Abstract
Aims
The diagnosis of periprosthetic joint infection (PJI) remains a challenge, as no single diagnostic test shows high diagnostic accuracy. Recently, the measurement of synovial biomarkers has shown promising results. The aim of this study was to present a novel multiplex micro-enzyme-linked immunosorbent assay (ELISA) method for the rapid and simultaneous measurement of alpha-defensin, interleukin (IL)-6, and calprotectin developed in a model buffer system and human spiked synovial fluid.
Methods
A microfluidics- and chemiluminescence-based multiplex micro-ELISA point-of-care testing system was employed for the rapid and simultaneous measurement of alpha-defensin, calprotectin, and IL-6 developed in a model buffer system and spiked human synovial fluid. Cut-off values of 1.56 µg/ml for alpha-defensin, 50 µg/ml for calprotectin, and 0.031 µg/ml for IL-6 were extracted from the literature as optimal cut-off values for the diagnosis of PJI, and were used for comparison.
Results
Limit of detection (LoD) was determined for each individual biomarker by means of calibration curves with serial dilutions in a model buffer system. LoDs of 0.008, 0.002, and 0.00014 µg/ml were determined for alpha-defensin, calprotectin, and IL-6, respectively. The spiked synovial fluid assay determined LoDs of 0.08, 0.31, and 0.004 µg/ml for alpha-defensin, calprotectin, and IL-6, respectively.
Conclusion
These findings highlight the proposed platform’s unique features, including simultaneous measurement of three key synovial biomarkers, minimal sample volume requirements (5 to 50 µl), lower LoDs compared to conventional tests, and a short processing turnaround time of 22 minutes. Further validation studies are necessary to confirm its clinical utility.
Cite this article: Bone Joint Res 2025;14(3):176–184.
Article focus
-
To present a novel multiplex micro-enzyme-linked immunosorbent assay (ELISA) for the rapid and simultaneous measurement of biomarkers developed in a model buffer system and human spiked synovial fluid.
Key messages
-
Simultaneous quantification of three key synovial biomarkers (alpha-defensin, calprotectin, and interleukin (IL)-6).
-
Lower LoDs compared to conventional tests and speed of processing can give this novel multiplex µ-ELISA platform great diagnostic potential as a point-of-care tool.
Strengths and limitations
-
Small sample volume requirements (5 to 50 µl).
-
Lower limits of detection compared to conventional tests.
-
Need for validation with clinical studies.
Introduction
Periprosthetic joint infection (PJI) is a rare but serious complication after joint arthroplasty.1,2 The diagnosis of PJI remains a challenge, especially in low-grade and culture-negative cases, as a diagnostic test with high accuracy is lacking. An accurate diagnosis is essential to guide appropriate surgical and antibiotic therapy.3 Biomarkers represent a set of diagnostic tools for monitoring a biological process or guiding therapeutic decisions. They can be measured and assessed as indicators of biological processes, disease agents, or therapeutic responses to treatment interventions.4 In the setting of infection, biomarkers are widely employed to enhance the accuracy and timing of diagnosis, assess prognosis, and optimize the choice of tailor-made treatment.5 The research for synovial biomarkers that can aid in the diagnosis of PJI is an ongoing field, and several molecules are currently under investigation.6-8 Of these, alpha-defensin, calprotectin, and interleukin (IL)-6 seem to be the most promising.9 The clinical gold standard for protein biomarker analysis is enzyme-linked immunosorbent assay (ELISA); however, standard ELISA assays are time-consuming (two to four hours) and do not allow random access testing at reasonable costs. Lateral flow immunoassay, mostly based on chromatography, has recently been proposed as a possible alternative to standard ELISA in point-of-care settings due to its ease of use and quick detection.10 Microfluidic chip-based ELISA methods present another innovative option, combining the ease of use and swift results of lateral flow assays with the precision and reliability of traditional ELISA.11-13
Despite extensive investigations, so far, no single biomarker test could provide a definitive and accurate diagnosis of PJI. On the other hand, the approach of merging multiple biomarkers into a single panel seemed to be more successful in ensuring improvement in diagnostic accuracy.14,15 Moreover, according to recent findings, the diagnostic potential increases if the association involves biomarkers belonging to different categories (cytokines, inflammatory proteins, and antimicrobial peptides).16
The aim of this study was to present a novel multiplex micro-ELISA (μ-ELISA) method for the rapid and simultaneous measurement of alpha-defensin, IL-6, and calprotectin developed in a model buffer system and human spiked synovial fluid.
Methods
The multiplex μ-ELISA system
The commercial rapid test system (GENSPEED Biotech, Austria) was used to detect alpha-defensin, calprotectin, and IL-6 on prototype test chips (Figure 1). This technology is a point-of-care chemiluminescence-based multiplex μ-ELISA system that can be employed in various applications.
Fig. 1
Analyzer and microfluidic test chip used in this study (for details refer to “Methods”).
The test system is based on proprietary technology that combines microfluidic test chips with integrated microarrays, miniaturized optoelectronics, and automated assay procedures. The test chip consists of an injection-moulded top part containing the microfluidic structures and a bottom foil (both polystyrene) onto which the capture antibodies are deposited (Figure 2).
Fig. 2
Left panel: Layout of capture antibodies immobilized on the multiplex microfluidic test chip for the respective biomarke. Sample and reagents enter through the chip inlet, pass through a channel with capture antibodies, and exit into the waste reservoir. Right panel: side view showing the sandwich assay setup for each biomarker: immobilized antibodies captured the biomarker, biotinylated detection antibodies bind, and streptavidin-horseradish peroxidase (HRP) interaction enables chemiluminescence detection. A biotinylated control antibody checks the assay’s workflow. α-D, alpha-defensin; CP, calprotecin; Fluo, fluorescein; IL-6, interleukin 6.
The immobilization of the capture antibodies (Table I) in the corresponding capture areas (1 × 2 mm) is achieved using a sciFLEXARRAYER S3 (SCIENION, Germany) by printing a pattern of spots (diameter 200 μm, pitch 500 μm) using proprietary printing buffers. Finally, the microfluidic top part and the bottom foil with the immobilized capture antibodies are connected by ultrasonic welding.
Table I.
List of materials used for the detection of human alpha-defensin, calprotectin, and interleukin-6 with the GENSPEED microfluidic test chip.
Biomarker | Capture antibody | Detection antibody | Antigen/Standard |
---|---|---|---|
Alpha-defensin | Human alpha-defensin 1 DuoSet ELISA (R&D Systems, USA) | Human alpha-defensin 1 DuoSet ELISA (R&D Systems) | Recombinant human alpha-defensin 1 protein (Abcam) |
Calprotectin | Anti-S100A9+ Calprotectin (S100A8/A9 complex) antibody (Abcam, UK) | Human S100A8 DuoSet ELISA (R&D Systems) | Recombinant Human S100A8/S100A9 (R&D Systems) |
IL-6 | Human IL-6 DuoSet ELISA (R&D Systems) | Human IL-6 DuoSet ELISA (R&D Systems) | Recombinant Human IL-6 (R&D Systems) |
-
ELISA, enzyme-linked immunosorbent assay; IL-6, interleukin-6.
The design of the microfluidic test chip, which includes an inlet port, a reaction channel, and a waste reservoir, facilitates a sequential, capillary force-driven flow of microlitre volumes of sample and reagents. After addition to the inlet, the fluids are transported through the channel by capillary action, halting when the inlet is empty and resuming with the introduction of new fluids, thereby displacing the preceding solution into the waste reservoir.
Before analysis, a sample filtration is performed, followed by applying a drop of the sample to the test chip’s inlet port. After incubation, a drop of biotinylated detection antibody solution is added (Table I). The GENSPEED R2 analyzer then automatically dispenses all necessary reagents for the multiplex μ-ELISA using micro-piezo pumps (mp6; Bartels Mikrotechnik, Germany), with results ready in 15 to 22 minutes, as shown in a previous study.17
For data readout, the chemiluminescence reaction’s emitted photons are captured by a custom photodiode array located beneath the test chip’s reaction channel. The one-dimensional array, with a spatial resolution of 1 mm, comprises 32 pixels and features a low-noise equivalent power (10-15 W/√Hz), and is used in combination with an appropriate analogue digital converter to integrate chemiluminescence signals over time. This setup has been successfully applied in standard multiplexing assays to detect various biomarkers or pathogens in a single test across different sample types, such as blood or swab materials. Examples include nucleic acid-based tests for identifying antibiotic resistance,18 and serological assays for detecting various immunoglobulins in patients infected with SARS-CoV-2.17
Data processing
The measurements in this study are expressed as readout in arbitrary units of mean chemiluminescence intensity originating from different locations of the microfluidic chip. They correspond to the positions of the respective immobilized capture antibodies for each of the biomarkers.
Measurement protocol
Either buffer samples or real synovial fluid samples from one patient were analyzed using the chemiluminescence-based multiplex μ-ELISA system in a point-of-care format. First, a dilution series of the corresponding markers (Table I) was prepared using either washing solution (WL) or a human synovial fluid sample as dilutor (Table II).
Table II.
List of buffer-based and synovial fluid-based samples with different concentrations of alpha-defensin, calprotectin, and interleukin-6.
Sample no. | α-Defensin, µg/ml | Calprotectin, µg/ml | Interleukin-6, µg/ml |
---|---|---|---|
Buffer-based samples | |||
1 | 1 | 100 | 0.1 |
2 | 0.1 | 10 | 0.01 |
3 | 0.01 | 1 | 0.001 |
4 | 0.001 | 0.1 | 0.0001 |
5 | 0.01 | 0.00001 | |
Blank | 0 | 0 | 0 |
Synovial fluid-based samples | |||
1 | 1,000 | 10,000 | 10 |
2 | 100 | 1,000 | 1 |
3 | 10 | 100 | 0.1 |
4 | 1 | 10 | 0.0 |
5 | 0.1 | 1 | 0.001 |
Blank | 0 | 0 | 0 |
For measurements of buffer-based samples, 78 µl of the sample was mixed with 78 µl of the detection antibody solution consisting of the corresponding biotinylated detection antibodies diluted in blocking solution (2 µg/ml). Subsequently, 50 µl of the sample-mix was transferred to the chip inlet using a laboratory pipette. The prepared sample-mix volume was sufficient to perform three replicates of each marker concentration. The rest of the analysis was obtained by the GENSPEED R2 Analyzer applying the test protocol as described in Table III.
Table III.
Automized measurement protocol of GENSPEED R2 Analyzer.
Sample mix | Enzyme solution | Washing solution | Chemiluminescence substrate | Total | |
---|---|---|---|---|---|
Sample | Detection antibody solution | ||||
25 µl | 25 µl | 30 µl | 4 × 10 µl | 30 µl | 150 µl |
10 mins | 5 mins | 4 × 1 min | 2.5 mins | 21.5 mins |
Spiked frozen synovial fluid (4 ml) taken from a patient undergoing arthrocentesis for routine diagnostic process was used and infection was excluded. Joint puncture was performed with sterile technique according to the standard institutional guidelines. The study was conducted according to the guidelines of the Declaration of Helsinki19 and approved by the Institutional Ethics Committee of the Charité – Universitätsmedizin Berlin, Germany (protocol code EA1/026/20, approved on 26 May 2020). For the measurements of samples based on spiked synovial fluid, a pre-dilution step was done to be able to deal with the high viscosity of the sample material. For samples containing alpha-defensin or calprotectin, a 1:100 dilution with washing solution was performed (5 µl spiked synovial fluid + 495 µl WL), whereas for samples containing IL-6 only a 1:10 dilution was used (50 µl spiked synovial fluid + 450 µl WL). All these dilutions were done in a GENSPEED dropper bottle before a special GENSPEED filter dropper was applied to it. Dropping the sample mix through this filter dropper allowed the removal of inhibiting components of the sample material such as erythrocytes or fibres that could negatively affect the flow rate of the sample through the reaction channel of the test chip. Further processing of the filtered sample material was done as described for buffer-based samples. Cut-off values of 1.56 µg/ml, 50 µg/ml, and 0.031 µg/ml for alpha-defensin, calprotectin, and IL-6, respectively, were extrapolated from literature data as optimal cut-offs for the diagnosis of PJI and, consequently, were used for comparison.20-22
Statistical analysis
Mean values and SDs for quantification of limit of blank (LoB) and limit of detection (LoD) were determined. Measure triplicates and respective mean values were considered. Also, the use of five parameter logistic regression for fitting and recalculation of concentrations was employed for statistical comparison in immunoassays. For the evaluation of the biomarker-specific performance of the system, calibration curves were generated using five parameter logistic regression methods.23 LoB and LoD were calculated using the following formulae:
LoB = øblank + 1.645∙σblank
LoD = LoB + 1.645∙σlow conc
Statistical analyses and graphics were conducted using Microsoft Excel for Microsoft 365 (Version 2402; Microsoft, USA), alongside the MyCurveFit Excel Add-in (Version 1.0.415.1001 (210304); MyAssays, UK).
Results
A three-plex μ-ELISA based technology for the simultaneous measurement of three biomarkers (alpha-defensin, calprotectin, and IL-6) has been developed in a model buffer system as well as in spiked human synovial fluid sample. A multiplex immunoassay microfluidic test chip has been designed to achieve a rapid and point-of-care result. First, LoD was determined for each individual biomarker by means of calibration curves with serial dilutions in a buffer-based system. Different concentrations of the three biomarkers are shown in Table II. The concentrations of detection antibody and capture antibody were 2 µg/ml and 0.1 mg/ml, respectively.
LoD of 0.008 µg/ml, 0.002 µg/ml, and 0.00014 µg/ml was determined for alpha-defensin, calprotectin, and IL-6, respectively. All measurements were performed in triplicates. The mean values of all three test results are shown in Figure 3.
Fig. 3
Buffer-based dilution series experiment for human alpha-defensin, calprotectin, and interleukin-6 (IL-6) diluted using GENSPEED washing solution measured with multiplex GENSPEED test chips. The figure displays the chemiluminescent signal intensities (AU) across various biomarker concentrations. The data represent mean values from three measurements: a) alpha-defensin; b) calprotectin; c) IL-6; and d) five parameters logistic (PL) regression. PBS, phosphate-buffered saline.
To assess the performance of this µ-ELISA assay, the experiment was reproduced using spiked human synovial fluid, collected from patients without joint infection. Then, predetermined concentrations of individual biomarker were added (Table II). LoD for alpha-defensin was 0.08 µg/ml and therefore clearly below 1.56 µg/ml, which is an important cut-off threshold for this marker. The same was shown for calprotectin with a LoD of 0.31 µg/ml and a cut-off threshold of 50 µg/ml, as well as for IL-6 with a LoD of 0.004 µg/ml and a cut-off threshold of 0.031 µg/ml. Considering the mean signal intensities of all three replicates, calibration curves for each marker in synovial fluid have been generated as described in the Methods section. The detection ranges are indicated with correspondingly coloured arrows in Figure 4d.
Fig. 4
Dilution series experiments on human synovial fluid spiked with alpha-defension (α-defensin), calprotectin, and interleukin-6 (IL-6) were conducted using multiplex GENSPEED test chips. α-defensin and calprotectin were diluted 1/100, while IL-6 was diluted 1/10 with GENSPEED washing solution. The figure shows chemiluminescent signal intensities (AU) for each biomarker, with all measurements done in triplicates. It includes the limit of detection (LoD) (green line) and a clinically relevant cut-off for periprosthetic joint infection (PJI). Data represent mean values from three tests: a) α-defensin; b) calprotectin; c) interleukin-6; and d) five parameters logistic (PL) regression.
The mean coefficient of variation (CV) values for alpha-defensin, calprotectin, and IL-6 were 10%, 6%, and 5%, respectively (Table IV).
Table IV.
Summary of key results evaluating the performance of the multiplex micro-enzyme-linked immunosorbent assay (µ-ELISA) for the simultaneous detection and quantification of alpha-defensin, calprotectin, and interleukin-6 in spiked synovial fluids.
Parameter | PJI-specific cut-off values | Limit of detection | CV* |
---|---|---|---|
Alpha-defensin | 1.56 µg/ml22 | 0.08 µg/ml† | 10% |
Calprotectin | 50 µg/ml23 | 0.31 µg/ml‡ | 6% |
IL-6 | 0.031 µg/ml24 | 0.004 µg/ml | 5% |
-
*
Mean value of all samples.
-
†
Detection limit Synovasure Lateral Flow Test for alpha-defensin detection: 11.7 µg/ml (for comparison).24
-
‡
Detection limit Lyfstone Lateral Flow Test for calprotectin detection: 14 µg/ml (for comparison).25
-
CV, coefficient of variation; IL-6, interleukin-6; PJI, periprosthetic joint infection.
Discussion
For accurate diagnosis of PJI, a combination of conventional (total white blood cell count, polymorphonuclear percentage)26 and emerging synovial fluid biomarkers (such as alpha-defensin or calprotectin) together with microbiological, histopathological, and molecular examinations is required. Despite the combination of multiple parameters, however, the diagnosis remains sometimes inconclusive. Alpha-defensin is the only biomarker included in the 2018 Philadelphia International Consensus Meeting (ICM) and 2021 European Bone and Joint Infection Society (EBJIS) diagnostic algorithms. Therefore, it was included in our panel. In 2019, the USA Food and Drug Administration (FDA) approved the detection of alpha-defensin in synovial fluid using the alpha-defensin lateral flow assay (Synovasure; Zimmer Biomet, USA) as the first biomarker for the diagnosis of PJI.27 However, a previous meta-analysis has shown that alpha-defensin ELISA immunoassay had superior overall diagnostic value compared with the lateral flow test (area under the curve, 0.98 vs 0.75) with higher sensitivity (96% (90% to 98%) vs 71% (55% to 83%), p < 0.001), but no significant difference in specificity (96% (93% to 97%) vs 90% (81% to 95%), p = 0.060).28 The immunoenzymatic assay was optimized to operate at a cut-off value of 5.2 mg/l, with a lower detection limit of 1.56 µg/ml.20 Although this cut-off has been used in several studies, it has not been officially recognized.29 A recent paper has proposed a high-performance liquid chromatography-mass spectrometry (LC-MS) method for the detection of alpha-defensin in synovial fluid, with good results in terms of accuracy. However, the routine use of this promising method is limited by the fact that it is only available in highly specialized laboratories and is time-consuming.30
Furthermore, previous published data have shown that the performance of this biomarker may be influenced by the clinical definition of PJI used, with a poorer consistency with EBJIS criteria compared to MusculoSkeletal Infection Society (MSIS).15,16 Consequently, regardless of the type of technique used (immunoenzymatic assay, lateral flow, or LC-MS), the use of alpha-defensin as a single marker for the diagnosis of PJI should be discouraged.31-33 Regarding other biomarkers, IL-6 and calprotectin, the choice of these molecules was based on a recent review by Fisher and Patel,9 which ranks these three markers among the five most promising in the field of PJI.
Synovial fluid pro-inflammatory cytokines have a key role in PJI, since they trigger inflammation in response to an infectious process. Among them, IL-6 has been quoted as promising with sensitivity and specificity ranging from 62.5% to 90% and 85.7% to 94.7% in several papers, with different cut-offs (2,100 pg/ml vs 30,750 pg/ml).22,34 According to Lenski and Scherer,22 synovial IL-6 lower than 10,000 pg/ml completely excludes the diagnosis of PJI, while IL-6 levels below 49,000 pg/ml strongly suggests it. However, when between 10,000 and 49,000 pg/ml, the suspicion of PJI remains uncertain, and the simultaneous detection of other markers remains strongly recommended to settle the dilemma of septic versus aseptic inflammation.22 Furthermore, the fact that standard IL-6 assays are designed to assess cytokine levels in serum rather than synovial fluid should be emphasized. Because of this, it is essential that future studies include IL-6 tests, whose consistency and effectiveness have been previously shown in a synovial biomarker investigation.34 Regarding inflammatory proteins, calprotectin has recently been successfully investigated in PJI.35
Different laboratory techniques have been studied to detect calprotectin in synovial fluid, including ELISA, lateral flow tests, and immunoturbidimetric immunoassays.21,36,37 A recent meta-analysis performed on eight studies with a sample size of 618 patients showed an outstanding cumulative diagnostic accuracy for PJI, with a sensitivity of 92% and specificity of 93% and a pooled positive and negative likelihood ratio of 9.91 and 0.07, respectively.38 However, a separate analysis of the single papers revealed a wide discrepancy according to joint site (hip, knee, shoulder prostheses), criteria used for the diagnosis of PJI (ICM 2018, MSIS), laboratory method (ELISA, immunochromatographic assay, immunoturbidimetric), and related cut-offs. Although five of the included studies used 50 µg/ml as the best cut-off using either ELISA,36,39 lateral flow immunoassay,40,41 or both,21 a standardized threshold for calprotectin in the context of PJI has not yet been universally recognized, as well as for the previous biomarkers. As more and more synovial fluid biomarkers have been discovered in recent years, it is becoming increasingly necessary to detect multiple biomarkers from a single sample in a single assay to get appropriate or conclusive findings. For this reason, multipanel technologies are becoming crucial in modern clinical diagnosis, and they are expected to improve the efficacy and accuracy of PJI detection. Therefore, the real challenge would lie in not only developing multiplex biomarker tests that can accurately and consistently distinguish between PJI and aseptic failures, but also in enabling their application in point of care. The most often used point-of-care methods for obtaining quick and ‘on-site’ results are paper-based diagnostic devices because of their versatility, sample manufacturing, visual interpretation of the result, affordability, and ease of transport. The presented assay represents an extremely promising platform since it offers a simultaneous measurement of three of the most promising biomarkers in the diagnosis of PJI, using only a minimal amount of synovial fluid (5 to 50 µl). The microfluidic architecture of the test chip provides a significant advantage in terms of time requirement for the analytical process. The capillary force-driven fluid flow of sample and reagents through the reaction channel, facilitated by the low height of the reaction channel (only 100 μm) which induces shear forces, reduces the time required for diffusion of the involved components and increases the likelihood of biomolecular interactions between the analytes and the immobilized capture probes as they pass through the channel.42 In contrast to laboratory-based ELISAs with turnaround times of two to four hours, this microfluidics-based platform allows for assay procedures to be completed within only 15 to 22 minutes and without employing trained laboratory personnel. Furthermore, the results obtained in the experimental model with spiked synovial fluid were extremely promising, since LoDs far lower than PJI-specific thresholds of other assays normally used in clinical practice were detected (Table IV). In summary, LoD for alpha-defensin was 0.08 µg/ml, thus considerably lower than 1.56 µg/ml. The same was shown for calprotectin with a LoD of 0.31 µg/ml and a cut-off threshold of 50 µg/ml, as well as for IL-6 with a LoD of 0.004 µg/ml and a cut-off threshold of 0.031 µg/ml.
Similar to other studies carried out under experimental conditions, the main limitation of this in vitro study is that it is based on spiked synovial fluid, in which known and progressive concentrations of individual biomarkers were added to mimic as closely as possible the joint environment during infection/inflammation. However, of course, the biomarker levels can be different in real patient samples, as a reason why clinical studies are obligatory to finally validate this assay.
In conclusion, the described multiplex µ-ELISA system considerably outperforms standard lateral flow point-of-care tests regarding multiplexing capabilities in terms of assay quality aspects. We therefore believe that the unique features of simultaneous measurement of the three most promising synovial biomarkers in PJI diagnosis using minimal sample volume, lower LoDs compared to conventional tests, and speed of processing can give the novel multiplex µ-ELISA platform, used in this study, great diagnostic potential as a point-of-care tool to help the clinician gain a broader picture for characterizing synovial fluids from PJI/aseptic failure doubtful cases.
However, while the system demonstrates promising benefits, it is important to acknowledge that clinical studies are necessary to confirm its effectiveness and validate its practical use in real-world clinical settings.
References
1. Patel R . Periprosthetic Joint Infection . N Engl J Med . 2023 ; 388 ( 3 ): 251 – 262 . Crossref PubMed Google Scholar
2. Rajput V , Meek RMD , Haddad FS . Periprosthetic joint infection: what next? Bone Joint J . 2022 ; 104-B ( 11 ): 1193 – 1195 . Crossref PubMed Google Scholar
3. van Sloten M , Gómez-Junyent J , Ferry T , et al. Should all patients with a culture-negative periprosthetic joint infection be treated with antibiotics? A multicentre observational study . Bone Joint J . 2022 ; 104-B ( 1 ): 183 – 188 . Crossref PubMed Google Scholar
4. Strimbu K , Tavel JA . What are biomarkers? Curr Opin HIV AIDS . 2010 ; 5 ( 6 ): 463 – 466 . Crossref PubMed Google Scholar
5. Heilmann E , Gregoriano C , Schuetz P . Biomarkers of infection: are they useful in the ICU? Semin Respir Crit Care Med . 2019 ; 40 ( 4 ): 465 – 475 . Crossref PubMed Google Scholar
6. Vrancianu CO , Serban B , Gheorghe-Barbu I , et al. The challenge of periprosthetic joint infection diagnosis: from current methods to emerging biomarkers . Int J Mol Sci . 2023 ; 24 ( 5 ): 4320 . Crossref PubMed Google Scholar
7. Maritati M , De Rito G , Rosta V , et al. Active myeloperoxidase: a promising biomarker to differentiate “acute” and “low-grade” peri-prosthetic joint infections from aseptic failures . Front Microbiol . 2024 ; 15 : 1417049 . Crossref PubMed Google Scholar
8. Cai Y , Liang J , Chen X , et al. Synovial fluid neutrophil extracellular traps could improve the diagnosis of periprosthetic joint infection . Bone Joint Res . 2023 ; 12 ( 2 ): 113 – 120 . Crossref PubMed Google Scholar
9. Fisher CR , Patel R . Profiling the immune response to periprosthetic joint infection and non-infectious arthroplasty failure . Antibiotics . 2023 ; 12 ( 2 ): 296 . Crossref PubMed Google Scholar
10. Li J , Macdonald J . Multiplexed lateral flow biosensors: technological advances for radically improving point-of-care diagnoses . Biosens Bioelectron . 2016 ; 83 : 177 – 192 . Crossref PubMed Google Scholar
11. Yin B , Yue W , Sohan ASMMF , et al. Construction of a desirable hyperbolic microfluidic chip for ultrasensitive determination of PCT based on chemiluminescence . J Mater Chem B . 2023 ; 11 ( 9 ): 1978 – 1986 . Crossref PubMed Google Scholar
12. Bhuiyan NH , Uddin MJ , Lee J , Hong JH , Shim JS . An internet-of-disease system for COVID-19 testing using saliva by an AI-controlled microfluidic ELISA device . Adv Mater Technol . 2022 ; 7 ( 9 ): 2101690 . Crossref PubMed Google Scholar
13. Mou L , Dong R , Hu B , Li Z , Zhang J , Jiang X . Hierarchically structured microchip for point-of-care immunoassays with dynamic detection ranges . Lab Chip . 2019 ; 19 ( 16 ): 2750 – 2757 . Crossref PubMed Google Scholar
14. Deirmengian C , Kardos K , Kilmartin P , Cameron A , Schiller K , Parvizi J . Combined measurement of synovial fluid α-Defensin and C-reactive protein levels: highly accurate for diagnosing periprosthetic joint infection . J Bone Joint Surg Am . 2014 ; 96-A ( 17 ): 1439 – 1445 . Crossref PubMed Google Scholar
15. Bottner F , Wegner A , Winkelmann W , Becker K , Erren M , Götze C . Interleukin-6, procalcitonin and TNF-alpha: markers of peri-prosthetic infection following total joint replacement . J Bone Joint Surg Br . 2007 ; 89-B ( 1 ): 94 – 99 . Crossref PubMed Google Scholar
16. Tsai T-T , Huang T-H , Ho NY-J , Chen Y-P , Chen C-A , Chen C-F . Development of a multiplex and sensitive lateral flow immunoassay for the diagnosis of periprosthetic joint infection . Sci Rep . 2019 ; 9 ( 1 ): 15679 . Crossref PubMed Google Scholar
17. Wechselberger C , Süßner S , Doppler S , Bernhard D . Performance evaluation of serological assays to determine the immunoglobulin status in SARS-CoV-2 infected patients . J Clin Virol . 2020 ; 131 : 104589 . Crossref PubMed Google Scholar
18. Kerschner H , Harrison EM , Hartl R , Holmes MA , Apfalter P . First report of mecC MRSA in human samples from Austria: molecular characteristics and clinical data . New Microbes New Infect . 2015 ; 3 : 4 – 9 . Crossref PubMed Google Scholar
19. World Medical Association . World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects . JAMA . 2013 ; 310 ( 20 ): 2191 – 2194 . Crossref PubMed Google Scholar
20. Bonanzinga T , Ferrari MC , Tanzi G , Vandenbulcke F , Zahar A , Marcacci M . The role of alpha defensin in prosthetic joint infection (PJI) diagnosis: a literature review . EFORT Open Rev . 2019 ; 4 ( 1 ): 10 – 13 . Crossref PubMed Google Scholar
21. Grassi M , Salari P , Farinelli L , D’Anzeo M , Onori N , Gigante A . Synovial biomarkers to detect chronic periprosthetic joint infection: a pilot study to compare calprotectin rapid test, calprotectin ELISA immunoassay and leukocyte esterase test . J Arthroplasty . 2022 ; 37 ( 4 ): 781 – 786 . Crossref PubMed Google Scholar
22. Lenski M , Scherer MA . The significance of interleukin-6 and lactate in the synovial fluid for diagnosing native septic arthritis . Acta Orthop Belg . 2014 ; 80 ( 1 ): 18 – 25 . PubMed Google Scholar
23. Dunn J , Wild D . Chapter 3.6 - Calibration Curve Fitting . In : Wild D . ed . The Immunoassay Handbook: Theory and Applications of Ligand Binding, ELISA and Related Techniques . Fourth ed . Oxford : Elsevier Science , 2013 : 323 – 336 . Crossref Google Scholar
24. No authors listed . Evaluation of Automatic Class III Designation for Synovasure Alpha Defensin Lateral Flow Test Kit Decision Summary . https://www.accessdata.fda.gov/cdrh_docs/reviews/DEN180032.pdf ( date last accessed 27 January 2025 ). Google Scholar
25. Trotter AJ , Dean R , Whitehouse CE , et al. Preliminary evaluation of a rapid lateral flow calprotectin test for the diagnosis of prosthetic joint infection . Bone Joint Res . 2020 ; 9 ( 5 ): 202 – 210 . Crossref PubMed Google Scholar
26. Jandl NM , Kleiss S , Mussawy H , Beil FT , Hubert J , Rolvien T . Absolute synovial polymorphonuclear neutrophil cell count as a biomarker of periprosthetic joint infection . Bone Joint J . 2023 ; 105-B ( 4 ): 373 – 381 . Crossref PubMed Google Scholar
27. Deirmengian C , Madigan J , Kallur Mallikarjuna S , Conway J , Higuera C , Patel R . Validation of the alpha defensin lateral flow test for periprosthetic joint infection . J Bone Joint Surg Am . 2021 ; 103-A ( 2 ): 115 – 122 . Crossref PubMed Google Scholar
28. Eriksson HK , Nordström J , Gabrysch K , Hailer NP , Lazarinis S . Does the alpha-defensin immunoassay or the lateral flow test have better diagnostic value for periprosthetic joint infection? A meta-analysis . Clin Orthop Relat Res . 2018 ; 476 ( 5 ): 1065 – 1072 . Crossref PubMed Google Scholar
29. Yuan J , Yan Y , Zhang J , Wang B , Feng J . Diagnostic accuracy of alpha-defensin in periprosthetic joint infection: a systematic review and meta-analysis . Int Orthop . 2017 ; 41 ( 12 ): 2447 – 2455 . Crossref PubMed Google Scholar
30. Balato G , Dall’Anese R , Balboni F , et al. Synovial fluid alpha-defensin in periprosthetic knee infection workup: liquid chromatography-mass spectrometry detection of alpha-defensin in synovial fluid . Bone Joint J . 2022 ; 104-B ( 9 ): 1047 – 1051 . Crossref PubMed Google Scholar
31. Lee YS , Koo K-H , Kim HJ , et al. Synovial fluid biomarkers for the diagnosis of periprosthetic joint infection: a systematic review and meta-analysis . J Bone Joint Surg Am . 2017 ; 99-A ( 24 ): 2077 – 2084 . Crossref PubMed Google Scholar
32. Shahi A , Parvizi J . The role of biomarkers in the diagnosis of periprosthetic joint infection . EFORT Open Rev . 2016 ; 1 ( 7 ): 275 – 278 . Crossref PubMed Google Scholar
33. Sigmund IK , Yermak K , Perka C , Trampuz A , Renz N . Is the enzyme-linked immunosorbent assay more accurate than the lateral flow alpha defensin test for diagnosing periprosthetic joint infection? Clin Orthop Relat Res . 2018 ; 476 ( 8 ): 1645 – 1654 . Crossref PubMed Google Scholar
34. Randau TM , Friedrich MJ , Wimmer MD , et al. Interleukin-6 in serum and in synovial fluid enhances the differentiation between periprosthetic joint infection and aseptic loosening . PLoS ONE . 2014 ; 9 ( 2 ): e89045 . Crossref PubMed Google Scholar
35. Cheok T , Smith T , Siddiquee S , Jennings MP , Jayasekera N , Jaarsma RL . Synovial fluid calprotectin performs better than synovial fluid polymerase chain reaction and interleukin-6 in the diagnosis of periprosthetic joint infection: a systematic review and meta-analysis . Bone Joint J . 2022 ; 104-B ( 3 ): 311 – 320 . Crossref PubMed Google Scholar
36. Salari P , Grassi M , Cinti B , Onori N , Gigante A . Synovial fluid calprotectin for the preoperative diagnosis of chronic periprosthetic joint infection . J Arthroplasty . 2020 ; 35 ( 2 ): 534 – 537 . Crossref PubMed Google Scholar
37. Grzelecki D , Walczak P , Szostek M , Grajek A , Rak S , Kowalczewski J . Blood and synovial fluid calprotectin as biomarkers to diagnose chronic hip and knee periprosthetic joint infections . Bone Joint J . 2021 ; 103-B ( 1 ): 46 – 55 . Crossref PubMed Google Scholar
38. Hantouly AT , Salameh M , Toubasi AA , et al. Synovial fluid calprotectin in diagnosing periprosthetic joint infection: a meta-analysis . Int Orthop . 2022 ; 46 ( 5 ): 971 – 981 . Crossref PubMed Google Scholar
39. Warren J , Anis HK , Bowers K , et al. Diagnostic utility of a novel point-of-care test of calprotectin for periprosthetic joint infection after total knee arthroplasty: a prospective cohort study . J Bone Joint Surg Am . 2021 ; 103-A ( 11 ): 1009 – 1015 . Crossref PubMed Google Scholar
40. Wouthuyzen-Bakker M , Ploegmakers JJW , Kampinga GA , Wagenmakers-Huizenga L , Jutte PC , Muller Kobold AC . Synovial calprotectin: a potential biomarker to exclude a prosthetic joint infection . Bone Joint J . 2017 ; 99-B ( 5 ): 660 – 665 . Crossref PubMed Google Scholar
41. Wouthuyzen-Bakker M , Ploegmakers JJW , Ottink K , et al. Synovial calprotectin: an inexpensive biomarker to exclude a chronic prosthetic joint infection . J Arthroplasty . 2018 ; 33 ( 4 ): 1149 – 1153 . Crossref PubMed Google Scholar
42. Squires TM , Messinger RJ , Manalis SR . Making it stick: convection, reaction and diffusion in surface-based biosensors . Nat Biotechnol . 2008 ; 26 ( 4 ): 417 – 426 . Crossref PubMed Google Scholar
Author contributions
M. Maritati: Writing – original draft
M. Vogl: Conceptualization, Formal analysis, Methodology, Writing – original draft, Writing – review & editing
M. Sonnleitner: Conceptualization, Methodology, Writing – original draft, Writing – review & editing
A. Trampuz: Conceptualization, Supervision, Writing – review & editing
Funding statement
The author(s) disclose receipt of the following financial or material support for the research, authorship, and/or publication of this article: this research was supported by: an unrestricted grant from Arthrex GmbH (Munich, Germany), as reported by M. Vogl, M. Sonnleitner, and A. Trampuz; and PRO-IMPLANT Foundation (www.pro-implant.org), a non-profit organization supporting research and education in life sciences and healthcare, as reported by M. Maritati.
ICMJE COI statement
M. Maritati reports sponsorship of the presentation by Arthrex GmbH, not related to this study. M. Maritati also reports funding from the PRO-IMPLANT Foundation, related to this study. M. Vogl is Head of Assay Development at GENSPEED Biotech GmbH. M. Sonnleitner is CEO of GENSPEED Biotech GmbH. M. Vogl, M. Sonnleitner, and A. Trampuz report a grant from Arthrex GmbH, related to this study. The study was supported by Arthrex GmbH (Munich, Germany), which had no role in the design, execution, interpretation, writing, or presentation of the results of the study.
Data sharing
The data that support the findings for this study are available to other researchers from the corresponding author upon reasonable request.
Acknowledgements
The authors express their gratitude to Arthrex for funding, with special recognition to Christoph Elser and Sebastian Konrad for ongoing discussions throughout the project.
Ethical review statement
The study was conducted according to the guidelines of the Declaration of Helsinki and approved by the Institutional Ethics Committee of Charité – Universitätsmedizin Berlin, Germany (protocol code EA1/026/20, approved on 26 May 2020).
Open access funding
The authors report that they received open access funding for their manuscript from GENSPEED Biotech GmbH.
© 2025 Maritati et al. This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial No Derivatives (CC BY-NC-ND 4.0) licence, which permits the copying and redistribution of the work only, and provided the original author and source are credited. See https://creativecommons.org/licenses/by-nc-nd/4.0/