbaric Oxygen for Lower Limb Trauma (HOLLT): an international multicentre randomized clinical trial; Is the T-shaped acetabular fracture really a “T”?; What causes
Objectives. The Sliding Hip Screw (SHS) is commonly used to treat trochanteric hip fractures. Fixation failure is a devastating complication requiring complex revision surgery. One mode of fixation failure is lag screw
Objectives. The paradoxical migration of the femoral neck element (FNE) superomedially against gravity, with respect to the intramedullary component of the cephalomedullary device, is a poorly understood phenomenon increasingly seen in the management of pertrochanteric hip fractures with the intramedullary nail. The aim of this study was to investigate the role of bidirectional loading on the medial migration phenomenon, based on unique wear patterns seen on scanning electron microscopy of retrieved implants suggestive of FNE toggling. Methods. A total of 18 synthetic femurs (Sawbones, Vashon Island, Washington) with comminuted pertrochanteric fractures were divided into three groups (n = 6 per group). Fracture fixation was performed using the Proximal Femoral Nail Antirotation (PFNA) implant (Synthes, Oberdorf, Switzerland; n = 6). Group 1 was subjected to unidirectional compression loading (600 N), with an elastomer (70A durometer) replacing loose fracture fragments to simulate surrounding soft-tissue tensioning. Group 2 was subjected to bidirectional loading (600 N compression loading, 120 N tensile loading), also with the elastomer replacing loose fracture fragments. Group 3 was subjected to bidirectional loading (600 N compression loading, 120 N tensile loading) without the elastomer. All constructs were tested at 2 Hz for 5000 cycles or until
Aims. Plating displaced proximal humeral fractures is associated with a high rate of screw perforation. Dynamization of the proximal screws might prevent these complications. The aim of this study was to develop and evaluate a new gliding screw concept for plating proximal humeral fractures biomechanically. Methods. Eight pairs of three-part humeral fractures were randomly assigned for pairwise instrumentation using either a prototype gliding plate or a standard PHILOS plate, and four pairs were fixed using the gliding plate with bone cement augmentation of its proximal screws. The specimens were cyclically tested under progressively increasing loading until perforation of a screw. Telescoping of a screw, varus tilting and screw migration were recorded using optical motion tracking. Results. Mean initial stiffness (N/mm) was 581.3 (. sd. 239.7) for the gliding plate, 631.5 (. sd. 160.0) for the PHILOS and 440.2 (. sd. 97.6) for the gliding augmented plate without significant differences between the groups (p = 0.11). Mean varus tilting (°) after 7500 cycles was comparable between the gliding plate (2.6; . sd. 1.9), PHILOS (1.2; . sd. 0.6) and gliding augmented plate (1.7; . sd. 0.9) (p = 0.10). Similarly, mean screw migration(mm) after 7500 cycles was similar between the gliding plate (3.02; . sd. 2.85), PHILOS (1.30; . sd. 0.44) and gliding augmented plate (2.83; . sd. 1.18) (p = 0.13). Mean number of cycles until failure with 5° varus tilting were 12702 (. sd. 3687) for the gliding plate, 13948 (. sd. 1295) for PHILOS and 13189 (. sd. 2647) for the gliding augmented plate without significant differences between the groups (p = 0.66). Conclusion. Biomechanically, plate fixation using a new gliding screw technology did not show considerable advantages in comparison with fixation using a standard PHILOS plate. Based on the finding of telescoping of screws, however, it may represent a valid approach for further investigations into how to avoid the
The augmentation of fixation with bone cement
is increasingly being used in the treatment of severe osteoporotic fractures.
We investigated the influence of bone quality on the mechanics of
augmentation of plate fixation in a distal femoral fracture model
(AO 33 A3 type). Eight osteoporotic and eight non-osteoporotic femoral
models were randomly assigned to either an augmented or a non-augmented
group. Fixation was performed using a locking compression plate.
In the augmented group additionally 1 ml of bone cement was injected
into the screw hole before insertion of the screw. Biomechanical
testing was performed in axial sinusoidal loading. Augmentation significantly
reduced the
We investigated a new intramedullary locking
nail that allows the distal interlocking screws to be locked to
the nail. We compared fixation using this new implant with fixation
using either a conventional nail or a locking plate in a laboratory
simulation of an osteoporotic fracture of the distal femur. A total
of 15 human cadaver femora were used to simulate an AO 33-A3 fracture
pattern. Paired specimens compared fixation using either a locking
or non-locking retrograde nail, and using either a locking retrograde
nail or a locking plate. The constructs underwent cyclical loading
to simulate single-leg stance up to 125 000 cycles. Axial and torsional
stiffness and displacement, cycles to failure and modes of failure
were recorded for each specimen. When compared with locking plate
constructs, locking nail constructs had significantly longer mean
fatigue life (75 800 cycles (. sd. 33 900) vs 12
800 cycles (. sd. 6100); p = 0.007) and mean axial stiffness
(220 N/mm (. sd. 80) vs 70 N/mm (. sd. 18);
p = 0.005), but lower mean torsional stiffness (2.5 Nm/° (. sd. 0.9) vs 5.1
Nm/° (. sd. 1.5); p = 0.008). In addition, in the nail group
the mode of failure was either
To evaluate if, for orthopaedic trainees, additional cadaveric simulation training or standard training alone yields superior radiological and clinical outcomes in patients undergoing dynamic hip screw (DHS) fixation or hemiarthroplasty for hip fracture. This was a preliminary, pragmatic, multicentre, parallel group randomized controlled trial in nine secondary and tertiary NHS hospitals in England. Researchers were blinded to group allocation. Overall, 40 trainees in the West Midlands were eligible: 33 agreed to take part and were randomized, five withdrew after randomization, 13 were allocated cadaveric training, and 15 were allocated standard training. The intervention was an additional two-day cadaveric simulation course. The control group received standard on-the-job training. Primary outcome was implant position on the postoperative radiograph: tip-apex distance (mm) (DHS) and leg length discrepancy (mm) (hemiarthroplasty). Secondary clinical outcomes were procedure time, length of hospital stay, acute postoperative complication rate, and 12-month mortality. Procedure-specific secondary outcomes were intraoperative radiation dose (for DHS) and postoperative blood transfusion requirement (hemiarthroplasty).Aims
Methods
We studied the biomechanical behaviour of three sliding fixation devices for trochanteric femoral fractures. These were a titanium alloy sideplate and lag screw, a titanium alloy sideplate and dome plunger with cement augmentation, and a stainless-steel sideplate and lag screw. We used 18 mildly osteoporotic cadaver femora, randomly assigned to one of the three fixation groups. Four displacement and two strain gauges were fixed to each specimen, and each femur was first tested intact (control), then as a two-part fracture and then as a four-part intertrochanteric fracture. A range of physiological loads was applied to determine load-bearing, load-sharing and head displacement. The four-part-fracture specimens were subsequently tested to failure to determine maximum fixation strengths and modes of failure. The dome-plunger group failed at a load 50% higher than that of the stainless-steel lag-screw group (p <
0.05) and at a load 20% higher than that of the titanium-alloy lag-screw group (NS). All 12 lag-screw specimens failed by
The management of fractures of the medial epicondyle is one of the greatest controversies in paediatric fracture care, with uncertainty concerning the need for surgery. The British Society of Children’s Orthopaedic Surgery prioritized this as their most important research question in paediatric trauma. This is the protocol for a randomized controlled, multicentre, prospective superiority trial of operative fixation versus nonoperative treatment for displaced medial epicondyle fractures: the Surgery or Cast of the EpicoNdyle in Children’s Elbows (SCIENCE) trial. Children aged seven to 15 years old inclusive, who have sustained a displaced fracture of the medial epicondyle, are eligible to take part. Baseline function using the Patient-Reported Outcomes Measurement Information System (PROMIS) upper limb score, pain measured using the Wong Baker FACES pain scale, and quality of life (QoL) assessed with the EuroQol five-dimension questionnaire for younger patients (EQ-5D-Y) will be collected. Each patient will be randomly allocated (1:1, stratified using a minimization algorithm by centre and initial elbow dislocation status (i.e. dislocated or not-dislocated at presentation to the emergency department)) to either a regimen of the operative fixation or non-surgical treatment.Aims
Methods
Total hip arthroplasty (THA) is considered the preferred treatment for displaced proximal femoral neck fractures. However, in many countries this option is economically unviable. To improve outcomes in financially disadvantaged populations, we studied the technique of concomitant valgus hip osteotomy and operative fixation (VOOF). This prospective serial study compares two treatment groups: VOOF versus operative fixation alone with cannulated compression screws (CCSs). In the first series, 98 hip fixation procedures were performed using CCS. After fluoroscopic reduction of the fracture, three CCSs were placed. In the second series, 105 VOOF procedures were performed using a closing wedge intertrochanteric osteotomy with a compression lag screw and lateral femoral plate. The alignment goal was to create a modified Pauwel’s fracture angle of 30°. After fluoroscopic reduction of fracture, lag screw was placed to achieve the calculated correction angle, followed by inter-trochanteric osteotomy and placement of barrel plate. Patients were followed for a minimum of two years.Aims
Methods
The Peri-Implant and PeriProsthetic Survival AnalysiS (PIPPAS) study aimed to investigate the risk factors for one-year mortality of femoral peri-implant fractures (FPIFs). This prospective, multicentre, observational study involved 440 FPIF patients with a minimum one-year follow-up. Data on demographics, clinical features, fracture characteristics, management, and mortality rates were collected and analyzed using both univariate and multivariate analyses. FPIF patients were elderly (median age 87 years (IQR 81 to 92)), mostly female (82.5%, n = 363), and frail: median clinical frailty scale 6 (IQR 4 to 7), median Pfeiffer 4 (1 to 7), median age-adjusted Charlson Comorbidity Index (CCI) 6 (IQR 5 to 7), and 58.9% (n = 250) were American Society of Anesthesiologists grade III.Aims
Methods
The aim of this study was to investigate if there are differences in outcome between sliding hip screws (SHSs) and intramedullary nails (IMNs) with regard to fracture stability. We assessed data from 17,341 patients with trochanteric or subtrochanteric fractures treated with SHS or IMN in the Norwegian Hip Fracture Register from 2013 to 2019. Primary outcome measures were reoperations for stable fractures (AO Foundation/Orthopaedic Trauma Association (AO/OTA) type A1) and unstable fractures (AO/OTA type A2, A3, and subtrochanteric fractures). Secondary outcome measures were reoperations for A2, A3, and subtrochanteric fractures individually, one-year mortality, quality of life (EuroQol five-dimension three-level index score), pain (visual analogue scale (VAS)), and satisfaction (VAS) for stable and unstable fractures. Hazard rate ratios (HRRs) for reoperation were calculated using Cox regression analysis with adjustments for age, sex, and American Society of Anesthesiologists score.Aims
Methods
We have prospectively compared the fixation of 100 intertrochanteric fractures of the proximal femur in elderly patients with random use of either a Dynamic Hip Screw (DHS) or a new intramedullary device, the Gamma nail. We found no difference in operating time, blood loss, wound complications, stay in hospital, place of eventual discharge, or the patients' mobility at final review. There was no difference in failure of proximal fixation:
In a prospective study we assessed the causes of mechanical failure in a series of 230 intertrochanteric femoral fractures which had been internally fixed with either a sliding hip screw or a Kuntscher Y-nail. The overall rate of mechanical failure was 16.5%; cutting-out of the implant from the femoral head was the cause in three-quarters of the instances. Implants placed posteriorly in the femoral head cut out more often (27%) than those placed centrally (7%). The