Advertisement for orthosearch.org.uk
Results 1 - 15 of 15
Results per page:
The Bone & Joint Journal
Vol. 96-B, Issue 8 | Pages 1070 - 1076
1 Aug 2014
Hannemann PFW van Wezenbeek MR Kolkman KA Twiss ELL Berghmans CHJ Dirven PAMGM Brink PRG Poeze M

We hypothesised that the use of pulsed electromagnetic field (PEMF) bone growth stimulation in acute scaphoid fractures would significantly shorten the time to union and reduce the number of nonunions in a randomised, double-blind, placebo-controlled multicentre trial. A total of 102 patients (78 male, 24 female; mean age 35 years (18 to 77)) from five different medical centres with a unilateral undisplaced acute scaphoid fracture were randomly allocated to PEMF (n = 51) or placebo (n = 51) and assessed with regard to functional and radiological outcomes (multiplanar reconstructed CT scans) at 6, 9, 12, 24 and 52 weeks. The overall time to clinical and radiological healing did not differ significantly between the active PEMF group and the placebo group. We concluded that the addition of PEMF bone growth stimulation to the conservative treatment of acute scaphoid fractures does not accelerate bone healing. Cite this article: Bone Joint J 2014; 96-B:1070–6


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1403 - 1408
1 Oct 2012
Hannemann PFW Göttgens KWA van Wely BJ Kolkman KA Werre AJ Poeze M Brink PRG

The use of pulsed electromagnetic fields (PEMF) to stimulate bone growth has been recommended as an alternative to the surgical treatment of ununited scaphoid fractures, but has never been examined in acute fractures. We hypothesised that the use of PEMF in acute scaphoid fractures would accelerate the time to union by 30% in a randomised, double-blind, placebo-controlled, multicentre trial. A total of 53 patients in three different medical centres with a unilateral undisplaced acute scaphoid fracture were randomly assigned to receive either treatment with PEMF (n = 24) or a placebo (n = 29). The clinical and radiological outcomes were assessed at four, six, nine, 12, 24 and 52 weeks. A log-rank analysis showed that neither time to clinical and radiological union nor the functional outcome differed significantly between the groups. The clinical assessment of union indicated that at six weeks tenderness in the anatomic snuffbox (p = 0.03) as well as tenderness on longitudinal compression of the scaphoid (p = 0.008) differed significantly in favour of the placebo group. We conclude that stimulation of bone growth by PEMF has no additional value in the conservative treatment of acute scaphoid fractures


Bone & Joint Research
Vol. 10, Issue 12 | Pages 767 - 779
8 Dec 2021
Li Y Yang Y Wang M Zhang X Bai S Lu X Li Y Waldorff EI Zhang N Lee WY Li G

Aims. Distraction osteogenesis (DO) is a useful orthopaedic procedure employed to lengthen and reshape bones by stimulating bone formation through controlled slow stretching force. Despite its promising applications, difficulties are still encountered. Our previous study demonstrated that pulsed electromagnetic field (PEMF) treatment significantly enhances bone mineralization and neovascularization, suggesting its potential application. The current study compared a new, high slew rate (HSR) PEMF signal, with different treatment durations, with the standard Food and Drug Administration (FDA)-approved signal, to determine if HSR PEMF is a better alternative for bone formation augmentation. Methods. The effects of a HSR PEMF signal with three daily treatment durations (0.5, one, and three hours/day) were investigated in an established rat DO model with comparison of an FDA-approved classic signal (three hrs/day). PEMF treatments were applied to the rats daily for 35 days, starting from the distraction phase until termination. Radiography, micro-CT (μCT), biomechanical tests, and histological examinations were employed to evaluate the quality of bone formation. Results. All rats tolerated the treatment well and no obvious adverse effects were found. By comparison, the HSR signal (three hrs/day) treatment group achieved the best healing outcome, in that endochondral ossification and bone consolidation were enhanced. In addition, HSR signal treatment (one one hr/day) had similar effects to treatment using the classic signal (three three hrs/day), indicating that treatment duration could be significantly shortened with the HSR signal. Conclusion. HSR signal may significantly enhance bone formation and shorten daily treatment duration in DO, making it a potential candidate for a new clinical protocol for patients undergoing DO treatments. Cite this article: Bone Joint Res 2021;10(12):767–779


Bone & Joint Research
Vol. 10, Issue 5 | Pages 298 - 306
1 May 2021
Dolkart O Kazum E Rosenthal Y Sher O Morag G Yakobson E Chechik O Maman E

Aims. Rotator cuff (RC) tears are common musculoskeletal injuries which often require surgical intervention. Noninvasive pulsed electromagnetic field (PEMF) devices have been approved for treatment of long-bone fracture nonunions and as an adjunct to lumbar and cervical spine fusion surgery. This study aimed to assess the effect of continuous PEMF on postoperative RC healing in a rat RC repair model. Methods. A total of 30 Wistar rats underwent acute bilateral supraspinatus tear and repair. A miniaturized electromagnetic device (MED) was implanted at the right shoulder and generated focused PEMF therapy. The animals’ left shoulders served as controls. Biomechanical, histological, and bone properties were assessed at three and six weeks. Results. Extension of the tendon from preload to the maximum load to failure was significantly better in the PEMF-treated shoulders at three weeks compared to controls (p = 0.038). The percentage strain was significantly higher in the PEMF group at both timepoints (p = 0.037). Collagen organization was significantly better (p = 0.034) as was tissue mineral density in the PEMF-treated group at three weeks (p = 0.028). Tendon immunohistochemistry revealed a prominent increase in type I collagen at the repair site at three weeks following continuous PEMF treatment compared with controls. None of the other tested parameters differed between the groups. Conclusion. MED-generated PEMF may enhance early postoperative tendon-to-bone healing in an acute rat supraspinatus detachment and repair model. Superior biomechanical elasticity parameters together with better collagen organization suggest improved RC healing. Cite this article: Bone Joint Res 2021;10(5):298–306


Bone & Joint Research
Vol. 7, Issue 2 | Pages 124 - 130
1 Feb 2018
Coric D Bullard DE Patel VV Ryaby JT Atkinson BL He D Guyer RD

Objectives. Pulsed electromagnetic field (PEMF) stimulation was evaluated after anterior cervical discectomy and fusion (ACDF) procedures in a randomized, controlled clinical study performed for United States Food and Drug Administration (FDA) approval. PEMF significantly increased fusion rates at six months, but 12-month fusion outcomes for subjects at elevated risk for pseudoarthrosis were not thoroughly reported. The objective of the current study was to evaluate the effect of PEMF treatment on subjects at increased risk for pseudoarthrosis after ACDF procedures. Methods. Two evaluations were performed that compared fusion rates between PEMF stimulation and a historical control (160 subjects) from the FDA investigational device exemption (IDE) study: a post hoc (PH) analysis of high-risk subjects from the FDA study (PH PEMF); and a multicentre, open-label (OL) study consisting of 274 subjects treated with PEMF (OL PEMF). Fisher’s exact test and multivariate logistic regression was used to compare fusion rates between PEMF-treated subjects and historical controls. Results. In separate comparisons of PH PEMF and OL PEMF groups to the historical control group, PEMF treatment significantly (p < 0.05, Fisher’s exact test) increased the fusion rate at six and 12 months for certain high-risk subjects who had at least one clinical risk factor of being elderly, a nicotine user, osteoporotic, or diabetic; and for those with at least one clinical risk factor and who received at least a two- or three-level arthrodesis. Conclusion. Adjunctive PEMF treatment can be recommended for patients who are at high risk for pseudoarthrosis. Cite this article: D. Coric, D. E. Bullard, V. V. Patel, J. T. Ryaby, B. L. Atkinson, D. He, R. D. Guyer. Pulsed electromagnetic field stimulation may improve fusion rates in cervical arthrodesis in high-risk populations. Bone Joint Res 2018;7:124–130. DOI: 10.1302/2046-3758.72.BJR-2017-0221.R1


Bone & Joint Research
Vol. 9, Issue 9 | Pages 613 - 622
1 Sep 2020
Perucca Orfei C Lovati AB Lugano G Viganò M Bottagisio M D’Arrigo D Sansone V Setti S de Girolamo L

Aims. In the context of tendon degenerative disorders, the need for innovative conservative treatments that can improve the intrinsic healing potential of tendon tissue is progressively increasing. In this study, the role of pulsed electromagnetic fields (PEMFs) in improving the tendon healing process was evaluated in a rat model of collagenase-induced Achilles tendinopathy. Methods. A total of 68 Sprague Dawley rats received a single injection of type I collagenase in Achilles tendons to induce the tendinopathy and then were daily exposed to PEMFs (1.5 mT and 75 Hz) for up to 14 days - starting 1, 7, or 15 days after the injection - to identify the best treatment option with respect to the phase of the disease. Then, 7 and 14 days of PEMF exposure were compared to identify the most effective protocol. Results. The daily exposure to PEMFs generally provided an improvement in the fibre organization, a decrease in cell density, vascularity, and fat deposition, and a restoration of the physiological cell morphology compared to untreated tendons. These improvements were more evident when the tendons were exposed to PEMFs during the mid-acute phase of the pathology (7 days after induction) rather than during the early (1 day after induction) or the late acute phase (15 days after induction). Moreover, the exposure to PEMFs for 14 days during the mid-acute phase was more effective than for 7 days. Conclusion. PEMFs exerted a positive role in the tendon healing process, thus representing a promising conservative treatment for tendinopathy, although further investigations regarding the clinical evaluation are needed. Cite this article: Bone Joint Res 2020;9(9):613–622


Bone & Joint Research
Vol. 6, Issue 5 | Pages 323 - 330
1 May 2017
Pijls BG Sanders IMJG Kuijper EJ Nelissen RGHH

Objectives. Infection of implants is a major problem in elective and trauma surgery. Heating is an effective way to reduce the bacterial load in food preparation, and studies on hyperthermia treatment for cancer have shown that it is possible to heat metal objects with pulsed electromagnetic fields selectively (PEMF), also known as induction heating. We therefore set out to answer the following research question: is non-contact induction heating of metallic implants effective in reducing bacterial load in vitro?. Methods. Titanium alloy cylinders (Ti6Al4V) were exposed to PEMF from an induction heater with maximum 2000 watts at 27 kHz after being contaminated with five different types of micro-organisms: Staphylococcus epidermidis; Staphylococcus aureus; Pseudomonas aeruginosa; spore-forming Bacillus cereus; and yeast Candida albicans. The cylinders were exposed to incremental target temperatures (35°C, 45°C, 50°C, 55°C, 60°C, 65°C, 70°C) for up to 3.5 minutes. Results. There was an average linear heating rate of 0.39°C per second up to the target temperature, and thereafter the target temperature was maintained until the end of the experiment. At 60°C and higher (duration 3.5 minutes), there was a 6-log reduction or higher for every micro-organism tested. At 60°C, we found that the shortest duration of effective induction heating was 1.5 minutes. This resulted in a 5-log reduction or higher for every micro-organism tested. Conclusion. Non-contact induction heating of a titanium disk is effective in reducing bacterial load in vitro. These promising results can be further explored as a new treatment modality for infections of metal orthopaedic implants. Cite this article: B. G. Pijls, I. M. J. G. Sanders, E. J. Kuijper, R. G. H. H. Nelissen. Non-contact electromagnetic induction heating for eradicating bacteria and yeasts on biomaterials and possible relevance to orthopaedic implant infections: In vitro findings. Bone Joint Res 2017;6:323–330. DOI: 10.1302/2046-3758.65.BJR-2016-0308.R1


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 4 | Pages 588 - 593
1 May 2003
Pickering SAW Bayston R Scammell BE

Infection of orthopaedic implants is a significant problem, with increased antibiotic resistance of adherent ‘biofilm’ bacteria causing difficulties in treatment. We have investigated the in vitro effect of a pulsed electromagnetic field (PEMF) on the efficacy of antibiotics in the treatment of infection of implants. Five-day biofilms of Staphylococcus epidermidis were grown on the tips of stainless-steel pegs. They were exposed for 12 hours to varying concentrations of gentamicin or vancomycin in microtitre trays at 37°C and 5% CO. 2. The test group were exposed to a PEMF. The control tray was not exposed to a PEMF. After exposure to antibiotic the pegs were incubated overnight, before standard plating onto blood agar for colony counting. Exposure to a PEMF increased the effectiveness of gentamicin against the five-day biofilms of Staphylococcus epidermidis. In three of five experiments there was reduction of at least 50% in the minimum biofilm inhibitory concentration. In a fourth experiment there was a two-log difference in colony count at 160 mg/l of gentamicin. Analysis of variance (ANOVA) confirmed an effect by a PEMF on the efficacy of gentamicin which was significant at p < 0.05. There was no significant effect with vancomycin


The Journal of Bone & Joint Surgery British Volume
Vol. 65-B, Issue 4 | Pages 478 - 492
1 Aug 1983
Raji A Bowden R

Apart from preliminary notices of present work, previous reports of experimental and clinical trials of the effects of a high-peak pulsed electromagnetic field (PEMF) on degeneration and regeneration of peripheral nerves lacked statistical analysis. Therefore, we designed experiments with standardised operative, histological, cytological and morphometric techniques to assess the effect of PEMF on lesions of the common peroneal nerves in paired male rats matched for age, environmental conditions and level and type of lesion. One of two types of lesion was induced in the left common peroneal nerve: in 12 pairs of rats the nerve was crushed just above the knee and in the remaining 12 pairs the nerve was cut and immediately sutured at the same level. The right common peroneal nerve of each rat served as a control. Animals received 15 minutes of PEMF produced by a Diapulse machine or sham treatment daily for periods ranging from three and a half days to eight weeks after injury. Healthy nerves were unaffected, but after damage there were statistically significant differences between PEMF treated and sham treated rats. PEMF accelerated the recovery of injured limbs and the degeneration, regeneration and maturation of myelinated axons; epineural, perineural and intraneural fibrosis was reduced; and the luminal cross-sectional area of intraneural vessels increased after both types of lesion. Findings are discussed and the need for clinical trials is stressed


Bone & Joint Research
Vol. 11, Issue 9 | Pages 629 - 638
1 Sep 2022
Pijls BG Sanders IMJG Kuijper EJ Nelissen RGHH

Aims

Here we used a mature seven-day biofilm model of Staphylococcus aureus, exposed to antibiotics up to an additional seven days, to establish the effectiveness of either mechanical cleaning or antibiotics or non-contact induction heating, and which combinations could eradicate S. aureus in mature biofilms.

Methods

Mature biofilms of S. aureus (ATCC 29213) were grown on titanium alloy (Ti6Al4V) coupons for seven days and were subjected to the following treatments or their combinations: antibiotics, mechanical cleaning, or heat shock by induction heating of 60°C for one minute. Experiments were repeated at least five times.


Bone & Joint Research
Vol. 9, Issue 4 | Pages 192 - 199
1 Apr 2020
Pijls BG Sanders IMJG Kujiper EJ Nelissen RGHH

Aims

Induction heating is a noninvasive, nonantibiotic treatment modality that can potentially be used to cause thermal damage to the bacterial biofilm on the metal implant surface. The purpose of this study was to determine the effectiveness of induction heating on killing Staphylococcus epidermidis from biofilm and to determine the possible synergistic effect of induction heating and antibiotics.

Methods

S. epidermidis biofilms were grown on titanium alloy (Ti6Al4V) coupons for 24 hours (young biofilm) and seven days (mature biofilm). These coupons with biofilm were heated to temperatures of 50°C, 55°C, 60°C, 65°C, 70°C, 80°C, and 90°C for 3.5 minutes and subsequently exposed to vancomycin and rifampicin at clinically relevant concentrations.


Bone & Joint Research
Vol. 7, Issue 11 | Pages 609 - 619
1 Nov 2018
Pijls BG Sanders IMJG Kuijper EJ Nelissen RGHH

Objectives

Prosthetic joint infection (PJI) is a devastating complication following total joint arthroplasty. Non-contact induction heating of metal implants is a new and emerging treatment for PJI. However, there may be concerns for potential tissue necrosis. It is thought that segmental induction heating can be used to control the thermal dose and to limit collateral thermal injury to the bone and surrounding tissues. The purpose of this study was to determine the thermal dose, for commonly used metal implants in orthopaedic surgery, at various distances from the heating centre (HC).

Methods

Commonly used metal orthopaedic implants (hip stem, intramedullary nail, and locking compression plate (LCP)) were heated segmentally using an induction heater. The thermal dose was expressed in cumulative equivalent minutes at 43°C (CEM43) and measured with a thermal camera at several different distances from the HC. A value of 16 CEM43 was used as the threshold for thermal damage in bone.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1587 - 1594
1 Dec 2013
Ibrahim MS Twaij H Giebaly DE Nizam I Haddad FS

The outcome after total hip replacement has improved with the development of surgical techniques, better pain management and the introduction of enhanced recovery pathways. These pathways require a multidisciplinary team to manage pre-operative education, multimodal pain control and accelerated rehabilitation. The current economic climate and restricted budgets favour brief hospitalisation while minimising costs. This has put considerable pressure on hospitals to combine excellent results, early functional recovery and shorter admissions.

In this review we present an evidence-based summary of some common interventions and methods, including pre-operative patient education, pre-emptive analgesia, local infiltration analgesia, pre-operative nutrition, the use of pulsed electromagnetic fields, peri-operative rehabilitation, wound dressings, different surgical techniques, minimally invasive surgery and fast-track joint replacement units.

Cite this article: Bone Joint J 2013;95-B:1587–94.


Bone & Joint 360
Vol. 3, Issue 6 | Pages 16 - 17
1 Dec 2014

The December 2014 Foot & Ankle Roundup360 looks at: Charcot feet, biomarkers and diabetes; weight bearing following Achilles tendon rupture; endobuttons and mal-reduced diastasis; evidence for stem cell therapies in osteochondral lesions of the talus; syndesmosis fixation in SER ankle fractures; and self-reporting for foot and ankle outcomes.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 123 - 128
1 Jan 2006
Fini M Giavaresi G Giardino R Cavani F Cadossi R

We investigated the effect of stimulation with a pulsed electromagnetic field on the osseointegration of hydroxyapatite in cortical bone in rabbits. Implants were inserted into femoral cortical bone and were stimulated for six hours per day for three weeks.

Electromagnetic stimulation improved osseointegration of hydroxyapatite compared with animals which did not receive this treatment in terms of direct contact with the bone, the maturity of the bone and mechanical fixation. The highest values of maximum push-out force (Fmax) and ultimate shear strength (σu) were observed in the treated group and differed significantly from those of the control group at three weeks (Fmax; p < 0.0001; σu, p < 0.0005).